Measurement Error in Grip and Pinch Force Measurements in Patients With Hand Injuries

Background and Purpose. There is limited documentation of measurement error of grip and pinch force evaluation methods. The purposes of this study were (1) to determine indexes of measurement error for intraexaminer and inter- examiner measurements of grip and pinch force in patients with hand injuries and (2) to investigate whether the measurement error differs between measure- ments of the injured and noninjured hands and between experienced and inexperienced examiners. Subjects. The subjects were a consecutive sample of 33 patients with hand injuries who were seen in the Department of Rehabilita- tion Medicine of Erasmus MC–University Medical Center Rotterdam in the Netherlands. Methods. Repeated measurements were taken of grip and pinch force, with a short break of 2 to 3 minutes between sessions. For the grip force in 2 handle positions (distance between handles of 4.6 and 7.2 cm, respectively), tip pinch (with the index finger on top and the thumb below, with the other fingers flexed) and key pinch force (with the thumb on top and the radial side of the index finger below) data were obtained on both hands of the subjects by an experienced examiner and an inexperienced examiner. Intraclass correla- tion coefficients (ICCs), standard errors of measurement (SEMs), and associ- ated smallest detectable differences (SDDs) were calculated and compared with data from previous studies. Results. The reliability of the measurements was expressed by ICCs between .82 and .97. For grip force measurements (in the second handle position) by the experienced examiner, an SDD of 61 N was found. For tip pinch and key pinch, these values were 12 N and 11 N, respectively. For measurements by the inexperienced examiner, SDDs of 56 N for grip force and 13 N and 18 N for tip pinch and key pinch were found. Discussion and Conclusion. Based on the SEMs and SDDs, in individual patients only relatively large differences in grip and pinch force measurements can be adequately detected between consecutive measurements. Measurement error did not differ between injured and noninjured hands or between experienced and inexperienced examiners. Criteria for judging whether the measurement error does allow application of the measurements in individual patients are discussed. [Schreuders TAR, Roebroeck ME, Goumans J, et al. Measurement error in grip and pinch force measurements in patients with hand injuries. Phys Ther. 2003;83:806 – 815.]

Key Words: Hand, Hand force, Hand injuries, Measurement error, Reliability of results.

Ton AR Schreuders, Marij E Roebroeck, Janine Goumans, Johan F van Nieuwenhuijzen, Theo H Stijnen, Henk J Stam

806 Physical Therapy . Volume 83 . Number 9 . September 2003

Re se

ar ch

Re po

rt �

�����������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

D ow

nloaded from https://academ

ic.oup.com /ptj/article/83/9/806/2805321 by guest on 19 February 2021

O nly data that have acceptable reliability and validity are valuable in the clinical decision- making process for determination of impaired function and comparison of surgi-

cal repair techniques to document progress during rehabilitation and to evaluate disability after injury. Such data will decrease the need for use of unsubstantiated opinions and will increase the physical therapist’s ability to obtain reproducible findings and meaningful results.1 Better reliability implies better precision of single mea- surements, which is a prerequisite for better tracking of changes in measurements in research and clinical practice.2

Generally accepted instruments in the evaluation of hand function measure grip and pinch force. Grip and pinch force measurements have been promoted as an important measure of outcome in, for example, the evaluation of peripheral nerve function.3–5 For a correct interpretation of these force measurements, we contend

information on the reliability (measurement error) of these measurements is required. The reliability of grip force measurements has been studied in patients with hand injuries6,7 and in people without impairments of the hand.8,9 It also has been studied in several other categories of patients, including women with nonspecific regional pain (NSRP)10 and patients with reflex sympa- thetic dystrophy (RSD)11 or epicondylitis,12 and in peo- ple after a stroke.13 The reliability of pinch force mea- surements has been studied in patients with RSD11 and repetitive strain injury (RSI).14

The studies of Spijkerman et al6 and Brown et al7 on the reliability of force measurements in patients with hand injuries have important drawbacks. The study by Spijk- erman et al included only 8 patients, and Brown et al, in their study of grip and pinch force measurements in patients with hand injuries, did not specify the measure- ment error. As in many studies, the intraclass correlation coefficient (ICC) was used as a measure of reliability. For

TAR Schreuders, PT, is Physical Therapist, Department of Rehabilitation Medicine, Hand Therapy Unit, Erasmus MC–University Medical Center Rotterdam, PO Box 2040, 3000 CA Rotterdam, the Netherlands ([email protected]). Address all correspondence to Mr Schreuders.

ME Roebroeck, PhD, is Assistant Professor, Department of Rehabilitation Medicine, Erasmus MC–University Medical Center Rotterdam.

J Goumans, MD, is Medical Student, Erasmus MC–University Medical Center Rotterdam.

JF van Nieuwenhuijzen, PT, is Research Therapist, Department of Rehabilitation Medicine, Erasmus MC–University Medical Center Rotterdam.

TH Stijnen, PhD, is Professor of Biostatistics, Department of Epidemiology and Biostatistics, Erasmus MC–University Medical Center Rotterdam.

HJ Stam, MD, PhD, is Professor of Rehabilitation Medicine and Chairman, Department of Rehabilitation Medicine, Erasmus MC–University Medical Center Rotterdam.

Mr Schreuders, Dr Roebroeck, and Dr Stam provided concept/idea/research design. Mr Schreuders and Dr Roebroeck provided writing. Dr Goumans and Mr van Nieuwenhuijzen provided data collection, and Mr Schreuders, Dr Roebroeck, Dr Goumans, and Dr Stijnen provided data analysis. Mr Schreuders and Dr Goumans provided project management. Dr Roebroeck and Dr Stam provided fund procurement and facilities/equipment. Dr Stijnen and Dr Stam provided consultation (including review of manuscript before submission).

The Ethical Committee of the Erasmus Medical Center, the Netherlands, approved this study.

This article was received February 24, 2003, and was accepted May 19, 2003.

Physical Therapy . Volume 83 . Number 9 . September 2003 Schreuders et al . 807

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

��� ���

D ow

nloaded from https://academ

ic.oup.com /ptj/article/83/9/806/2805321 by guest on 19 February 2021

clinical use, however, we believe the most relevant information is the magnitude of change between tests that is required to detect a real change, preferably quantified in the same units as the force measure- ment.10,15 This magnitude of change is specified in the smallest detectable difference (SDD), which is calculated from the standard error of measurement (SEM), the absolute error component of the measurements.16,17 The smaller the measurement error, the better the measure.18

Various factors relating to the patient’s condition may influence the measurement error, such as the amount of pain and loss of normal function of the fingers or thumb after injury. The experience of the examiner also might influence the measurement error. However, studies on these aspects of grip and pinch force measurements for patients with hand injuries are scarce.

One aim of our study was to assess the SEM and related indexes of reliability for intraexaminer and interexam- iner applications of grip (2 handle positions, with dis- tances between the handles of 4.6 and 7.2 cm) and pinch (tip pinch and key pinch) force measurements for patients with hand injuries obtained with Lode handgrip and pinch-grip dynamometers (Fig. 1).* A second aim was to investigate whether the measurement error of grip and pinch force measurements differs between specific applications, such as measurements of injured and noninjured hands, measurements obtained by expe- rienced and inexperienced examiners, and measure- ments between 2 handle positions for grip force and 2 different types of pinch force (ie, tip and key pinch

force). A third aim was to judge whether the measure- ment error is small enough to justify use of these measurements to discern real changes in grip and pinch force in individual patients. Because no clear criteria to judge the SDD are available for grip and pinch force measurements, we compared our findings with the results of studies in which grip and pinch force measure- ments were examined in different patient categories (ie, patients with hand injuries, pain, RSD, or epicondy- litis and people after stroke).

Method

Subjects A consecutive sample of 33 patients (20 male, 13 female) who were seen in the Department of Rehabilitation Medicine of Erasmus MC–University Medical Center Rotterdam in the Netherlands participated in the study. They had a mean age of 36 years (SD 13.7, range�17– 67). All patients had an injury on only one hand (6 with tendon lacerations, 9 with nerve lesions, 4 with fractures, and 14 with a combination of lesions or surgeries [eg, crush injury, finger amputation, arthrodesis]). During a period of 5 months, patients who were capable of being evaluated for grip and pinch force were asked to partic- ipate in this study. Patients who were not permitted to grip or pinch with maximum force because of injured tissue and patients who complained of pain were excluded. We presumed that pain would inhibit the patients from exerting full maximum force and that force measurements may reflect the amount of pain rather than the force-generating capacity of the patients. The participants were informed about the purpose of the study and gave informed consent.

The participants were in different phases of their reha- bilitation process. Study entry was, on average, 22 months after injury or surgery when including one patient who had a 16-year period between the date of injury and time of examination. Excluding this patient, the average period was 9 months (SD�7.0, range 1–28) since injury or surgery. Force measurements of the hand are routinely obtained in our department during differ- ent phases of the rehabilitation process. Such measure- ments can be used to determine progression of muscle force recovery, the ability to perform in a vocation, and indications for continuing therapy and to assess long- term outcome after hand injuries.

Instrumentation We used Lode handgrip and pinch-grip dynamometers (Fig. 1). The distances between the handles of the Lode handgrip dynamometer were adapted to create spaces between the handles comparable to those of the Jamar

* Lode Medical Technology, Zernikepark 16, 9747 AN Groningen, the Nether- lands. Distributed in the United States by ElectraMed Corp, G-5332 Hill-23 Dr, Flint, MI 48507.

Figure 1. Measurements were performed with the participant seated at a table on which the dynamometers were positioned.

808 . Schreuders et al Physical Therapy . Volume 83 . Number 9 . September 2003

D ow

nloaded from https://academ

ic.oup.com /ptj/article/83/9/806/2805321 by guest on 19 February 2021

hand dynamometer.† For our study, the grip handle positions 2 (distance between handles�4.6 cm) and 4 (distance between handles�7.2 cm), which we desig- nated “grip 2” and “grip 4,” respectively, were used with the handgrip dynamometer.

With the Lode pinch-grip dynamometer, which is similar in design to the Preston pinch dynamometer,†,19 2 types of pinch were measured: tip pinch and key pinch. Tip pinch force was measured with only the index finger on top and the other fingers flexed with the thumb below, and key pinch force was measured with the thumb on top and the radial side of the index finger below.

The grip handle is connected to an amplifier from which the values give a digital readout on a display. According to the manufacturer, the measuring range is 0 to 1,000 N, with an accuracy of 1% deviation for 2 to 500 N and 2% for 500 to 900 N. Calibration was done with sus- pended weights according to the method described by Mathiowetz et al.8 We believe the use of strain gauge technology is preferable to dynamometers, which are spring-based or use a hydraulic pressure system, because such dynamometers might produce erroneous data due to the wear and tear of metal, slow leaks, and hysteresis.20

Procedure Measurements were taken according to the recommen- dations of the American Society of Hand Therapists (ASHT).21 Participants were seated at a table on which the dynamometers were positioned (Fig. 1). The subjects were told to keep their elbow flexed without resting their arm or the grip handle of the dynamometer on the table. The digital display was not visible to the subjects. Corre- sponding to the ASHT recommendations for each mea- surement, the mean of 3 repetitions was recorded. Measurements were obtained of the left and right hands alternately, which is the method of testing used at our clinic. Testing both hands, we contend, enables compar- isons of the injured and noninjured hands, as proposed by Gaul.22

The side (injured or noninjured) at which measure- ments were started and the order of tests (grip 2, grip 4, tip pinch, or key pinch) were randomly selected. Both examiners obtained all the measurements twice for each subject; the order of examiners was randomly selected. During each measurement session, the examiners obtained 12 measurements (2 grip and 2 pinch measure- ments � 3 repetitions from one hand. Then the exam- iners were changed while the subject took a short break (2–3 minutes) and moved the arm and hand out of the testing position. In this way, we attempted to reflect practice with examiners independently testing subjects,

except there was a short time interval between examin- ers. Both examiners participated in 2 sessions per patient, which resulted in 48 measurements per hand. The total time required for testing, including the breaks, was 35 to 40 minutes per patient.

Examiners Because we were interested in the effect of experience of the examiner on measurement error, we selected 2 examiners with widely differing experience. The first examiner was a 54-year-old male physical therapist ( JFvN) with 20 years of experience in testing hand function. The second examiner ( JG) was a 22-year-old female medical student with no previous experience with these tests. During a 2-week period, the inexperi- enced examiner had 5 hours of training by an experi- enced physical therapist to become familiarized with the equipment and testing protocol. During the training, measurements were practiced on several subjects with- out hand injuries and some patients with hand injuries. We believe that the results obtained by this student are reasonably representative of those that could be obtained by an inexperienced examiner, regardless of whether this is a starting physician or physical therapist, because at this stage they have comparable (ie, limited) knowledge, attitudes, and skills concerning force measurements.

Data Analysis Statistical analyses were done to determine intraexam- iner reliability of data obtained in the 2 sessions per- formed by the same examiner both for the experienced examiner (Texp1 and Texp2) and the inexperienced examiner (Tinx 1 and Tinx 2) separately. Interexaminer reliability was calculated between the first sessions of the experienced examiner (Texp1) and the inexperienced examiner (Tinx1). Analyses of the measurements of the injured hands (n�33) and noninjured hands (n�33) were conducted separately.

An analysis of variance (ANOVA) was conducted with an SPSS/PC� program‡ to determine the multiple sources of measurement error. The variance attributed to differ- ences among participants was designated “Var (P).” The variance attributed to differences between sessions by the same examiner (Texp1 and Texp2, Tinx1 and Tinx2) was designated “Var (S).” The variance ascribed to the different examiners (Texp1 and Tinx1) was designated “Var (T).” Variance due to interaction among participants for intraexaminer sessions was des- ignated “Var (P�S),” and variance due to interaction among participants for interexaminer sessions was des- ignated “Var (P�T).” The interaction components (P�S) and (P�T) were confounded by the residual

† Sammons Preston/Rolyan, 4 Sammons Ct, Bolingbrook, IL 60440. ‡ SPSS Inc, 233 S Wacker Dr, Chicago, IL 60606.

Physical Therapy . Volume 83 . Number 9 . September 2003 Schreuders et al . 809

��� ���

��� ���

��� ���

��� ���

��� �

D ow

nloaded from https://academ

ic.oup.com /ptj/article/83/9/806/2805321 by guest on 19 February 2021

error. In the intraexaminer analyses, Var (S) and Var (P�S) (ie, the variance within participants) constituted the error variance. Correspondingly, for the interexam- iner analyses, the variance components Var (T) and Var (P�T) constituted the error variance.

From the error variance, the SEM was computed as its square root.17 Based on the SEM, the SDD with 95% confidence was calculated as 1.96 � �2 � SEM.23 This SDD can be applied in such a way that only differences between 2 consecutive measurements greater than the SDD can be interpreted with 95% certainty as real change in grip or pinch force. Intraclass correlation coefficients were computed as the ratio of variances among participants (ie, Var [P]) and the total variance.15

Comparing SEMs In order to test whether differences existed in error variances of different applications of grip and pinch force measurements, a test of equality of variances for paired samples was applied. In this analysis, scores between sessions were compared by calculating the correlation coefficient (Pearson r) between the sum and difference of the difference scores and testing whether this correlation differed from zero.16,24(pp171–172) In this way, we tested whether differences existed in error variance, and thus SEM and SDD, between the experi- enced and inexperienced examiners, between measure- ments of the injured and noninjured hands, and between the 2 handle positions of the grip force measure- ments and between the 2 pinch force measurements.

Commonly accepted criteria to judge whether the SEM and SDD of measurements are adequate for application of the measurements to individual patients do not exist. Therefore, we compared the SEMs found in our study with those from other studies in which grip and pinch force measurements were investigated. For an adequate comparison, We selected studies from the literature in

which values for SEM or SDD were reported and studies from which SEM and SDD can be estimated by using the formulas: SEM � SD� (1�ICC)15(p119),23 or SEM � SDdifference /�2,15(p120) where SD�standard deviation. These studies included similar groups of patients as in our study (ie, patients with hand injuries or patients from different groups such as patients with pain, RSD, epicondylitis, and stroke).

Results Descriptive values (mean and standard deviations) were calculated for all measurements (2 grip and 2 pinch measurements) for the injured and noninjured hands and for the 4 sessions conducted by the experienced and inexperienced examiners (Tab. 1). Evidently noninjured

Table 1. Grip and Pinch Force Measurements (in Newtons) Obtained by Experienced and Inexperienced Examiners for Both Hands of Participants With Hand Injury of One Hand (N�33)

Hand Testa

Experienced Examiner Inexperienced Examiner

Session 1 Session 2 Session 1 Session 2

X SD X SD X SD X SD

Injured Grip 2 250 123 253 125 241 122 245 117 Grip 4 231 110 231 111 231 113 223 109 Tip pinch 39 17 39 16 37 16 38 16 Key pinch 63 22 64 22 63 25 62 24

Noninjured Grip 2 370 116 364 115 345 115 359 118 Grip 4 326 110 323 108 325 112 317 109 Tip pinch 55 13 53 12 54 12 54 12 Key pinch 87 18 86 19 86 20 84 20

a Grip 2�measurements of grip force in the second handle position, grip 4�measurements of grip force in the fourth handle position.

Figure 2. Scatterplot of intraexaminer grip force measurements with the second handle position obtained in 2 sessions by an experienced examiner (Texp1 and Texp2). ‚ represents measurements of the injured hands (N�33); F represents measurements of the noninjured hands (N�33), which are the higher values.

810 . Schreuders et al Physical Therapy . Volume 83 . Number 9 . September 2003

D ow

nloaded from https://academ

ic.oup.com /ptj/article/83/9/806/2805321 by guest on 19 February 2021

hands have stronger grip and pinch forces. The mean grip force in position 4 was less than in position 2. Key pinch had a considerably greater mean force than tip pinch.

Two scatterplots of measurements obtained for both injured and noninjured hands (n�66) are presented in Figures 2 and 3. Figure 2 shows the measurements of grip force with the second handle position for the 2 sessions conducted by the experienced examiner (Texp1 and Texp2). Figure 3 shows the measurements of the tip pinch force obtained by the 2 different examiners (Texp1 and Tinx1). Both scatterplots demon- strate that there was a high level of correspondence between the 2 sessions and the 2 examiners.

Table 2 contains the results of the ANOVA: the sum of squares, mean of squares, and estimated variance com- ponents attributed to different sources. Reliability indexes (ie, SEM, SDD, and ICC) for the intraexaminer grip and pinch measurements are presented, ordered by examiner (experienced and inexperienced) and by hand (injured and noninjured). The SEMs ranged from 4 to 29 N, and the corresponding SDDs ranged from 11 to 80 N. The lowest ICC was .82 and the highest ICCs were .97 which, according to the scale suggested by Shrout and Fleiss,25 should be classified as excellent. The interexaminer data are presented in Table 3 and show comparable values.

No differences were found in the error variances of grip and pinch force measurements between injured and noninjured hands or performed by experienced and

inexperienced examiners. Similarly, no differences were found in the error variances of the measurements between the 2 handle positions of grip force or between the 2 pinch techniques.

Tables 4 and 5 present comparisons among studies investigating grip and pinch force measurements, including the type of dynamometer used, characteristics of the subjects, means and standard deviations of force measurements, and reliability indexes (ie, SEM, SDD, and ICC). The measurement errors of the SEMs of grip force measurements are comparable in most of the studies, except for those of Nitschke et al10 and Smidt et al,12 in which relatively low values were found (Tab. 4). For pinch force measurements, the SEMs were 4 or 5 N, with the exception of the higher values of 11 N found by Geertzen et al11 (Tab. 5).

Discussion In our study, high ICCs were found for the reliability of grip and pinch force measurements in subjects with hand injuries. In our view, however, a high ICC should not be interpreted as a small measurement error because, in an ICC, measurement error and real variabil- ity between subjects are expressed in relative terms. Expressing the reliability as a dimensionless ratio of variances does not allow us to interpret the reliability in terms of an individual score. To decide whether a person’s grip force has changed after a period of reha- bilitation, a physical therapist must know which part of the change measured was real and which part by is due to measurement error. To accomplish this, we calculated the SEM and SDD.

In our study, the SDD of the grip force measurements (second handle position) of the injured hands was 61 N for measurements obtained by the experienced exam- iner (Tab. 2). This finding means that for grip force measurements taken by the experienced examiner (Texp), a change in force of at least 61 N is needed between 2 sessions to be 95% confident that a real change has occurred. The 2 different types of pinch force measurements (tip pinch and key pinch) showed a much smaller SDD; for the experienced examiner, the SDDs for the injured hands were 11 to 12 N (Tab. 2).

In the case of our 2 examiners with widely differing experience, comparison of the SEMs showed no differ- ences between them. Thus, for grip and pinch force measurements, it appears that a brief training period was sufficient for an inexperienced examiner to reach a level of reliability comparable to that of an experienced examiner. Given the small number of examiners in our study, however, this cannot be stated with certainty.

Figure 3. Scatterplot of interexaminer tip pinch force measurements by an expe- rienced examiner (Texp1) and an inexperienced examiner (Tinx1). ‚ represents measurements of the injured hands (N�33); F represents measurements of the noninjured hands (N�33).

Physical Therapy . Volume 83 . Number 9 . September 2003 Schreuders et al . 811

��� ���

��� ���

��� ���

��� ���

��� �

D ow

nloaded from https://academ

ic.oup.com /ptj/article/83/9/806/2805321 by guest on 19 February 2021

More examiners, however, would entail a longer mea- surement time, which might fatigue the patient or decrease his or her concentration, both of which may influence the measurement error. We attempted to compensate for the small number of examiners by selecting 2 examiners with what we considered maxi- mally different experience. Given that no differences in measurement error between the 2 examiners were found, we contend that having more examiners (eg, with intermediate levels of experience) obtain the measure- ments would not have affected our conclusions. Although we do not have data to support this view, we believe it is a reasonable interpretation.

Other factors also may influence the measurement error. For example, we also examined the difference in SEMs between grip and pinch force measurements of the injured and noninjured hands. The measurement error, however, did not differ between the injured and

noninjured hands. No differences were found between the SEMs of grip force measurements in the 2 handle positions or between both types of pinch force measure- ments. Therefore, for applications in clinical practice, we contend one SEM value can be applied for grip force measurements for both handle positions and one SEM value can be applied for pinch force measurements, irrespective of injury and experience of the examiner.

Accepted criteria to judge the SDD of grip and pinch force measurements do not exist. Therefore, we com- pared our findings with those of 6 studies in which grip and pinch force measurements were investigated in similar and different patient categories (ie, patients with hand injuries,6,7 women with NSRP,10 patients with RSD,11 patients with epicondylitis,12 and patients after a stroke23) (Tab. 4). The SEMs of grip force measure- ments in these different studies are comparable. How- ever, the differences in means and ranges of the mea-

Table 2. Intraexaminer Measurements of Grip and Pinch Force (in Newtons) of Injured Hands of Participants (N�33) Obtained by Experienced and Inexperienced Examinersa

Examiner Hand Test Source SS MS Estimated Variance SEM SDD ICC

Experienced Injured Grip 2 Between patients 972270 30383 14947 Within patients 15841 490 490 22 61 .97

Grip 4 Between patients 758462 23702 11534 Within patients 20257 633 633 25 70 .95

Tip pinch Between patients 16035 501 241 Within patients 603 19 18 4 12 .93

Key pinch Between patients 30373 949 483 Within patients 510 16 16 4 11 .97

Inexperienced Injured Grip 2 Between patients 899327 28104 13849 Within patients 12989 406 402 20 56 .97

Grip 4 Between patients 772010 24125 11868 Within patients 12443 389 406 20 56 .97

Tip pinch Between patients 15234 476 228 Within patients 657 21 21 5 13 .92

Key pinch Between patients 36315 1135 547 (n�32) Within patients 1316 41 41 6 18 .93

Experienced Noninjured Grip 2 Between patients 835322 26104 12705 Within patients 22198 694 694 26 73 .95

Grip 4 Between patients 738201 23069 11114 Within patients 26918 841 841 29 80 .93

Tip pinch Between patients 9482 296 140 Within patients 539 17 18 4 12 .89

Key pinch Between patients 21601 675 326 Within patients 754 24 24 5 13 .93

Inexperienced Noninjured Grip 2 Between patients 822792 26542 12922 (n�32) Within patients 21624 698 698 28 77 .94 Grip 4 Between patients 769145 24036 11823

Within patients 13478 408 408 20 56 .97 Tip pinch Between patients 8858 277 124

Within patients 892 28 28 5 15 .82 Key pinch Between patients 23848 745 351

Within patients 1383 43 43 7 18 .89

a Grip 2�measurements of grip force in the second handle position, grip 4�measurements of grip force in the fourth handle position. SS�sum of squares, MS�mean squares, SEM�standard error of measurement, SDD�smallest detectable difference, ICC�intraclass correlation coefficient.

812 . Schreuders et al Physical Therapy . Volume 83 . Number 9 . September 2003

D ow

nloaded from https://academ

ic.oup.com /ptj/article/83/9/806/2805321 by guest on 19 February 2021

Table 4. Comparison of Intratester Measurements From 6 Grip Force Studies With Data From Our Studya

Study Handheld Dynamometer Subjects n X SD SEM SDD ICC

Spijkerman et al6 Prototype Lode No injuries 16 231 18 14 39 .98 Hand injuries 8 434 10 25 59 .98

Brown et al7 Dexter gripb Hand injuries 30 267 115 30 84 .93

Nitschke et al10 Jamar Women with no injuries

32 325 69 20 57 .93

Women with NSRP

10 174 62 18 59 .95

Geertzen et al11 CITEC gripc RSD Unaffected hand 29 123 66 24 66 .97 Affected hand 84 51 25 71 .94

Smidt et al12 Jamar Epicondylitis Uninvolved hand 50 340 100 13 37e .98 Involved hand 300 110 14 39e .98

Boissy et al13 Lafayette gripd Not affected 10 382 87 33 92 .86 Stroke 15 130 84 25 69 .91

Our study Lode grip handle position 2

Hand injuries Noninjured hand 33 367 115 26 73 .95 Injured hand 252 123 22 61 .97

Lode grip handle position 4

Noninjured hand 324 109 29 80 .93 Injured hand 231 109 25 70 .95

a NSRP�nonspecific regional pain syndrome, RSD�reflex sympathetic dystrophy, SEM�standard error of measurement, SDD�smallest detectable difference, ICC�intraclass correlation coefficient (except for study by Nitschke et al, in which Pearson correlation coefficients were calculated). All data converted to newtons (data in studies by Nitschke et al, Geertzen et al, and Smidt et al were in kilograms of force and were converted to newtons by multiplying by 10; Brown and colleagues’ within-patient data were in pounds and were multiplied by 4.448). Data are for intratester measurements, except for Geertzen et al and Smidt et al, who reported intertester data. Data from Brown et al are the mean of 3 testers. Figures in italics are estimated from ICC values presented in the studies. b Dexter Extremity Evaluation & Therapy Systems, Cedaron Medical Inc, PO Box 2100, Davis, CA 95617. c CITEC hand-held dynamometer, CIT Technics, Rijksstraatweg 384, 9752 CR Haren, the Netherlands. d Lafayette Instrument Co, 3700 Sagamore Pkwy N, PO Box 5729, Lafayette, IN 47903. e Figures from study by Smidt et al are estimated from the SDdiff values they reported. These values of the SDD differ from the values reported by Smidt et al, who submitted an erratum on this point.

Table 3. Interexaminer Measurements of Grip and Pinch Force of Injured and Noninjured Hands of Participants (N�33)a

Hand Test Source SS MS Estimated Variance

SEM (N)

SDD (N) ICC

Injured Grip 2 Between patients 948096 29628 14584 Within patients 16279 493 493 22 62 .97

Grip 4 Between patients 763813 23869 11508 Within patients 27279 852 852 29 81 .93

Tip pinch Between patients 15805 494 234 Within patients 855 28 29 5 15 .89

Key pinch Between patients 33541 1048 509 (n�32) Within patients 955 30 30 5 15 .94

Noninjured Grip 2 Between patients 835379 26106 12899 Within patients 35320 1104 1111 33 92 .92

Grip 4 Between patients 774343 24198 11857 Within patients 15495 484 484 22 61 .96

Tip pinch Between patients 9579 299 136 Within patients 857 27 27 5 14 .84

Key pinch Between patients 21771 680 315 Within patients 1594 49 50 7 20 .86

a Grip 2�measurements of grip force in the second handle position, grip 4�measurements of grip force in the fourth handle position. SS�sum of squares, MS�mean squares, SEM�standard error of measurement, SDD�smallest detectable difference, ICC�intraclass correlation coefficient, N�newtons.

Physical Therapy . Volume 83 . Number 9 . September 2003 Schreuders et al . 813

��� ���

��� ���

��� ���

��� ���

��� �

D ow

nloaded from https://academ

ic.oup.com /ptj/article/83/9/806/2805321 by guest on 19 February 2021

sured force levels in these studies are remarkable (eg, the mean grip force levels in the affected hands ranged from 84 to 300 N). To determine the value of a partic- ular measurement in a population during clinical prac- tice, we believe it is essential to know in which range the changes of muscle force take place. For example, in people with hand injury, the grip force will be lower shortly after the trauma. During later phases of rehabil- itation, the increased muscle force will generally reach a plateau after many months. If the difference in the grip force between the start and end of the rehabilitation is, for example, 100 N, an SDD of 80 N will mean the measurement is inadequate to detect changes. If the range in which the forces change is as wide as 300 N, an SDD of 30 N is certainly sufficient to detect changes during the rehabilitation process. In this latter example, a virtual scale of 10 steps is achieved.

To determine whether the calculated measurement error is small enough to make a test valuable in clinical practice, different methods have been applied in the studies we cite. Geertzen et al11 concluded that the SDD is too large in relation to the mean force measured to be useful in patients with RSD. Smidt et al12 suggested that an SDD less than 10% of the total range of the measure- ments would be acceptable. The range, however, is determined by the extreme values of the measurement, whereas the standard deviation is less influenced by the extreme values. In our study, therefore, we re-examined the SDD in relation to the standard deviation of the measurements by calculating the SDD/standard devia- tion ratios. The SDD/standard deviation ratios were 0.5 to 0.6 for grip force measurements using the second handle position and 0.6 to 0.7 when using the fourth handle position. These findings mean that 6 to 7 steps of changes in force level can be detected within the 95% distribution (“virtual scale”) of impaired to normal levels of grip force. In our opinion, this SDD/standard devia- tion ratio is useful in order to gain better insight as to whether measurements can be used to detect clinical

meaningful changes. A more definitive conclusion con- cerning the usefulness of the measurements can be made by relating the SDD relative to the magnitude of changes in force measurements in a prospective study of patients throughout the entire rehabilitation process.

Conclusions According to the ICC values obtained in our study, the reliability of grip and pinch force measurements is excellent. However, assessment of the measurement error and detectable change in muscle force between 2 consecutive measurements demonstrated that, in grip and pinch force measurements in individual patients, only relatively large changes can be adequately detected.

Measurement error did not differ between the experi- enced and inexperienced examiners or the injured and noninjured hands. Similarly, no differences were found in the grip force measurements for the 2 handle posi- tions or in the tip pinch and key pinch force measure- ments. The limited number of examiners and the use of subjects who were pain-free should be considered when applying our results. Further study of the SDD/standard deviation ratio is needed to develop clear criteria to judge whether measurement error is acceptable to detect changes in individual patients.

References 1 Bear-Lehman J, Abreu BC. Evaluating the hand: issues in reliability and validity. Phys Ther. 1989;69:1025–1033.

2 Beckerman H, Roebroeck ME, Lankhorst GJ, et al. Smallest real difference: a link between reproducibility and responsiveness. Qual Life Res. 2001;10:571–578.

3 Dellon AL. A numerical grading scale for peripheral nerve function. J Hand Ther. 1993;6:152–160.

4 Rosen B, Dahlin LB, Lundborg G. Assessment of functional outcome after nerve repair in a longitudinal cohort. Scand J Plast Reconstr Surg Hand Surg. 2000;34:71–78.

Table 5. Comparison of Data From Earlier Pinch Force Measurement Studies With Data From Our Study (All Data in Newtons)a

Study Handheld Dynamometer Subjects n Tester X SD SEM SDD ICC

Geertzen et al11 CITEC RSD Unaffected side Key pinch 29 67 19 11 30 . . .b

Affected side Intertester 56 23 11 31 . . .b

Brown et al7 Dexter Hand injuries

Tip pinch 30 Intratester (mean of 48 18.2 4 7.1 .96 Jaw pinch 3 testers) 64 22.7 4 8.5 .96

Our study Lode Hand injuries

Noninjured hand Tip pinch 33 Intratester 53 11.7 4 12 .89 Injured hand 39 15.6 4 12 .93

Lode Noninjured hand Key pinch 86 19 5 14 .93 Injured hand 64 22.4 4 11 .97

a SEM�standard error of measurement, SDD�smallest detectable difference, ICC�intraclass correlation coefficient, RSD�reflex sympathetic dystrophy. b No data available.

814 . Schreuders et al Physical Therapy . Volume 83 . Number 9 . September 2003

D ow

nloaded from https://academ

ic.oup.com /ptj/article/83/9/806/2805321 by guest on 19 February 2021

5 Strickland JW, Idler, RS, DelSignore JL. Ulnar nerve repair. In: Gelberman RH, ed. Operative Nerve Repair and Reconstruction. Vol 1. Philadelphia, Pa: JB Lippincott Co; 1991;425– 436.

6 Spijkerman DC, Snijders CJ, Stijnen T, Lankhorst GJ. Standardiza- tion of grip strength measurements: effects on repeatability and peak force. Scand J Rehabil Med. 1991;23:203–206.

7 Brown A, Cramer LD, Eckhaus D, et al. Validity and reliability of the Dexter hand evaluation and therapy system in hand-injured patients. J Hand Ther. 2000;13:37– 45.

8 Mathiowetz V, Weber K, Volland G, Kashman N. Reliability and validity of grip and pinch strength evaluations. J Hand Surg [Am]. 1984;9:222–226.

9 Hamilton A, Balnave R, Adams R. Grip strength testing reliability. J Hand Ther. 1994;7:163–170.

10 Nitschke JE, McMeeken JM, Burry HC, Matyas TA. When is a change a genuine change? A clinically meaningful interpretation of grip strength measurements in healthy and disabled women. J Hand Ther. 1999;12:25–30.

11 Geertzen JH, Dijkstra PU, Stewart RE, et al. Variation in measure- ments of grip strength: a study in reflex sympathetic dystrophy patients. Acta Orthop Scand Suppl. 1998;279:4 –11.

12 Smidt N, van der Windt DA, Assendelft WJ, et al. Interobserver reproducibility of the assessment of severity of complaints, grip strength, and pressure pain threshold in patients with lateral epicon- dylitis. Arch Phys Med Rehabil. 2002;83:1145–50.

13 Boissy P, Bourbonnais D, Carlotti MM, et al. Maximal grip force in chronic stroke subjects and its relationship to global upper extremity function. Clin Rehabil. 1999;13:354 –362.

14 MacDermid JC, Kramer JF, Woodbury MG, et al. Interrater reliabil- ity of pinch and grip strength measurements in patients with cumula- tive trauma disorders. J Hand Ther. 1994;7:10 –14.

15 Streiner DL, Norman GR. Health Measurement Scales: A Practical Guide to Their Development and Use. 2nd ed. Oxford, England: Oxford Univer- sity Press; 1995.

16 Stratford PW, Goldsmith CH. Use of the standard error as a reliability index of interest: an applied example using elbow flexor strength data. Phys Ther. 1997;77:745–750.

17 Roebroeck ME, Harlaar J, Lankhorst GJ. The application of gener- alizability theory to reliability assessment: an illustration using isomet- ric force measurements. Phys Ther. 1993;73:386 –395; discussion 396 – 401.

18 Hopkins WG. Measures of reliability in sports medicine and science. Sports Med. 2000;30:1–15.

19 Young VL, Pin P, Kraemer BA, et al. Fluctuation in grip and pinch strength among normal subjects . J Hand Surg [Am]. 1989;14:125–129.

20 Pronk CNA. Technical note: measuring hand-grip force, using a new application of strain gauges. Med Biol Eng Comput. 1981;19: 127–128.

21 Clinical Assessment Recommendations. 2nd ed. Garner, NC: American Society of Hand Therapists; 1981.

22 Gaul JS Jr. Intrinsic motor recovery: a long-term study of ulnar nerve repair. J Hand Surg [Am]. 1982;7:502–508.

23 Roebroeck ME, Harlaar J, Lankhorst GJ. Reliability assessment of isometric knee extension measurements with a computer-assisted hand-held dynamometer. Arch Phys Med Rehabil. 1998;79:442– 448.

24 Armitage P, Berry G. Statistical Methods in Medical Research. 3rd ed. Boston: Mass: Blackwell Scientific Publications; 1994.

25 Shrout P, Fleiss J. Intraclass correlation: uses in assessing rater reliability. Psychol Bull. 1979;86:420 – 428.

Physical Therapy . Volume 83 . Number 9 . September 2003 Schreuders et al . 815

��� ���

��� ���

��� ���

��� ���

��� �