Lab extended abstract

profileJack11
exp-6-temperature_measurement_and_calibration_s161.pdf

Last Rev.: August 2014 Calibration and Temperature Measurement Page 2

ME 495—Thermo Fluids Laboratory ~~~~~~~~~~~~~~

Temperature Measurement and First-

Order Dynamic Response ~~~~~~~~~~~~~~

PREPARED BY: GROUP LEADER’S NAME

LAB PARTNERS: NAME

NAME

NAME

TIME/DATE OF EXPERIMENT: TIME , DATE ~~~~~~~~~~~~~~

OBJECTIVE — The objectives of this laboratory are:

• To learn basic concepts and definitions associated with the temperature and temperature measurements.

• To learn how to calibrate a Thermocouple and a Thermistor. • To determine and compare the time constants of a

thermocouple and a thermometer.

• To determine how a thermocouple and a thermometer responds to different inputs. You will also observe the

response of a thermocouple to an oscillatory input.

• To develop awareness for sources of error in temperature measurements.

THEORY – In this lab, we will use first-order models to approximate the response of a thermometer, thermocouple, and a

thermistor to temperature inputs, as these temperature sensors

measure temperatures in a different way.

 A thermometer senses a change in temperature as a change in the density of a fluid.

 A thermocouple consists of two wires of different metals joined at one end (the junction). When a voltage is applied

across the free ends of the two wires, the differing properties

of the wires create an induced voltage that it proportional to

the temperature change at the junction.

 A thermistor is a type of resistor whose resistance is dependent on temperature, more so than in standard resistors.

The change in resistance is linear with respect to change in

temperature, thus making a thermistor an accurate

temperature measuring device.

EXPERIMENT PREPARATION - Get a thermometer, a K (or J) type thermocouple, and a thermistor from the TA. Identify the

positive and negative terminals for the thermocouple.

• Verify that the thermocouple is functioning well. This can be done by connecting the thermocouple to a DMM and ensuring that

the voltage changes when you hold the thermocouple weld

between your fingers.

• Be familiar with all of the instruments you will be using for this experiment. Knowing your equipment well is essential.

• Prepare an ice bath. Most EMF (electromotive force) tables use

ice point (0C) as the reference temperature and this traditional

fixed point temperature is preferred for accurate and reliable

measurements. To prepare the ice bath:

o Crush or flake the ice (Ice is available in the white icebox located on the measurement table).

o Fill the thermos (the blue with white lid) half with crushed-ice, add water and stir it until the mixture becomes a slush without

having the ice float. [Recall: If the ice floats, the bottom

temperature could be higher than 0C –Anomalous expansion of

water.]

PROCEDURE - Part 1: Modify a VI for temperature measurements In this lab, you will build on the skills that were developed in

Introduction to Benchtop Equipment and Data Acquisition. It is

assumed that the student understands the concepts of a Front

Panel, a Block Diagram, the Control Palette, and the Functions

Palette. It is also assumed that the student can search the control

and functions palette for necessary features.

1. Connect either terminal of the thermocouple to either of the

positive or negative terminals of the TC0 channel of the 9213

module.

2. Open LabVIEW 10.0.

3. Open the VI that you developed for Introduction to Benchtop

Equipment and Data Acquisition.

4. Click File >> Save As… >> Substitute Copy for Original to

save your VI with a new name.

5. On the Block Diagram, double click the DAQ Assistant to

open the configuration dialogue box.

6. Configure the DAQ Assistant as shown in Figure 1.

a. DELETE all previous measurement channels.

c. Add a new measurement channel.

 Click Add Channels >> Voltage.

 Select the channel that the thermocouple is connected to NI-9213>>ai0.

 Click OK. d. The input range should be 1.0mV max to -1.0mV min,

but this range can be adjusted if needed.

(Thermocouples have a low voltage output.)

e. The Terminal Configuration should be Differential.

f. Change Acquisition Mode to “Continuous Samples.”

Note: A dialog box will open asking if you would like a While

Loop to be made around the DAQ Assistant. Click No. (You

will place a While Loop in the next step that includes the

subVIs that are already on the Block Diagram.

Figure 1: DAQ Assistant Configuration Box

7. Place a While Loop around the entire Block Diagram as

shown below. LabVIEW While Loops are similar to While

Loops in any text-based programming language. The routine

inside the loop will run repeatedly until a true value is passed to

the loop condition, represented by a red button at the bottom

right corner of the loop.

Last Rev.: August 2014 Calibration and Temperature Measurement Page 3

8. A toggle button can be used to end the While Loop routine.

Right click on the loop condition icon and click

Create>>Control.

9. Double click the Write to Measurement File ExpressVI to

open the configuration box. Under If a file already exists, select

Append to File. Click OK. (Because the Write to File function

now occurs in a loop, the data must be appended to the end of

the file, rather than overwriting the data from the previous

loop.)

10. On the front panel, place a Thermometer.

11. Right Click the Thermometer. Click Visible Items >>

Digital Display.

12. On the Block Diagram, inside the While Loop, place a Time

Delay Express VI. When the Configuration Box opens, change

the Time Delay (seconds) to 0.001. (Without a delay in the

While Loop, the program will repeat as quickly as it can and

consume processor power.)

13. Your Block Diagram should resemble Figure 3.

14. Click File >> Save As >> Copy >> Create Unopened Disk

Copy. Save the VI with “Lab#” in the file name. (You will use

the VI as it is now as a starting point in Operational Amplifiers

and Signal Conditioning).

Figure 2: Block diagram with while loop

Figure 3: Completed block diagram

Part 2: Calibration Curve

You will create a static calibration curve to correlate voltage

measurements with temperature readings. We will use the

thermometer as the standard. (Note: We use the thermometer as a

standard for convenience. Do not assume that an alcohol-in-glass

thermometer is a temperature measurement device with superior

accuracy.)

2.1 Calibrate the thermocouple

1. Take temperature readings with the thermometer in ice water

and boiling water. Be careful not to allow the devices to come into

contact with the hot plate or the bottom of the beaker.

2. Record the temperatures measured.

3. Use the thermocouple to measure both temperatures. Again, be

careful not to allow the devices to come in contact with the hot

plate or the bottom of the beaker.

4. In Excel:

1. a. Plot the temperatures measured with the known standard

values on the y-axis and the ADC output (voltage) resulting

from the thermocouple input on the x-axis.

b. Create a linear curve fit using the two data points.

The linear fit results in an equation that relates the ADC

output obtained from the thermocouple to the known

temperature from the thermometer:

T=k(VTC) + b

where “T” is the temperature, “k” is the gain (°C/V), “Vadc” is the

voltage read in to the computer, and “b” is the zero-voltage offset.

The gain “k” is the inverse of the static sensitivity, which has units

of V/ °C.

2.2 Modify your VI for calibration

Figure 4: Modified block diagram

In this section, you will multiply the data by the conversion scalar

“k” that you found experimentally. You will also add an offset

“b”.

1. Delete the Tone Measurement ExpressVI. Delete the

Frequency and Amplitude indicators.

2. Delete the wires that connect the DAQ Assistant to the

thermometer, the Waveform Graph, and the Write to File

ExpressVI. Press Ctrl+b to remove any broken wire remnants.

3. Place a Multiply and an Add on the block diagram.

4. Create a control and connect it to the y terminal of the

Multiply icon. Rename the control “k”.

5. Create a control and connect it to the y terminal of the Add

icon. Rename the control “b”.

6. Wire your block diagram as shown in Figure 4.

7. On the front panel, enter the values you found experimentally

for the “k” and “b” values.

8. When you run the VI, your measurements should now be

displayed as temperatures.

NOTE: This VI will also be used for the calibration of the

thermistor, with the exception that the DAQ Assistant will be

modified.

Modifying VI for Thermistor

Configure the DAQ Assistant as shown below in Figure 5.

a) DELETE all previous measurement channels. b) Add a “Vex Thermistor” measurement. c) Select the appropriate channels on the 9021 module (ex:

channel ai0).

d) Set the max and min temperature to 100 and 0 C respectively.

e) For the A, B, and C constants, input 1.295361E-3, 2.343159E-4, and 1.018703E-7 respectively.

f) For the R1 value, input 10k (i.e. 10,000 ohms). Ensure that the configuration is set to “4-Wire.”

g) For the Vex Value (V), input 5V. h) Ensure that the Acquisition Mode is set to “Continuous

Samples.”

Last Rev.: August 2014 Calibration and Temperature Measurement Page 4

Figure 1: Modified DAQ Assistant for Thermistor

Part 3: Time Constants

The calibration curves you found represent static calibrations. In

the following section, you will consider the dynamic behavior of

the sensors. The dynamics of all the temperature sensors used in

this lab can be modeled accurately as first-order systems. The

natural response of a first-order system can be characterized by a

single parameter - the time constant. Time constants can be found

by observing the response of the instrument to a step input in

temperature.

For each of the three sensors, you will determine the time

constant for transfers from ambient air to ice water and from hot

air to ambient air.

An ice water bath and a heat gun (to generate hot air) will be

provided for your tests. Caution: Keep all temperature sensors

(particularly the thermometer and the thermistor) at least 9 inches

away from the heat gun. Holding them closer to the heat gun will

result in temperatures outside the operating range of the sensors.

Three methods will be used to calculate various time constants.

Table 1 summarizes the tests to be performed. The testing methods

are described below.

Table 1. Methods to be used to calculate time constants.

thermometer thermocouple thermistor

Ambient air to

ice water

1, 2 3, 4 2, 3

Hot air to

ambient air

1, 2 3 2, 3

Ice water to

ambient air

- 3 2, 3

NOTE: Methods 1 through 4 are listed and described below.

3.1 Method 1- Time Elapsed at 63.2% Sensor Accuracy

Thermometer: Perform this test for ambient air to ice water.

Repeat for hot air to ambient air.

1. Record the temperature of the initial medium, T0, and the

temperature of the final medium, T.

2. Calculate the intermediate reference temperature, Tr that is

63.2 percent of the way from T0 to T.

3. With the instrument at steady state (reading T0) in the initial

medium, move it quickly to the final medium.

4. The time from when the step in temperature began to when

the instrument reads Tr is the time constant, τ. This time can

be measured using a stopwatch.

3.2 Method 2 - Mathematic Definition of τ (Tau)

Thermometer: Perform this test moving from ambient air to ice

water.

1. Record the temperature of the initial medium, T0, and the

temperature of the final medium, T.

2. Choose an arbitrary reference temperature, Tr that is between

60 and 80 percent of the way from T0 to T.

3. With the instrument at steady state (reading T0) in the initial

medium, move it quickly to the final medium.

4. Record the time tr that it takes to reach the reference

temperature Tr.

5. The following expression can be used to calculate the time

constant tau:

0TT

TT e r

tr

 

  

  

 

 (1)

3.3 Method 3 - Analyze Acquired Data

Thermocouple: Perform this test for ambient air-to-ice water, hot

air-to-ambient air, and ice water-to ambient air.

1. Run the VI.

2. Insert the thermocouple into the initial medium.

3. Click the Enable toggle control as you transfer the

thermocouple to the final medium.

4. When the sensor comes to equilibrium, stop the VI.

5. Open your measurement file in Excel.

6. Determine the time required for the sensor to complete 63.2

percent of the transient from the initial temperature to the

final temperature.

7. Change the file path to prevent overwriting data. (You can

either change the name of the file after it is written, or you

can change the name of the path that the next _le will be

written to.) You will use the saved data for subsequent

calculations.

3.4 Method 4- Error Fraction

Thermocouple: Complete this step using the data from the ambient

air-to-ice water transfer.

Equation 2 below gives an expression for the error fraction.

   

0TT

tTT t

 

 (2)

1. Calculate the error fraction at each instant of time for the data

you have gathered.

2. Take the natural log of the error fraction data.

3. Plot the natural log of the error fraction versus time.

4. Find the negative reciprocal of the slope of this curve. This is

the time constant.

The error fraction versus time curve provides an easy way to

judge the accuracy of a first-order model for the data gathered.

If the error fraction plot is linear, then the first-order model is a

good approximation of the dynamic behavior of the system.

(See pages 87-88 of Figliola and Beasley for a more detailed

discussion of the error fraction and its use to calculate the time

constant.) Is the thermocouple in water adequately modeled by

the first-order system approximation?

Part 4: Additional Tests

Once you have performed all of the tests outlined above,

consider the following:

 Transfer the thermocouple from ice water to room temperature air.

How does the time constant compare with the one obtained

from the air-to-ice-water transition? Is it different? Why or

why not?

 When calculating the time constant for the thermocouple in transition from air to ice water, you measured the time to go

63.2 percent of the way from the initial temperature (around

Last Rev.: August 2014 Calibration and Temperature Measurement Page 5

21°C) to the final temperature (around 0°C). Using the same

graph, recalculate the time constant using14°C as the starting

temperature (i.e., what is the time required to go 63.2 percent

of the way from 14°C to 0°C?). Is this time constant value

significantly different from the initial value you computed?

Why or why not?

Part 5: Thought Questions

Document your response to these questions in your lab report.

1. How do the time constants of the thermometer and

thermocouple compare? Why is the thermometer “slow”

compared to the thermocouple?

2. How do the time constants obtained for air and water

compare? Are they significantly different? If so, explain why.

3. Does the instrument need to be at steady state before

transferring it to the final medium?

4. How accurate (qualitatively) is the first-order model

approximation for the different sensors and media?

Part 6: Frequency Response

From the measured time responses of the thermometer and the

thermocouple, it can be seen that the response time of these

sensors is not instantaneous. From this we can infer that these

measurement systems may have some difficulty following inputs

that change quickly. Frequency response is a means of

characterizing the dynamic response capabilities of a system.

The frequency response of a system can be determined by

calculating the response of a system to sinusoidal inputs at

different frequencies. Two important characteristics are:

 the magnitude ratio between the output and input waveforms

 the phase difference between the output and input waveforms

For a first-order system, we can create the frequency response

plot using only our knowledge of the system time constant. The

magnitude ratio can be calculated from the following expression.

   21

1

 

M (4)

The phase difference is determined from equation 5

    1tan (5) In Excel, create a frequency response plot for the thermometer,

thermocouple, and thermistor using the time constant from the ice

water tests. Plot the magnitude ratio on a log-log scale and the

phase difference on a semilog scale (log scale of frequency axis).

Consider frequencies from 0.1 rad/s to 1000 rad/s. Your plots will

look nice if you use ten points per decade of frequency (i.e., 0.1,

0.2, 0.3, : : : ,1, 2, 3, : : : , 10, 20, 30, : : :). Overlay the three

magnitude responses on one plot and the three phase responses on

another. How do the frequency response characteristics of the

thermometer, thermocouple, and thermistor compare?

Part 7: Dynamic Temperature Measurement Problem

Include the analysis for this problem in your lab report.

Assume that the thermocouple you used is placed in a water

stream at a point where the temperature varies sinusoidally

between 60°C and 80°C at a frequency of 40 rad/s. Assuming that

the thermocouple time constant is the same as the value you

determined in the ice water case, what will be the range of

sinusoidal variation in the temperature readings? Repeat the

calculations for a temperature frequency of 2 rad/s.

Now assume the thermocouple is in air instead of water.

Determine the range of the variation using input frequencies of 10

rad/s and 1 rad/s.

QUESTIONS - Document your response to these questions and those asked above in your lab report.

1) How do the time constants of the thermometer,

thermocouple, and thermistor compare? Why is the

thermometer “slow” compared to the other devices?

2) How do the time constants obtained for air and water

compare? Are they significantly different? If so, explain why.

3) Does the instrument need to be at steady state before

subjecting it to the step to the final medium?

4) How accurate (qualitatively) is the first-order model

approximation for the different sensors and media?

5) What is cold junction compensation?

6) What are sources of measurement error?

When submitting the report, each team member must also

submit a peer evaluation form. The form is in the appendix of

this handout.

Clean Up:

Please straighten up your lab work area. Turn off the data

acquisition hardware and leave your work area clean. Your help in

keeping the lab neat is appreciated.

(3)

Name: ME 495 Lab

Group #:

Peer Evaluation Grade your teammates – be honest: A – Work is exemplary, exciting, engaging. This student made a positive, active, and essential contribution to the team. Outstanding effort

B – Student was a willing participant. Contribution was positive and exactly what was expected. Very good effort and solid work.

C – Fair to average effort. Only worked on tasks when they were assigned. Not much volunteering.

D – Irresponsible and didn’t contribute to the team effort. Hurt team.

F – Who is this person? Were they on our team? Never heard of him/her….. Grade Name