ECU LogoEHST 3701: Industrial Hygiene Laboratory

LABORATORY EXERCISE ON LIGHT MONITORING

I. Introduction

II. ObjectiveS

At the end of the session each student should be able:

1. To conduct illumination measurements

2. To analyze and interpret the results obtained

3. To give recommendations based on findings

III. Equipment and materials

1. Light meter, Extech Model EA30

IV. Methodology

A. Use of the Instrument

1. Make sure that the protective cap is on the white domed light sensor. Press the orange power button to turn power on. The meter will automatically perform a zero calibration.

2. Select the desired unit of measure (Lux or Fc). Select LUX for this experiment.

3. Remove the protective cap from the sensor and place the white domed light sensor on the surface (i.e. counter top) you would like to monitor. The light being measured must encompass the entire white sensor dome. NOTE: Do not block the light source while reading the display.

4. Read the light level on the LCD. If the light measurement is out of range, an “OL” indication will appear on the display. If the “OL” appears, select another range by pressing the RANGE button.

5. Take note of the illumination level measurements (minimum and maximum).

a. Press the MX/MN button 3 times to have the meter continue to record the highest and lowest readings (MX and MN icons is displayed on the upper left-hand portion of the LCD).

b. Press the MX/MN button once to put the meter in MX mode to display only the maximum reading (MX is displayed).

c. Press the MX/MN button again to put the meter in MN mode to display only the minimum reading (MN is displayed).

d. To exit the MX MN mode, press and hold the MX/MN button until the MX and/or MN icons disappear.

6. Press the orange power button to turn power off.

B. Illumination Monitoring

Sampling location: Laboratory

1. Perform a brief walk-through of the areas for monitoring.

2. Draw a layout of the area to be monitored.

3. Discuss and determine the location of sampling points. Measure at least 15 sampling points. Prioritize the working surfaces that are likely to be used by the lab user.

4. Perform actual illumination measurement using the light meter. Obtain the illumination levels in lux unit and slow response time.

5. Record data obtained from the monitoring and develop an illumination map.

V. Points For discussion

Note: Please use complete sentences when addressing the questions below for your discussion.

1. Describe the illumination levels in the area monitored (i.e. where the lowest and highest lighting occurs, source of lighting, etc.)

2. Are the illumination levels measured acceptable? What recommended illumination level/s did you compare your results with? Explain why you chose those levels.

3. What control measures do you recommend?

VI. RECOMMENDED ILLUMINATION LEVELS

Activity

Illumination (lux)

Public areas with dark surroundings

20 – 50

Simple orientation for short visits

50 – 100

Working areas where visual tasks are only occasionally performed

100 – 200

Performance of visual tasks of high contrast or large size

200 – 500

Performance of visual tasks of medium contrast or small size

500 – 1000

Performance of visual tasks of low contrast or very small size

1000 - 2000

Performance of visual tasks of low contrast and very small size for prolonged periods of time

2,000 – 5,000

Performance of very prolonged and exacting visual tasks

5,000 – 10,000

Performance of very special visual tasks of extremely low contrast and small size

10,000 – 20,000

Activity

Minimum Illumination (lux)

Warehouses, homes, theaters, archives

150

Easy office work, classes

250

Normal office work, PC work, study library, groceries, show rooms, laboratories

500

Supermarkets, mechanical workshops, office landscapes

750

Normal drawing work, detailed mechanical workshops, operation theaters

1,000

Detailed drawing work, very detailed mechanical works

1,500 – 2,000

VII. CONCLUSION

ILLUMINATION mAP Illumination Monitoring Worksheet

Date of Monitoring: ___________________________________________

Instrument Used: ____________________________________________

Sampling Location

Sampling Pt. No.

Operation / Activity

Time of Msmt

Illumination level (lux)

Remarks

Min

Max

Carol belk building room 3501 for sample 1-15

12:06pm

890

968

12:08pm

886

1013

12:09pm

1022

1030

12:10pm

656

879

12:11pm

684

692

12:12pm

1140

1163

12:13pm

1201

1231

12:14pm

1072

1123

12:15pm

853

904

12:15pm

1002

1035

12:16pm

1005

1014

12:16pm

807

827

12:17pm

155

205

12:18pm

755

782

12:19pm

998

1055