Lab Principles Report
Physiology Lessons
for use with the
Biopac Student Lab
PC under Windows
98SE, Me, 2000 Pro
or Macintosh
8.6 – 9.1
Lesson 9
GALVANIC SKIN RESPONSE
&
THE POLYGRAPH
Manual Revision
PL3.6.7-ML3.0.7/081103
Richard Pflanzer, Ph.D. Associate Professor
Indiana University School of Medicine
Purdue University School of Science
William McMullen Vice President
BIOPAC Systems, Inc.
BIOPAC Systems, Inc. 42 Aero Camino, Goleta, CA 93117
(805) 685-0066, Fax (805) 685-0067
Email: [email protected]
Web Site: http://www.biopac.com
Page 2 Lesson 9: GSR & Polygraph BIOPAC Student Lab
I. INTRODUCTION
Electricity flows through an electrical circuit because of a difference in electrical pressure
between the beginning and the end of a circuit. Electrical pressure or electromotive force (E)
is measured in volts (V). The flow of electricity, called current (I), is measured in amperes
(A) or amps for short. As electricity flows through the circuit, resistance to flow occurs.
Electrical resistance (R) is measured in ohms ().
In a simple circuit of direct electrical current, the relationship between the electromotive
force causing the electrical current, the resistance to flow of electricity, and the resultant
magnitude of the current is described by Ohm‘s Law.
Ohm’s Law: I (Amps) = E (Volts) / R (Ohms)
If two of the three variables are known, the unknown third variable can be calculated.
For example, if voltage and resistance values for a simple circuit are known,
the above formula can be used to calculate the value for current; if the values
for current and resistance are known, then the formula for computing voltage
is E = IR.
Ohm‘s Law implies that if a constant current is applied across a resistance, changes in the
resistance will produce a voltage change directly proportional to the resistance change.
For example, if a constant current of 1.0 ampere is applied across a resistance
of 2.0 ohms, the measured voltage would be 2.0 volts (I = E/R, 1.0 ampere =
2.0 volts/2.0 ohms). If the resistance dropped to 0.5 ohm, the voltage would
also fall to 0.5 volt (I = E/R, 1.0 amperes = 0.5 volt/0.5 ohm).
In this lesson, you will apply principles of Ohm‘s Law and record changes in the electrical
resistance of the skin.
The human skin displays several forms of bioelectric phenomena, especially in areas of the
extremities such as the fingers, palms of the hands, and soles of the feet.
Galvanic skin resistance (GSR) — When a feeble electric current is steadily applied
between two electrodes placed about an inch apart, the recorded electrical resistance
between them, referred to as the galvanic skin resistance (GSR), varies in accordance
with the emotional state of the subject.
Galvanic skin potential (GSP) — Similarly, if the electrodes are connected to a
suitable voltage amplifier, but without any externally applied current, the voltage
measured between them, referred to as the galvanic skin potential (GSP), varies with
the emotional state of the subject.
The combined changes in the GSR and GSP related to the emotion of the subject constitute
the galvanic skin response.
Biopac Student Lab Lesson 9: GSR & Polygraph Page 3
The physiological basis of the galvanic skin response is a change in autonomic tone, largely
sympathetic, occurring in the skin and subcutaneous tissue in response to a change in the
affective state of the subject. Changes in peripheral autonomic tone alter sweating and
cutaneous blood flow, which in turn change GSR and GSP.
For example, if a painful stimulus such as a pinprick is applied to the skin in
an area distant to the electrode, the stimulus will reflexively elicit a general
phasic sympathetic discharge to sweat glands, increasing secretion. The
increase in sweat, although generally small, lowers the electrical resistance of
the skin because sweat contains water and electrolytes, both of which
increase electrical conductivity of the skin.
As in the case of somatic sensory stimuli (e.g., pain, pressure, touch), changes in emotion
elicit changes in peripheral autonomic tone and hence the galvanic skin response. A common
example is the vasodilation of cutaneous blood vessels of the face (blushing) and increased
sweating that often occur in the emotional state of embarrassment.
The detection and recording of the galvanic skin response is often combined with the
detection and recording of other autonomic-dependent psychophysiological variables such as
heart rate, respiratory rate, and blood pressure. The device that detects and records these
variables is called a polygraph. Although many people think polygraph is synonymous with
lie detector, the literal meaning is ―many measures‖ (poly - many, graph – write). This lesson
is a polygraph in the true sense of the word since it uses three types of measures: (a) GSR, (b)
respiration, and (c) heart rate.
One of the underlying principles involved in using the polygraph as a lie detector is that
autonomic nervous system control of heart rate, respiratory rate, blood pressure and flow, and
sweating cannot consciously be altered. Another principle is that changes in emotion
associated with intentional falsification of answers to carefully selected and worded questions
involuntarily and subconsciously alters autonomic output in such a way as to cause
recognizable changes in recorded physiological variables.
In the experiments that follow, you will record respiration, GSR, and heart rate under various
experimental procedures so as to gain a better understanding of polygraphy, its applications,
and its limitations.
It is important to keep in mind that although the recording procedures and measures
used are similar to those that might be used in a real polygraph recording, this is not a ―lie
detector test.‖ All you will do here is record the Subject‘s physiological responses to
certain questions. Some types of physiological responses are typically associated with
―lying,‖ although even under the best conditions about one-third of innocent people ―fail‖
lie detector tests. The best you can hope for here is to get a better understanding of how
these types of procedures work.
Page 4 Lesson 9: GSR & Polygraph BIOPAC Student Lab
II. EXPERIMENTAL OBJECTIVES
1) To become familiar with procedures for recording the galvanic skin response.
2) To observe and record changes in respiratory rate, heart rate, and skin resistance
associated with somatic and special sensory stimuli.
3) To observe and record changes in respiratory rate, heart rate, and skin resistance
associated with cognitive behavior and emotion.
4) To analyze a 3-channel polygram recorded under various experimental conditions to gain
a better understanding of polygraphy and its potential for use and misuse.
III. MATERIALS
BIOPAC disposable vinyl electrodes (EL503) – 3 electrodes per Subject
BIOPAC Electrode lead set (SS2L)
BIOPAC Electrode gel (GEL1)
BIOPAC GSR transducer (SS3L)
BIOPAC Respiration transducer (SS5LB or older SS5LA or SS5L)
BIOPAC PAPER1 or nine sheets of different colored paper. Recommended: 8-1/2‖x 11‖
sheets of: white, black, green, red, blue, yellow, orange, brown, purple.
Computer system
Biopac Student Lab software v3.6.7 PC or v3.0.7 Mac or greater
BIOPAC acquisition unit (MP30)
BIOPAC wall transformer (AC100A)
BIOPAC serial cable (CBLSERA) or USB cable (USB1W) if using a USB port.
Biopac Student Lab Lesson 9: GSR & Polygraph Page 5
IV. EXPERIMENTAL METHODS
Overview
As you complete the Experimental Methods (Set Up, Calibration, and Recording) and the Analysis,
you may need to use the following tools and/or display options. The window display shown below is
only a reference sample — it does not represent any lesson specific data. The sample screen shows 3
channels of data and four channel measurement boxes, but your screen display may vary between
lessons and at different points within the same lesson.
channel measurement boxes (channel # measurement type result)
)
marker label
channel boxes (Data analysis mode only)
marker marker tools
selection tool
I-Beam cursor
zoom tool horizontal (time) scroll bar
horizontal scale
vertical scales
vertical (amplitude) scroll bar
channel labels
The symbols explained below are used throughout Experimental Methods and Analysis.
Key to Symbols
If you encounter a problem or need further explanation of a concept, refer to the
Orientation Chapter for more details.
The data collected in the step needs to be recorded in the Data Report (in the section
indicated by the alpha character). You can record the data individually by hand or choose
Edit > Journal > Paste measurements to paste the data to your journal for future
reference.
Most markers and labels are automatic. Markers appear at the top of the window as inverted triangles. This symbol is used to indicate that you need to insert a marker and key
in a marker label similar to the text in quotes. You can insert and label the marker during
or after acquisition. On a Mac, press ―ESC‖ and on a PC, press ―F9.‖
Each section is presented in a two-column format, as described below.
FAST TRACK STEPS
This side of the lesson (left, shaded
column) is the ―FAST TRACK‖ through
the lesson, which contains a basic
explanation of each step.
DETAILED EXPLANATION OF STEPS
This side of the lesson contains more detailed
information to clarify the steps and/or concepts in the
FAST TRACK, and may include reference diagrams,
illustrations, and screen shots.
Page 6 Lesson 9: GSR & Polygraph BIOPAC Student Lab
A. SET UP
FAST TRACK Set Up Detailed Explanation of Set Up Steps
1. Turn the computer ON. The desktop should appear on the monitor. If it does not
appear, ask the laboratory instructor for assistance.
2. Make sure the BIOPAC MP30 unit
is turned OFF.
3. Plug the transducers in as follows:
Respiration (SS5LB) — CH 1
Electrode lead set (SS2L) — CH 2
GSR (SS3L/SS3LA) — CH 3
BIOPAC MP30 unit
Respiratory Transducer
(SS5LB or SS5LA
or SS5L) plugs into
CHannel 1
Electrode lead set(SS2L)
plugs into CHannel 2
GSR (SS3L or SS3LA)
plugs into CHannel 3
Fig. 9.1 Equipment Connections
4. Turn the MP30 Data Acquisition
Unit ON.
5. Attach the respiratory transducer
(SS5LB) to the Subject (Fig. 9.2).
IMPORTANT USE NOTE
If using the SS5LA transducer,
you must be very careful to not
pull or yank on the rubber bow
tie portion that contains the
sensor element.
Set up continues…
Attach the respiratory transducer around the chest below
the armpits and above the nipples (Fig. 9.2). The correct
tension is critical. The respiratory transducer must be
slightly tight at the point of maximal expiration. The
respiration transducer can be applied over thin clothing,
such as a t-shirt.
Fig. 9.2 SS5LB Placement
Biopac Student Lab Lesson 9: GSR & Polygraph Page 7
If using the SS5LB or SS5L, simply attach the Velcro
ends together at the correct tension.
If using the SS5LA, attach the nylon belt by threading
the nylon strap through the corresponding slots on the
rubber bow tie such that the strap clamps into place when
tightened (Fig. 9.3)
Fig. 9.3
6. Fill both cavities of the GSR
transducer (SS3L/SS3LA) with gel
and attach to the Subject (Fig. 9.4).
IMPORTANT
You must fill both sensor cavities
with electrode gel (GEL1) before
placing on the fingers.
Set up continues…
You must fill the each cavity of the SS3L/SS3LA GSR
transducer with electrode gel to obtain accurate
recordings.
The SS3L and SS3LA attach to the fingertips in an
identical manner (Fig. 9.4) and should be in place for at
least five minutes prior to the start of recording.
Sensors attach to bottom of fingertips
Velcro straps wrap around
fingers
Fig. 9.4 SS3L/SS3LA attachment and connection
The SS3L/SS3LA is typically placed on the index and
middle finger of the left hand.
Page 8 Lesson 9: GSR & Polygraph BIOPAC Student Lab
Position the transducer so that the sensor is on the bottom
of your fingertip (the part without the fingernail) and
wrap the Velcro
tape around the finger so the transducer
fits snugly but not so tight that blood circulation is cut
off. It‘s a fine line between tight and too tight.
HINT For a good signal to be picked up, it helps if the subjects have a little sweat on their hands (not a lot, but enough so that their hands are not completely smooth or cold). If subjects wash their hands just prior to the recording or
if they have been sitting in a cold room, then they must do something to activate the sweat glands before beginning calibration or recording. If subjects begin with colder hands, the scale will be diminished and the signal will be easily saturated once they “warm up” during the lesson.
7. Place three electrodes on the Subject
(Fig. 9.5).
Set up continues…
Place three electrodes at the positions shown (Fig. 9.5).
one on right forearm (just above wrist)
one on inside right leg (just above ankle bone)
one on inside left leg (just above ankle bone)
Fig. 9.5
Place one electrode on the medial surface of the
right leg, just above the ankle bone
Place one electrode on the medial surface of the
left leg, just above the ankle bone
Place one electrode on the right anterior forearm
just above the wrist (same side of arm as the palm
of hand).
Note: For optimal electrode adhesion, the electrodes should
be placed on the skin at least 5 minutes before the start of
the Calibration procedure.
Biopac Student Lab Lesson 9: GSR & Polygraph Page 9
8. Attach the electrode lead set (SS2L)
to the electrodes (Fig. 9.6)
right forearm WHITE lead
right leg BLACK lead
(ground)
left leg RED lead
Fig. 9.6
Each of the pinch connectors on the end of the electrode
cable needs to be attached to a specific electrode. The
electrode cables are each a different color. Follow Fig.
9.6 to ensure that you connect each cable to the proper
electrode.
The pinch connectors work like a small clothespin, but
will only latch onto the nipple of the electrode from one
side of the connector.
When the electrode cable is connected properly, the
LEAD II electrode configuration will be established.
9. Start the BIOPAC Student Lab
Program.
10. Choose Lesson 9 (L09-Poly-1).
11. Type in your filename. Use a unique identifier.
12. Click OK.
END OF SET UP
This ends the Set Up procedure.
Page 10 Lesson 9: GSR & Polygraph BIOPAC Student Lab
B. CALIBRATION
The Calibration procedure establishes the hardware‘s internal parameters (such as gain, offset, and
scaling) and is critical for optimum performance. Pay close attention to the Calibration procedure.
FAST TRACK Calibration Detailed Explanation of Calibration Steps
1. Seat the Subject facing the
Director and away from screen.
Fig. 9.7
Ideally, the Subject should sit in a chair facing the
Director, with arms on the armrest and be in a relaxed
state, breathing normally.
The Subject should not be able to see the screen during
recording.
2. Click on Calibrate. The Calibrate button is in the upper left corner of the
Setup window.
3. Three seconds into the recording, a
beep will sound and Subject should
inhale and exhale deeply for one
cycle, then return to normal
breathing.
The program needs to see a change in the GSR recording
during calibration.
4. Wait for the Calibration to stop.
Calibration continues…
The Calibration will run for 10 seconds and then stop
automatically, so let it run its course.
Biopac Student Lab Lesson 9: GSR & Polygraph Page 11
5. Check your calibration data. At the end of the 10-sec calibration recording, your screen
should resemble Fig. 9.8.
Figure 9.8
If similar, proceed to the Data
Recording section.
All three recording channels should show some
fluctuation. There should be some variation 4-6 seconds
into the GSR recording from the deep inhale.
If different, Redo Calibration.
END OF CALIBRATION
If a channel does not show fluctuation, then it is possible
that a transducer is not connected properly or the Subject
did not inhale deeply enough, and you must redo
calibration by clicking on the Redo Calibration button
and repeating the entire calibration sequence.
Page 12 Lesson 9: GSR & Polygraph BIOPAC Student Lab
C. RECORDING LESSON DATA
FAST TRACK Recording Detailed Explanation of Recording Steps
1. Prepare for the recording. You will record three segments of data, 120 seconds each.
In order to work efficiently, read this entire section so you
will know what to do for each recording segment.
Check the last line of the journal and note the total amount
of time available for the recording. Stop each recording
segment as soon as possible so you don‘t use an excessive
amount of time (time is memory). Hints for obtaining optimal data:
a) The Subject must not be able to see the record as it is
being recorded.
b) The environment must be quiet.
c) Sensory input to the Subject must be kept at a
minimum since almost any change in the environment
may evoke a response.
d) Subject needs to be relaxed with arms resting on the
armrest.
e) The Subject should remain as still as possible during
recording.
f) Subject should answer question in a quiet tone with
minimal movement of the mouth.
g) Subject should be at his/her resting heart rate in a
relaxed mental and physical state, and should not have
performed any recent physical or mental exertion.
Segment 1
2. Take position as follows:
a) Subject faces Director.
b) Recorder faces screen and
listens for Director‘s
instructions to Subject.
Subject should be in a chair in a relaxed state, with arms on
the armrest, breathing normally.
Recorder will need to listen for Director‘s instructions to
Subject so s/he knows when to place markers.
3. Click on Record. The recording will begin.
4. Wait five seconds, then Director
instructs Subject to perform the
following, and pauses between
requests to reestablish a baseline.
Recording continues…
The 5-second wait establishes a baseline.
This segment should be completed within 120 seconds.
To insert Markers: Mac = Esc key, PC = F9 key. Markers can be entered or edited after the data is recorded.
Biopac Student Lab Lesson 9: GSR & Polygraph Page 13
Recorder inserts markers to indicate
event change.
a) Quietly say his/her name.
―name‖ b) Quietly count backward from 10.
―count from 10‖ c) Count backward from 30 by
subtracting increasing odd
numbers (e.g., 30, 29, 26, 21).
―count from 30‖ d) Director touches Subject on the
side of the face.
―face touched‖
It may be too difficult to type in the marker label text while
you are recording. The important thing is to get the marker
inserted during recording, at the precise moment the event
changes. Labels can be keyed after the recording is done.
Subtract 1 from 30, then 3 from 29, then 5 from 26 and so
on, by using increasing odd numbers as the subtracted
factor.
5. Click on Suspend. The recording should halt, giving you time to review the
data and prepare for the next recording segment.
6. Review the data on the screen.
If correct, go to Step 7.
If all went well, your data should look similar to Fig. 9.9
and you can proceed to Step 7.
Fig. 9.9
If incorrect, click on Redo. The data would be incorrect if:
a) the Suspend button was pressed prematurely.
b) A transducer slipped off or an electrode peeled up,
causing a large baseline drift, spike, or loss of
signal.
c) The markers were not inserted or were inserted at
the wrong time(s).
In this case, you should redo the recording by clicking on
―Redo‖ and repeating Steps 2-6. Note that once you press
Redo, the data you have just recorded will be erased.
Recording continues…
Page 14 Lesson 9: GSR & Polygraph BIOPAC Student Lab
Segment 2
7. Click on Resume. This segment should be completed within 120 seconds.
The recording will continue from the point where it last
stopped, and a marker labeled ―Concentration on colored
squares‖ will automatically come up when Resume is
pressed.
8. Director holds PAPER1 colored
squares about two feet from
Subject’s face.
Director instructs Subject to look at
and concentrate on each square for
about 10 seconds, and pauses
between requests to reestablish a
baseline.
Recorder inserts markers to indicate
color change and records the time
required for a stimulus to generate a
response.
Display the colors and insert
markers in the following order:
a) ―white‖
b) ―black‖
c) ―red‖
d) ―blue‖
e) ―green‖
f) ―yellow‖
g) ―orange‖
h) ―brown‖
i) ―purple‖
To insert Markers: Mac = Esc key, PC = F9 key. Markers
can also be entered or edited after the data is recorded.
9. Click on Suspend. The recording should halt, giving you time to review the
data and prepare for the next recording segment.
Recording continues…
Biopac Student Lab Lesson 9: GSR & Polygraph Page 15
10. Review the data on the screen.
If correct, and more segments are
required, go to Step 11.
If all went well, your data should look similar to Fig. 9.10
and you can proceed to Step 11.
Note: In the sample recording, only four colored sheets
were used.
Fig. 9.10
Variation in the data will completely depend on the Subject
and recording conditions.
If incorrect, click on Redo. The data would be incorrect for the reasons in Step 6.
If incorrect, you should redo the recording by clicking on
―Redo‖ and repeating Steps 7-10. Note that once you press
Redo, the data you have just recorded will be erased.
Segment 3
11. Click on Resume. This segment should be completed within 120 seconds.
The recording will continue from the point where it last
stopped, and a marker labeled ―series of ‗Yes/No‘
questions‖ will automatically come up when Resume is
pressed.
12. Director asks Subject the ten
questions below, and notes Subject‘s
response.
Subject responds ―yes‖ or ―no.‖
Recorder inserts a marker when the
question is asked and another marker
when Subject begins to answer.
―Q‖ when question asked
―A‖ when answer starts
Recording continues…
Each question-answer should take about 10 seconds.
Subject‘s replies should be limited to ―yes‖ or ―no.‖
Subject may answer truthfully or dishonestly.
To insert Markers: Mac = Esc key, PC = F9 key.
Markers can be edited after the data is recorded.
a) Are you currently a student? Y N Director should note Subject‘s responses here by circling Y
Page 16 Lesson 9: GSR & Polygraph BIOPAC Student Lab
b) Are your eyes blue? Y N
c) Do you have any brothers? Y N
d) Did you earn an ―A‖ on the last
physiology exam? Y N
e) Do you drive a motorcycle? Y N
f) Are you less than 25 years of
age? Y N
g) Have you ever traveled to
another planet? Y N
h) Have aliens from another planet
visited you? Y N
i) Do you watch ―Seinfeld‖? Y N
j) Have you answered all of the
preceding questions truthfully?
Y N
for yes and N for No
13. Click on Suspend. The recording should halt, allowing you to review the data.
14. Review the data on the screen.
If correct, go to Step 15.
If all went well, your data should look similar to Fig. 9.11
and you can proceed to Step 15.
Fig. 9.11
Variation in the data will completely depend on the Subject
and recording conditions.
If incorrect, click on Redo. The data would be incorrect for the reasons in Step 6.
If incorrect, you should redo the recording by clicking on
―Redo‖ and repeating Steps 11-14. Note that once you
press Redo, the data you have just recorded will be erased.
Recording continues…
15. Ask the Subject to answer the
questions honestly and note the
Flip to Table 9.3 in your Data Report and record the
Biopac Student Lab Lesson 9: GSR & Polygraph Page 17
truthful answer in the Data Report.
C
Subject‘s truthful answers to each question.
16. Click on Done. A pop-up window with four options will appear. Make your
choice, and continue as directed.
If choosing the ―Record from another Subject‖ option:
a) Attach the sensors per Set Up Steps 5, 6, and 7 and
continue the entire lesson from Set Up Step 10.
b) Each person will need to use a unique file name.
17. Remove the sensors.
END OF RECORDING
Page 18 Lesson 9: GSR & Polygraph BIOPAC Student Lab
V. DATA ANALYSIS
FAST TRACK Data Analysis Detailed Explanation of Data Analysis Steps
1. Enter the Review Saved Data mode
and choose the correct file.
Enter Review Saved Data from the Lessons menu.
Note Channel Number (CH)
designations:
Channel Displays
CH 3 GSR
CH 40 Respiration
CH 41 Heart Rate
Fig 9.12
2. Setup your display window for
optimal viewing of the first 5
seconds of the recording.
The following tools help you adjust the data window:
Autoscale horizontal Horizontal (Time) Scroll Bar
Autoscale waveforms Vertical (Amplitude) Scroll Bar
Zoom Tool Zoom Previous
3. Set up the measurement boxes as
follows:
Channel Measurement
CH 41 value
CH 40 BPM
CH 3 value
CH 3 none
Data Analysis continues…
The measurement boxes are above the marker region in the
data window. Each measurement has three sections:
channel number, measurement type, and value. The first
two sections are pull-down menus that are activated when
you click on them. The following is a brief description of
these specific measurements.
value: displays the amplitude value for the channel at the
selected point. If a single point is selected, the value is for
that point, if an area is selected, the value is the endpoint
of the selected area.
BPM: In this lesson, the BPM measurement stands for
Breaths Per Minute and calculates the difference in time
between the end and beginning of the selected area (same
as T), then divides this value into 60 seconds/minute.
none: turns off the measurement channel.
The ―selected area‖ is the area selected by the I-Beam tool
(including the endpoints).
Biopac Student Lab Lesson 9: GSR & Polygraph Page 19
4. Using the I-Beam cursor, choose a
point at the 2-second mark and
record the heart rate and GSR values
(Fig. 9.13).
A
Fig 9.13
5. Using the I-Beam cursor, select an
area from the start of one inhale to
the start of the next inhale (Fig.
9.14), and record the respiration rate
(BPM).
A
The respiration transducer records chest expansion
(inhalation) as positive values, and chest deflation
(exhalation) as negative values. Therefore, the start of
inhalation is recorded as the beginning of the ascending
positive waveform.
Note: This measurement may be difficult to perform,
depending on your data, because small dips in chest
expansion can occur within the normal cycle. You must
be able to distinguish the small dips from the big dips.
Fig 9.14
6. Scroll to view a 10-second interval
beginning at the first marker.
This 10-second interval should show the Subject‘s response
to the first instruction.
7. Find the point of maximal GSR
within this 10-second segment and
record the heart rate and GSR values
at this point.
A
Data Analysis continues…
Fig 9.15
Page 20 Lesson 9: GSR & Polygraph BIOPAC Student Lab
8. Using the I-Beam cursor, select an
area from the start of one inhale to
the start of the next inhale, closest to
the point used in Step 7 and record
the respiration rate (BPM).
A
The respiration transducer records chest expansion
(inhalation) as positive values, and chest deflation
(exhalation) as negative values. Therefore, the start of
inhalation is recorded as the beginning of the ascending
positive waveform.
Fig 9.16
9. Repeat Steps 5-7 for each condition
in Segment 1 of your data.
A
Each condition in the data record should be separated by a
marker (―‖).
10. Repeat Steps 7-8 for Segment 2
data.
B
11. Repeat Steps 7-8 for Segment 3
data, using a 5-second interval
beginning at the ―A‖ marker.
C
Measurements should be taken in the interval that begins
when the Subject started to answer.
12. Save or print the data file. You may save the data to a floppy drive, save notes that are
in the journal, or print the data file.
13. Exit the program.
END OF DATA ANALYSIS
END OF LESSON 9 Complete the Lesson 9 Data Report that follows.
Biopac Student Lab Lesson 9: GSR & Polygraph Page 21
Lesson 9
GALVANIC SKIN RESPONSE & THE POLYGRAPH
DATA REPORT
Student’s Name:
Lab Section:
Date:
Subject Profile
Name Height
Age Weight
Gender: Male / Female
I. Data and Calculations
A. Complete Table 9.1 with Segment 1 data.
Mark I for increase, D for decrease, and NC for no change relative to baseline.
Table 9.1 Segment 1 Data
Procedure Heart Rate Respiratory Rate GSR
[CH 41 Value] [CH 40 BPM] [CH 3 Value]
Resting (baseline)
Quietly say name
Count from 10
Count from 30
Face touched
Page 22 Lesson 9: GSR & Polygraph BIOPAC Student Lab
B. Complete Table 9.2 with Segment 2 data.
Mark I for increase, D for decrease, and NC for no change relative to baseline.
Table 9.2 Segment 2 Data
Square Color Heart Rate Respiratory Rate GSR
[CH 41 Value] [CH 40 BPM] [CH 3 Value]
white
black
red
blue
green
yellow
orange
brown
purple
C. Complete Table 9.3 with Segment 3 data.
Mark I for increase, D for decrease, and NC for no change relative to baseline.
Table 9.3 Segment 3 Data
Question Answer Truth Heart Rate Respiratory Rate GSR
[CH 41 Value] [CH 40 BPM] [CH 3 Value]
Student? Y N Y N
Blue eyes? Y N Y N
Brothers? Y N Y N
Earn “A”? Y N Y N
Motorcycle? Y N Y N
Less than 25? Y N Y N
Another planet? Y N Y N
Aliens visit? Y N Y N
“Seinfeld”? Y N Y N
Truthful? Y N Y N
Biopac Student Lab Lesson 9: GSR & Polygraph Page 23
II. Questions
D. Of what practical value is the GSR information obtained from the color experiment?
E. What major physiological changes account for the galvanic skin response?
F. Give three reasons why polygraph testing of a person‘s sincerity and honesty may yield inconclusive results.
End of Lesson 9 Data Report