TECHNICAL PROCEDURE

SOP# 313

Title:

The Alere BinaxNOW ® Influenza A & B Card Product Instructions

Purpose:

The BinaxNOW ® Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasopharyngeal (NP) swab and nasal wash/aspirate specimens. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. Negative test results should be confirmed by cell culture.

Scope:

This policy applies to Laboratories with a Certificate of Waiver must follow the manufacturer’s instructions for performing the test. Any modifications by the laboratory to the test system or FDA cleared test system instructions will result in the test no longer meeting the requirements for waived categorization.

Location:

Inpatient Ambulatory Clinics Emergency Services

tyy

Population:

All ages Adults Only Pediatrics

Personnel:

All clinically trained staff or MLS RNs / LPNs RTs

Definitions:

The Alere BinaxNOW ® Influenza A & B Card is a highly contagious, acute viral infection of the respiratory tract. It is a communicable disease easily transmitted through the coughing and sneezing of aerosolized droplets containing live virus.

Influenza outbreaks occur each year during the fall and winter months.

Rapid diagnosis of influenza A and B has become more important due to the availability try of effective antiviral therapy. Rapid diagnosis of influenza can lead to reduced hos- pital stays, antimicrobial use and cost of hospital care.

The Alere BinaxNOW ® Influenza A & B Card provides a simple, rapid method for the diagnosis of influenza A and B using NP swab, nasal swab and nasal wash/aspirate specimens. The easy-to-use format and rapid results allow for its use in “STAT” testing where it can provide information to assist with treatment and hospitalization decisions.

I. Specimen Requirements

A. Swab specimens require a sample preparation step, in which the sample is eluted or washed off the swab into an elution solution, saline or transport media. Nasal wash/ aspirate samples require no preparation. Sample is added to the top of the test strip and the test card is closed. Test results are interpreted at 15 minutes based on the presence or absence of pink-to-purple colored Sample Lines. The blue Control Line turns pink in a valid assay.

B. Specimen Collection and Handling:

1. Use fresh NP swabs and nasal wash/aspirates for best test performance.

2. Collect nasal washes in standard containers. Test as soon as possible. Washes can be held at 36-46°F (2-8°C) for up to 24 hours prior to testing in the BinaxNOW ® Test.

3. Use sterile cotton, rayon, foam or polyester flexible swabs to collect NP sample. Do not use calcium alginate swabs. Elute swab within one hour of collection. Test as soon as possible. Eluted swab samples can be held at 36-46 °F (2-8°C) for up to 24 hours before testing in the BinaxNOW ® Test. If needed, transport sample at 36-46 ° F (2-8°C) in a leak-proof container.

4. Allow samples to war m to room temperature before testing in the BinaxNOW ® Test. Swirl gently to mix (without creating bubbles) before testing.

5. Transport Media: The following transport media were tested and are acceptable for use in te BinaxNOW ® Test. Amies Media, Hank’s Balanced Salt Solution, M4 Media, M4-RT Media, M5 Media, and Stuart’s Media Saline.

C. Specimen Acceptability / Rejection Criteria

1. For in vitro diagnostic use.

2. Leave test sealed in its foil pouch until just before use.

3. Do not use kit past its expiration date.

4. Do not mix components from different kit lots.

5. The white sample pad at the top of the test strip contains reagents that extract the target antigen from the virus. To ensure best performance, add the sample SLOWLY (drop-by-drop) to the MIDDLE of this pad, without touching with the pipet, such that the entire sample absorbs into the pad.

6. Solutions used to make the control swabs are inactivated using standard methods. However, patient samples, controls, and tests should be handled as though they could transmit disease. Observe established precautions against microbial hazards. The use of lab coats, gloves, and safety eye glasses is recommended.

7. All transfer pipettes and test vials are single use items – do not use with more than one specimen.

8. The ability of this test to detect avian influenza was determined using cultured avian influenza viruses, the performance characteristics of this test with specimens collected from humans infected with H5N1 or other avian influenzas is unknown.

9. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive culture specimens.

10. Performance characteristics for influenza A were established when influenza A/H3 and A/H1 were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.

11. The elution solution packaged in this kit contains saline, detergents and preservatives that will inactivate cells and virus particles. Samples eluted in this solution are not suitable for culture.

II. Equipment and Materials:

A. Materials Provided:

1. Alere BinaxNOW ® Influenza A & B Card.

2. Test Cards: A cardboard, book-shaped, hinged, test card containing the test strip. A/Texas/1/77 was the master influenza virus strain used to develop the monoclonal antibodies incorporated into the test card to detect the influenza A virus.

3. Transfer Pipettes: Fixed volume (100 μl), transfer pipettes used to transfer sample to the test cards. Use only pipettes provided by Alere.

4. Positive Control Swab: Inactivated influenza A/Beijing, influenza A/Texas/1/77 (H3N2) or influenza A/T/W/66 (H9N2) virus and inactivated influenza B/Harbin or influenza B/Hong Kong 5/72 virus dried onto the swab. The influenza viruses are originally grown in embryonic eggs and are Formalin or gamma radiation or Beta propiolactone inactivated. Viruses are tested for inactivation and non-infectiousness by re-growing virus in embryonic eggs or by cytopathic effect (CPE) in culture. Viruses are considered inactivated when no viral propagation is seen in eggs or cells.

5. Negative Control Swab: Inactivated Streptococcus Group A dried onto swab. Organism used to inoculate the swab is heat inactivated, and then tested for inactivation and non-infectiousness by standard culture. The organisms are determined to be inactivated when no growth is present on the plate.

6. Elution Solution Vials for Control Swabs/Swab Specimens: Vials containing elution solution used to prepare the Control Swabs/Swab Specimens for testing.

7. NP Swabs: Sterile swabs for use in the Alere BinaxNOW ® Influenza A & B Card.

B. Material recommended but not provided:

Clock, timer or stopwatch; nasal wash/aspirate collection containers.

C. Storage and stability:

Store kit at room temperature (59-86°F, 15-30°C). The BinaxNOW ® Influenza A & B Test Kit and reagents are stable until the expiration dates marked on their outer packaging and containers.

D. Sample preparation procedure:

1. Nasal Wash/Aspirate: Nasal wash/aspirates do not need preparation.

2. Precaution: When testing nasal wash/aspirate samples, avoid thick areas of the sample when drawing it into the transfer pipette. If the pipette becomes clogged, and the lower part of the pipette is not full, put the sample back into container by squeezing the top bulb. Redraw the sample into the pipette. Use a new pipette if needed.

3. Nasopharyngeal Swabs: Remove sample from swab in 0.5 to 3.0 ml of saline or transport media/fluid by vigorously rotating the swab in the liquid. Go to Test Procedure.

4. Swab (Control & Patient) Elution using Binax Elution Solution:

a. Twist off the test vial cap.

b. Put the swab to be tested into test vial. Rotate the swab vigorously (without making a lot of bubbles) three (3) times in the liquid.

c. Press the swab against the side of the vial and turn as you remove it from the vial. This removes sample from the swab.

d. Discard the swab into a container intended for contagious material.

e. Test the liquid sample (from the test vial) in the BinaxNOW® Test as soon as possible.

III. Procedure

A. Influenza A and B test:

WARNING: INVALID RESULTS can occur when too little sample is added to the test. Be sure that the lower part of the transfer pipette is full and does not have any air spaces before you add the sample to the Sample Pad. If there are air spaces, put the sample back into the container by squeezing the top bulb. Redraw the sample from the bottom of the container into the pipette. Use a new pipette if needed.

1. Remove device from the pouch just prior to testing and lay flat on work bench.

2. Fill pipette by firmly squeezing the top bulb and then placing pipette tip into sample. Slowly release bulb while tip is still in sample. This will pull liquid into the pipette. Make sure there are no air spaces in the lower part of the pipette.

3. See arrow on test to find White Sample Pad. SLOWLY (dropby- drop) add entire contents of pipette (100 μl) to the MIDDLE of this pad by squeezing the top bulb.

4. Immediately peel off adhesive liner from the test device. Close and securely seal the device. Read result in window 15 minutes after closing the device. Note: When reading test results, tilt the device to reduce glare on the result window if necessary.

B. Reporting of results:

Result

Suggested Report

Positive for Flu A

Positive for Flu A protein antigen.

Positive for Flu B

Positive for Flu B protein antigen.

Positive for Flu A & B

Positive for both Flu A and B protein antigens.

Negative

Negative for Flu A and Flu B protein antigens. Infection due to Flu A and Flu B cannot be ruled out. Flu A and/or Flu B antigen in the sample may be below the detection limit of the test. Binax suggests culture of negative samples.

IV. Control Procedures

A. Quality Control:

Procedure Controls:

1. An untested strip has a blue line at the “Control” position. If the test flows correctly and the reagents work, this blue line will always turn pink on the strip.

2. The clearing of background color from the result window is a negative background control. The background color in the window should change from light pink to white within 15 minutes. Background color should not interfere with the reading of the test.

B. External Positive and Negative Controls:

1. Good laboratory practice suggests the use of positive and negative controls to ensure that:

• Test reagents are working, and

• the test is correctly performed.

2. Alere BinaxNOW® Influenza A & B Card kits contain Positive and Negative Control Swabs. These swabs will verify the entire assay. Test these swabs once with each new shipment received. Other controls may be tested in order to conform with:

• Local, state and/or federal regulations,

• accrediting groups, and/or,

• your lab’s standard QC procedures.

C. Linearity

Daily Quality Control: The BinaxNOW® Test has built-in (internal) procedural controls. For daily quality control, Binax suggests that you record results of these controls for each test run.

V. Maintenance

A. Routine Maintenance:

1. The cuvette holder should be cleaned after each day of use. Pull the cuvette holder out to the loading position. Clean the cuvette holder with alcohol’

2. The background color in the window should change within 15 minutes.

B. Wait 15 minutes before replacing the cuvette holder and using the analyzer.

VI. Limitations / Additional Considerations

A. A negative test result does not exclude infection with influenza A and/or B. Therefore, the results obtained with the Alere BinaxNOW ® Influenza A & B Card should be used in conjunction with clinical findings to make an accurate diagnosis. Additional testing is required to differentiate any specific influenza A and B subtypes or strains, in consultation with state or local public health departments.

B. The Alere BinaxNOW ® Influenza A & B Card detects both viable (live) and non-viable influenza A and B.Test performance depends on the amount of virus (antigen) in the specimen and may or may not compare with cell culture results performed on the same specimen.

C. Monoclonal antibodies may fail to detect, or detect with less sensitivity, influenza A and B viruses that have undergone minor amino acid changes in the target epitope region.

D. Inadequate specimen collection or improper sample handling/transport may yield a false-negative result.

E. Performance of the Alere BinaxNOW ® Influenza A & B Card has not been established for monitoring antiviral treatment of influenza.

F. Positive and negative predictive values of in vitro diagnostic tests are highly dependent on prevalence. False negative test results are more likely during peak activity when prevalence of disease is high. False positive test results are more likely during periods of low influenza activity when prevalence is moderate to low.

G. Individuals who have received nasally administered influenza A vaccine may test positive in commercially available influenza rapid diagnostic tests for up to three days after vaccination.

H. Children tend to shed virus more abundantly and for longer periods of time than adults.Therefore, in vitro diagnostic tests for influenza may have lower sensitivity in adults than in children.

I. Use of visibly bloody samples is not recommended with the Alere BinaxNOW ® Influenza A & B Card.

VII. Result Interpretation

A. Reference Intervals

Note: Do not read test results before or after 15 minutes as they may not be correct.

Sample

Result Interpretation

For a NEGATIVE SAMPLE,

The BLUE Control Line in the BOTTOM THIRD of the window turns a pink to purple color. No other line appears.

For a FLU A POSITIVE SAMPLE,

The BLUE Control Line turns a pink to purple color. A SECOND pink to purple Sample Line appears above it in the MIDDLE THIRD of the window. Any shade of a pink to purple Sample Line indicates a positive result.

For a FLU B POSITIVE SAMPLE,

The BLUE Control Line turns a pink to purple color. A SECOND pink to purple Sample Line appears above it in the TOP THIRD of the window. Any shade of a pink to purple Sample Line indicates a positive result.

For a FLU A and FLU B POSITIVE SAMPLE,

The BLUE Control. Line turns a pink to purple color, AND two pink to purple Sample. Lines appear above it in the MIDDLE and TOP thirds of the window. Any shade of pink to purple Sample Lines indicates positive results.

A test is INVALID,

if the Control Line remains BLUE or is not present

at all, whether a Sample Line(s) is present or not. Repeat

invalid tests. Call Binax if the problem continues.

B. Measurement Range

1. Analytical Sensitivity:

The Alere BinaxNOW ® Influenza A & B Card limit of detection (LOD), defined as the concentration of influenza virus that produces positive Alere BinaxNOW ® Influenza A & B Card results approximately 95% of the time, was identified by evaluating different concentrations of inactivated Flu A/Beijing and inactivated Flu B/Harbin in the Alere BinaxNOW ® Influenza A & B Card. Twelve (12) different operators each interpreted 2 tests run at each concentration for a total of 24 determinations per level. The following results identify a concentration of 1.03 x 10 2 ng/ml as the LOD for Flu A/Beijing and 6.05 x 10 1 ng/ml for Flu B/Harbin.

2. Analytical Reactivity:

The first seven (7) live influenza A strains and the five (5) live influenza B strains listed tested positive in the BinaxNOW® Influenza A & B Test at concentrations ranging from 102 - 106 CEID50/ml. The last two (2) influenza A strains listed below, A/Hong Kong/156/97 and A/Vietnam/1194/04, tested positive in the Binax test at concentrations of 1.3 x 102 and 1.0 x 104 TCID50/mL respectively. Although the specific influenza strains causing infection in humans can vary year to year, all contain the conserved nucleoproteins targeted by the BinaxNOW® test.

3. Analytical Specificity (Cross-Reactivity):

mensal and pathogenic microorganisms (27 bacteria, 8 viruses and 1 yeast) that may be present in the nasal cavity or nasopharynx were tested. All of the following microorganisms were negative when tested at concentrations ranging from 104 to 108 TCID50/ml (viruses), 107 to 108 organisms/ml (bacteria) and 106 organisms/ml (yeast).

Related Documents:

Appendix A: Influenza A Card

Appendix B: Influenza B Card

Effective Date:

04/24/2017

Scope:

Medical Director / Designee

Date

Your hospital

Your Medical director’s signature

Scope:

Patient Care Services

Date

Your hospital

Your signature as lab manager

References

1) Williams, KM, Jackson MA, Hamilton M. (2002) Rapid Diagnostic Testing for URIs in Children: Impact on Physician Decision Making and Cost. Infect. Med. 19(3): 109-111.

2) Dowdle, W.R, Kendal, A.P., and Noble, G.R. (1980). Influenza Virus, p 836-844. Manual of Clinical Microbiology, 3rd edition, in Lennette, et. al (ed.). American Society for Microbiology, Washington, D.C.

3) “Key Facts about Avian Influenza (Bird Flu) and Avian Influenza A (H5N1) Virus” CDC Publication, May 24, 2005. http://www.cdc.gov/flu/avian/gen-info/facts.htm

4) “Avian Influenza Infection in Humans” CDC Publication, May 24, 2005. http://www.cdc.gov/flu/avian/gen-info/avian-flu-humans.htm

Al Matar The Alere BinaxNOW ® Influenza A B Training Checklist .docx

The Alere BinaxNOW ® Influenza A & B

Skills Training

St. Luke’s Magic Valley Point of Care

Employee Name / Employee Number: _________________________________________________ Date: _________________________________ Location: __________________________________

Instructor’s Name: _________________________________________________________________

Training Points - Critical Skills

Introduction & Demonstration

(√ & Instructor’s Initials)

Employee Performance Observed

(√ & Instructor’s Initials)

Test Purpose: For the qualitative detection of influenza A and B nucleoprotein antigens in nasopharyngeal (NP) swab and nasal wash/aspirate specimens. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections.

Specimen Requirements / Storage: Swab specimens require a sample preparation step, in which the sample is eluted or washed off the swab into an elution solution, saline or transport media. Nasal wash/ aspirate samples require no preparation. Sample is added to the top of the test strip and the test card is closed. Test results are interpreted at 15 minutes based on the presence or absence of pink-to-purple colored Sample Lines. The blue Control Line turns pink in a valid assay.

Reagents / Supplies: Alere BinaxNOW ® Influenza A & B Card. Test Cards.

Transfer Pipettes. Positive Control Swab. Negative Control Swab. NP Swabs.

Test Procedure:

1. Remove device from the pouch just prior to testing and lay flat on work bench.

2. Fill pipette by firmly squeezing the top bulb and then placing pipette tip into sample. Slowly release bulb while tip is still in sample. This will pull liquid into the pipette. Make sure there are no air spaces in the lower part of the pipette.

3. See arrow on test to find White Sample Pad. SLOWLY (dropby- drop) add entire contents of pipette (100 μl) to the MIDDLE of this pad by squeezing the top bulb.

4. Immediately peel off adhesive liner from the test device. Close and securely seal the device. Read result in window 15 minutes after closing the device. Note: When reading test results, tilt the device to reduce glare on the result window if necessary.

Expected (Normal) Results:

Positive for Flu A: Positive for Flu A protein antigen.

Positive for Flu B: Positive for Flu B protein antigen.

Positive for Flu A & B: Positive for both Flu A and B protein antigens.

Negative: Negative for Flu A and Flu B protein antigens. Infection due to Flu A and Flu B cannot be ruled out. Flu A and/or Flu B antigen in the sample may be below the detection limit of the test. Binax suggests culture of negative samples.

Internal Quality Control: An untested strip has a blue line at the “Control” position. If the test flows correctly and the reagents work, this blue line will always turn pink on the strip.The clearing of background color from the result window is a negative background control. The background color in the window should change from light pink to white within 15 minutes. Background color should not interfere with the reading of the test.

Training Points - Critical Skills

Introduction & Demonstration

(√ & Instructor’s Initials)

Employee Performance Observed

(√ & Instructor’s Initials)

External Quality Control: Good laboratory practice suggests the use of positive and negative controls to ensure that:

• Test reagents are working, and

• the test is correctly performed.

Alere BinaxNOW® Influenza A & B Card kits contain Positive and Negative Control Swabs. These swabs will verify the entire assay. Test these swabs once with each new shipment received. Other controls may be tested in order to conform with:

• Local, state and/or federal regulations,

• accrediting groups, and/or,

• your lab’s standard QC procedures.

Instrument Cleaning: The cuvette holder should be cleaned after each day of use. Pull the cuvette holder out to the loading position. Clean the cuvette holder with alcohol’. The background color in the window should change within 15 minutes.

Wait 15 minutes before replacing the cuvette holder and using the analyzer.

I have read, observed, and been trained on the preceding skills sets.

Employee’s Signature and Date

All Common Checklist.pdf

Master

All Common Checklist

CAP Accreditation Program

College of American Pathologists 325 Waukegan Road Northfield, IL 60093-2750 www.cap.org 08.17.2016

2 of 52

All Common Checklist 08.17.2016

Disclaimer and Copyright Notice On-site inspections are performed with the edition of the Checklists mailed to a facility at the completion of the application or reapplication process, not necessarily those currently posted on the website. The checklists undergo regular revision and a new edition may be published after the inspection materials are sent.

For questions about the use of the Checklists or Checklist interpretation, email [email protected] or call 800-323-4040 or 847-832-7000 (international customers, use country code 001).

The Checklists used for inspection by the College of American Pathologists' Accreditation Programs have been created by the CAP and are copyrighted works of the CAP. The CAP has authorized copying and use of the checklists by CAP inspectors in conducting laboratory inspections for the Commission on Laboratory Accreditation and by laboratories that are preparing for such inspections. Except as permitted by section 107 of the Copyright Act, 17 U.S.C. sec. 107, any other use of the Checklists constitutes infringement of the CAP's copyrights in the Checklists. The CAP will take appropriate legal action to protect these copyrights.

All Checklists are ©2016. College of American Pathologists. All rights reserved.

3 of 52

All Common Checklist 08.17.2016

All Common Checklist

TABLE OF CONTENTS

SUMMARY OF CHANGES....................................................................................................................4 UNDERSTANDING THE CAP ACCREDITATION CHECKLIST COMPONENTS................................ 6 INTRODUCTION.................................................................................................................................... 6 DEFINITION OF TERMS....................................................................................................................... 6 ALL COMMON CHECKLIST............................................................................................................... 10

PROFICIENCY TESTING.................................................................................................................................................. 10 QUALITY MANAGEMENT................................................................................................................................................. 17

GENERAL ISSUES..................................................................................................................................................... 17 SPECIMEN COLLECTION AND HANDLING............................................................................................................. 21 PROCEDURE MANUAL.............................................................................................................................................. 23 RESULTS REPORTING.............................................................................................................................................. 27 REAGENTS................................................................................................................................................................. 28 INSTRUMENTS AND EQUIPMENT............................................................................................................................32

Instrument and Equipment Maintenance/Function Checks.................................................................................. 32 Thermometers....................................................................................................................................................... 34 Temperature-Dependent Instruments, Equipment, and Environments................................................................. 35

TEST METHOD VALIDATION AND VERIFICATION........................................................................................................36 METHOD PERFORMANCE SPECIFICATIONS - NONWAIVED TESTS................................................................... 38 REFERENCE INTERVALS..........................................................................................................................................46

INDIVIDUALIZED QUALITY CONTROL PLAN (IQCP).....................................................................................................47

4 of 52

All Common Checklist 08.17.2016

ON-LINE CHECKLIST AVAILABILITY

Participants of the CAP accreditation programs may download the checklists from the CAP website (www.cap.org) by logging into e-LAB Solutions. They are available in different checklist types and formatting options, including:

● Master — contains ALL of the requirements and instructions available in PDF, Word/XML or Excel formats

● Custom — customized based on the laboratory's activity (test) menu; available in PDF, Word/XML or Excel formats

● Changes Only — contains only those requirements with significant changes since the previous checklist edition in a track changes format to show the differences; in PDF version only. Requirements that have been moved or merged appear in a table at the end of the file.

SUMMARY OF CHECKLIST EDITION CHANGES All Common Checklist

08/17/2016 Edition

The information below includes a listing of checklist requirements with significant changes in the current edition and previous edition of this checklist. The list is separated into three categories:

1. New 2. Revised:

● Modifications that may require a change in policy, procedure, or process for continued compliance; or

● A change to the Phase 3. Deleted/Moved/Merged:

● Deleted ● Moved — Relocation of a requirement into a different checklist (requirements that have been

resequenced within the same checklist are not listed) ● Merged — The combining of similar requirements

NOTE: The listing of requirements below is from the Master version of the checklist. The customized checklist version created for on-site inspections and self-evaluations may not list all of these requirements.

NEW Checklist Requirements

Requirement Effective Date COM.01950 08/17/2016 COM.06200 07/28/2015 COM.06300 07/28/2015 COM.30980 08/17/2016 COM.40350 07/28/2015 COM.40610 07/28/2015 COM.50200 07/28/2015 COM.50300 07/28/2015 COM.50400 07/28/2015 COM.50600 07/28/2015

REVISED Checklist Requirements

Requirement Effective Date COM.01300 07/28/2015 COM.01400 08/17/2016

5 of 52

All Common Checklist 08.17.2016

COM.01600 08/17/2016 COM.01700 08/17/2016 COM.01800 08/17/2016 COM.01900 08/17/2016 COM.04000 08/17/2016 COM.04050 08/17/2016 COM.04200 07/28/2015 COM.06100 08/17/2016 COM.10000 08/17/2016 COM.10500 08/17/2016 COM.30000 08/17/2016 COM.30250 08/17/2016 COM.30300 07/28/2015 COM.30350 08/17/2016 COM.30400 07/28/2015 COM.30550 07/28/2015 COM.30575 07/28/2015 COM.30625 07/28/2015 COM.30750 08/17/2016 COM.40000 08/17/2016 COM.40200 07/28/2015 COM.40300 08/17/2016 COM.40620 08/17/2016 COM.40630 08/17/2016 COM.40640 07/28/2015 COM.40700 07/28/2015 COM.50500 08/17/2016

DELETED/MOVED/MERGED Checklist Requirements None

6 of 52

All Common Checklist 08.17.2016

UNDERSTANDING THE CAP ACCREDITATION CHECKLIST COMPONENTS

All checklist requirements contain a requirement number, subject header, phase, and a declarative statement. Some requirements also contain a NOTE and/or Evidence of Compliance.

The NOTE portion of a checklist requirement provides additional detail to assist in interpreting the requirement.

Evidence of Compliance (EOC) is intended to: ● Suggest specific examples of acceptable records; some elements are required ● Assist in inspection preparation and for managing ongoing compliance ● Drive consistent understanding of requirements

If a policy or procedure is referenced within a requirement, it is only repeated in the Evidence of Compliance if such statement adds clarity. All policies or procedures covered in the CAP checklists must be a written document. A separate policy or procedure may not be needed for items in EOC if it is already addressed by an overarching policy.

The Master version of the checklist also contains references and the inspector R.O.A.D. instructions (Read, Observe, Ask, Discover), which can provide valuable insight for the basis of requirements and on how compliance will be assessed.

INTRODUCTION

The All Common Checklist (COM) contains a core set of requirements that apply to all areas performing laboratory tests and procedures. In some instances, the same requirement exists in both the COM Checklist and in a discipline-specific checklist, but with more specificity in the discipline-specific checklist. In these situations, the discipline-specific requirement takes precedence.

One COM Checklist is provided for inspection of each laboratory section or department. If more than one inspector is assigned to inspect a section, each inspector must be familiar with the COM requirements and ensure that all testing is in compliance.

Certain requirements are different for waived versus nonwaived tests. Refer to the checklist headings and explanatory text to determine applicability based on test complexity. The current list of tests waived under CLIA may be found at http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfClia/analyteswaived.cfm

Note for non-US laboratories: Checklist requirements apply to all laboratories unless a specific disclaimer of exclusion is stated in the checklist.

DEFINITION OF TERMS

Addendum - Information appended to a final report without changing any of the language in the original report (original report is intact and unchanged).

Alternative assessment - A system for determining the reliability of laboratory examinations for which no commercial proficiency testing products are available, are not appropriate for the method or patient population served by the laboratory, or participation is not required by the accrediting organization.

Amended/amendment - Any change in the diagnosis, narrative text, or other content of a report that has been issued (minor or major). The change in an anatomic pathology report is usually in the diagnosis or narrative, but occasionally may involve a change in a number or some other quality.

7 of 52

All Common Checklist 08.17.2016

Analytical validation - The process used to confirm with objective evidence that a laboratory-developed or modified FDA-cleared/approved test method or instrument system delivers reliable results for the intended application.

Analytical verification - The process by which a laboratory determines that an unmodified FDA-cleared/ approved test performs according to the specifications set forth by the manufacturer when used as directed.

Annual - Every 12 calendar months

Biennial - Every 24 calendar months

Authority - The power to give orders or make decisions: the power or right to direct someone or control a process

Calibrator, historical - The set of archived results of a single-point calibrator that demonstrates stability of the assay over time

Check - Examination to determine the accuracy, quality or presence of any attribute of a test system

Clinical validation - The determination of the ability of a test to diagnose or predict risk of a particular health condition or predisposition, measured by sensitivity, specificity, and predictive values

Commutable - The property of a reference material that yields the same numeric result as would a patient's specimen containing the same quantity of analyte in that analytic method under discussion (i.e. matrix effects are absent).

Confirmation - Substantiation of the correctness of a value or process

Corrected/correction - Errors in test results that may include incorrect patient identification, test results, reference interval, interpretive information, or other significant information, but not minor typographical errors of no consequence.

Corrective Action - Action taken to eliminate the cause of a detected nonconformity or other undesirable situation

Correlation - Establishment of agreement between two or more measured values

Credentialing - The process of obtaining, verifying, and assessing the qualifications of a practitioner to provide care in a health care organization

Device - Any reagent, reagent product, kit, instrument, apparatus, equipment or related product, whether used alone or in combination, intended by the manufacturer to be distributed for use in vitro for the examination of human specimens

Digital image analysis - The computer-assisted detection or quantification of specific features in an image following enhancement and processing of that image, including immunohistochemistry, DNA analysis, morphometric analysis, and in situ hybridization

Equipment - Single apparatus or set of devices or apparatuses needed to perform a specific task

Examination - In the context of checklist requirements, examination refers to the process of inspection of tissues and samples prior to analysis. An examination is not an analytical test.

FDA - In the context of checklist requirements, FDA should be taken to mean the national, state, or provincial authority having jurisdiction over in vitro diagnostic test systems.

8 of 52

All Common Checklist 08.17.2016

Function Check - Confirmation that an instrument or item of equipment operates according to manufacturer's specifications before routine use, at prescribed intervals, or after minor adjustment (e.g. base line calibration, balancing/zero adjustment, thermometer calibration, reagent delivery).

High complexity - Rating given by the FDA to commercially marketed in vitro diagnostic tests based on their risks to public health. Tests in this category are seen to have the highest risks to public health.

Instrument - An analytical unit that uses samples to perform chemical or physical assays (e.g. chemistry analyzer, hematology analyzer)

Instrument platform - Any of a series of similar or identical analytical methods intended by their manufacturer to give identical patient results across all models

Laboratory Director - The individual who is responsible for the overall operation and administration of the laboratory, including provision of timely, reliable and clinically relevant test results and compliance with applicable regulations and accreditation requirements. This individual is listed on the laboratory's CAP and CLIA certificate (as applicable).

Maintenance - Those activities that prolong the life of an instrument or minimize breakdowns or mechanical malfunctions. Examples include cleaning, changing parts, fluids, tubing, lubrication, electronic checks, etc.

Moderate complexity - Rating given by the FDA to commercially marketed in vitro diagnostic tests based on their risks to public health

Modification of manufacturer's instructions - Any change to the manufacturer's supplied ingredients or modifications to the assay as set forth in the manufacturer's labeling and instructions, including specimen type, instrumentation or procedure that could affect its performance specifications for sensitivity, specificity, accuracy, or precision or any change to the stated purpose of the test, its approved test population, or any claims related to interpretation of the results

Nonwaived - Tests categorized as either moderately complex (including provider-performed microscopy) or highly complex by the US Food and Drug Administration (FDA), according to a scoring system used by the FDA

Performance verification - The set of processes that demonstrate an instrument or an item of equipment operates according to expectations upon installation and after repair or reconditioning (e.g. replacement of critical components)

Policy - 1) Set of basic principles or guidelines that direct or restrict the facility's plans, actions, and decisions; 2) Statement that tells what should or should not be done

Preventive action - Action taken to eliminate the cause of a potential nonconformity or any other undesirable potential situation

Primary source verification report - A document, usually prepared by a third party agent or company that confirms that a job applicant's degree, certificate, or diploma is authentic, licenses were granted, and reported work history (company names, locations, dates and positions held) is accurate. The confirmation is obtained through direct contact with an institution, former employer, or their authorized agents.

Primary specimen - The body fluid, tissue, or sample submitted for examination, study or analysis. It may be within a collection tube, cup, syringe, swab, slide, data file, or other form as received by the laboratory.

Procedure - 1) Specified way to carry out an activity of a process (also referred to by ISO as "work instructions"; 2) Set of steps performed that tells "how to do it" to achieve a specified outcome, including decisions to be made

Process - 1) Set of interrelated or interacting activities that transforms inputs into outputs; 2) Series of events, stages, or phases that takes place over time that tells "what happens" or "how it works"

9 of 52

All Common Checklist 08.17.2016

Proficiency testing - Evaluation of participant (laboratory or individual) performance against pre-established criteria by means of interlaboratory comparisons. In some countries, the PT programs for clinical laboratories are called "external quality assessment" programs.

Reagent - Any substance in a test system other than a solvent or support material that is required for the target analyte to be detected and its value measured in a sample.

Report errors - A report element (see GEN.41096) that is either incorrect or incomplete

Responsibility - A duty or task that an individual is required or expected to do

Secondary specimen - Any derivative of the primary specimen used in subsequent phases of testing. It may be an aliquot, dilution tube, slide, block, culture plate, reaction unit, data extract file, image, or other form during the processing or testing of a specimen. (The aliquots or images created by automated devices and tracked by internal electronic means are not secondary specimens.)

Section Director - The individual who is responsible for the medical, technical and/or scientific oversight of a specialty or section of the laboratory.

Semiannual - Every 6 calendar months

Subject to U.S. Regulations - Laboratories located within the United States and laboratories located outside of the US that have obtained or applied for a CLIA certificate to perform laboratory testing on specimens collected in the US for the assessment of the health of human beings.

Telepathology - The practice in which the pathologist views digitized or analog video or still image(s), and renders an interpretation that is included in a formal diagnostic report or is recorded in the patient record.

Testing personnel - Individuals responsible for performing laboratory assays and reporting laboratory results

Test - A qualitative, semiqualitative, quantitative, or semiquantitative procedure for detecting the presence of, or measuring an analyte

Test system - The process that includes pre-analytic, analytic, and post-analytic steps used to produce a test result or set of results. A test system may be manual, automated, multi-channel or single-use and can include reagents, components, equipment or instruments required to produce results. A test system may encompass multiple identical analyzers or devices. Different test systems may be used for the same analyte.

Waived - A category of tests defined as "simple laboratory examinations and procedures which have an insignificant risk of an erroneous result." Laboratories performing waived tests are subject to minimal regulatory requirements.

10 of 52

All Common Checklist 08.17.2016

ALL COMMON CHECKLIST

PROFICIENCY TESTING

Inspector Instructions:

● Sampling of proficiency testing policies and procedures ● Sampling of evaluations of unacceptable proficiency testing results ● Sampling of proficiency testing records including intermediate worksheets, instrument

printouts or interfaced results, proficiency testing result forms (paper or online), physically signed attestation statement and laboratory director/designee review

● Records of semi-annual alternative assessment testing, if applicable ● Evaluations of ungraded proficiency testing results, if applicable

● How are testing personnel selected to perform a proficiency testing challenge? ● What steps do you follow when you are assigned to perform proficiency testing? ● In what situations would you repeat a proficiency testing sample? ● What do you consider unacceptable proficiency testing performance and how do you

determine corrective action? ● How do you evaluate ungraded proficiency testing?

● Select a representative sample of proficiency testing results and follow records from kit receipt, selection of personnel to perform testing, test performance, use of intermediate worksheets, instrument printouts or interfaced results, and completion of proficiency testing result forms (paper or online), to the submission of results to the proficiency testing provider. Determine if the samples and results are being handled in compliance with requirements and following laboratory policies and procedures.

● Select a representative sample of unacceptable proficiency testing results and follow records from original testing to final determination of root cause. Determine if the procedures and processes produce a thorough investigation with appropriate corrective action taken

● Select an analyte with an unacceptable proficiency testing event and review records for the subsequent proficiency testing event for that analyte. Closely examine the testing records to confirm that the samples were handled and reported in the same manner as patient specimens (COM.01600) and following laboratory policy and procedure. Determine if any inappropriate actions were taken to ensure an acceptable event score such as duplicate testing of samples when not indicated in laboratory policy.

COM.01000 PT Procedure Phase II

The laboratory has written procedures for proficiency testing sufficient for the extent and complexity of testing done in the laboratory.

NOTE: The laboratory must have written procedures for the proper handling, analysis, review and reporting of proficiency testing materials. There must be written procedures for investigation and correction of problems that are identified by unacceptable proficiency testing results. The laboratory should also have procedures for investigation of results that, although acceptable show bias or trends suggesting a problem.

11 of 52

All Common Checklist 08.17.2016

CAP-accredited laboratories must participate in proficiency testing (PT) (when available through CAP or a CAP-accepted alternate provider) for all patient tests designated by CAP. The current list of analytes for which CAP requires PT is available on the CAP website [http://www.cap.org/] or by phoning 800-323-4040 (or 847-832-7000), option 1.

The CAP office audits PT participation to assure that accredited laboratories participate in PT as appropriate.

REFERENCES 1) Clinical and Laboratory Standards Institute. Using Proficiency Testing to Improve the Clinical Laboratory; Approved Guideline. 2nd

ed. CLSI Document GP27-A2. Clinical and Laboratory Standards Institute. Wayne, PA; 2007.

COM.01100 Ungraded PT Challenges Phase II

The laboratory has a procedure for assessing its performance on PT challenges that were intended to be graded, but were not.

NOTE: This requirement addresses PT challenges that were intended to be graded, but were not, for reasons such as: 1) the laboratory submitted its results after the cut-off date, 2) the laboratory did not submit results, 3) the laboratory did not complete the result form correctly (for example, submitting the wrong method code or recording the result in the wrong place). Also, if possible, the laboratory should assess its performance on PT challenges that were not graded because of lack of consensus. For guidance on the approach to these situations, refer to appendix I in the CAP Laboratory Accreditation Manual for listing of PT exception codes and actions.

Evidence of Compliance: ✓ Records of review and evaluation of ungraded PT challenges

REFERENCES 1) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Clinical laboratory improvement amendments

of 1988; final rule. Fed Register. 2003(Jan 24):3705 [42CFR493.1236(a)(2)]

COM.01200 Activity Menu Phase I

The laboratory's current CAP Activity Menu accurately reflects the testing performed.

NOTE: The Activity Menu must reflect the laboratory’s current testing, including removal of discontinued tests. The accuracy of the Activity Menu can be assessed by inquiry of responsible individuals, and by examination of the laboratory's test requisition(s), computer order screens, procedure manuals, or patient reports.

In order to ensure proper customization of the checklists, the laboratory must also ensure that the activity menu is accurate for non-test activities, such as methods and types of services offered.

Some activities are included on the Master Activity Menu using more generic groupings or panels instead of listing the individual tests. The Master Activity Menu represents only those analytes that are directly measured. Calculations are not included, with a few exceptions (e.g. INR, hematocrit).

If any tests omitted from the laboratory’s Activity Menu are not covered by the checklists provided for the inspection, the inspector should contact the CAP (800-323-4040) for instructions and record on the appropriate section page in the Inspector's Summation Report (ISR) whether those tests were inspected or not inspected.

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2004(Oct 1): 985 [42CFR493.51]

**REVISED** 07/28/2015 COM.01300 PT Participation Phase II

12 of 52

All Common Checklist 08.17.2016

The laboratory participates in the appropriate required proficiency testing (PT)/external quality assessment (EQA) program accepted by CAP for the patient testing performed.

NOTE 1: Information on analytes that require enrollment and participation in a CAP-accepted PT program is available on the CAP website [http://www.cap.org/] through e-LAB Solutions Suite under CAP Accreditation Resources, Master Activity Menu Reports. Also, the inspection packet includes a report with this information for each laboratory section/department.

NOTE 2: This checklist requirement applies to both waived and nonwaived tests.

NOTE 3: For laboratories subject to US regulations, participation in proficiency testing may be through CAP PT Programs or another proficiency testing provider accepted by CAP. Laboratories will not be penalized if they are unable to participate in an oversubscribed program. If unable to participate, however, the laboratory must implement an alternative assessment procedure for the affected analytes. For regulated analytes, if the CAP and CAP-accepted PT programs are oversubscribed, CMS requires the laboratory to attempt to enroll in another CMS-approved PT program.

NOTE 4: For laboratories not subject to US regulations, participation in proficiency testing must be through CAP PT Programs. Laboratories may use acceptable alternatives when the CAP is unable to deliver PT due to oversubscribed programs, stability issues or customs denial, contingent on CAP approval. (This went into effect as of the 2014 Proficiency Testing Program year.) If unable to participate, however, the laboratory must implement an alternative assessment procedure for the affected analytes.

NOTE 5: Proficiency testing for HER2 (ERBB2) is method specific. If the laboratory performs HER2 (ERBB2) testing by multiple methods, the laboratory must participate in PT for each method.

A. HER2 interpretation by immunohistochemistry (IHC): If the laboratory interprets its HER2 test results from IHC stains prepared at another facility, the laboratory must:

● Enroll in an appropriate PT Program ● Send PT materials to the staining facility for preparation, and ● Interpret the resulting stains using the same procedures that are used for patient

specimens

B. HER2 (ERBB2) interpretation by FISH (or ISH): If the laboratory sends its FISH (or ISH) slides for hybridization to another facility, the laboratory must perform an alternative assessment of the test twice annually and may not participate in formal (external) PT.

NOTE 6: For purposes of photograph/image identification in CAP PT Programs, it is strongly recommended that the current CAP Surveys Hematology Glossary be readily available to the bench technologist in the hematology and urinalysis sections.

Evidence of Compliance: ✓ Records such as CAP order form or purchase order indicating that the laboratory is enrolled

in CAP PT Programs for all analytes that CAP requires PT OR record of completed/submitted result forms for all analytes on the activity menu

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 1992(Feb 28):7146 [42CFR493.801] 2) Tholen DW. Reference values and participant means as targets in proficiency testing. Arch Pathol Lab Med. 1993;117:885-889 3) Borsotti M. External quality assessment scheme in Tuscany, Italy. Ann 1st Super Sanita. 1995;31:175-186 4) Westgard JO, et al. Laboratory precision performance. State of the art versus operating specifications that assure the analytical

quality required by clinical laboratory improvement amendments proficiency testing. Arch Pathol Lab Med. 1996;120:621-625 5) Ross JW, et al. The accuracy of laboratory measurements in clinical chemistry. A study of 11 routine chemistry analytes in the

College of American Pathologists chemistry survey with fresh frozen serum, definitive methods, and reference materials. Arch Pathol Lab Med. 1998;122:587-608

6) College of American Pathologists, Commission on Laboratory Accreditation. Standards for laboratory accreditation; standard III. Northfield, IL: CAP, 1998

13 of 52

All Common Checklist 08.17.2016

7) Dale JC, Hamrick HJ. Neonatal bilirubin testing practices. Reports from 312 laboratories enrolled in the College of American Pathologists Excel proficiency testing program. Arch Pathol Lab Med. 2000;124:1425-1428

8) Plebani M, et al. External quality assessment for serum proteins: state of the art. Clin Chem. 2001;47(suppl):A35 9) Panteghini M, et al. External quality assessment scheme for biochemical markers of cardiac damage. Clin Chem.

2001;47(suppl):A195 10) Wilson JF, et al. Primary standardization of assays for anticonvulsant drugs: comparison of accuracy and precision. Clin Chem.

2002;48:1963-1969 11) Taylor A, et al. Comparison of procedures for evaluating laboratory performance in external quality assessment schemes for lead in

blood and aluminum in serum demonstrates the need for common quality specifications. Clin Chem. 2002;48:2000-2007 12) Wolff AC, Hammond ME, Schwartz JN, et al. American Society of Clinical Oncology/College of American Pathologists guideline

recommendations for human epidermal growth factor receptor 2 testing in breast cancer. Arch Pathol Lab Med 2007;131:18-43 13) Nakhleh RE, Grimm EE, Idowu MO, et al. Laboratory compliance with the American Society of Clinical Oncology/College of American

Pathologists (ASC/CAP) guidelines for human epidermal growth factor 2 (HER2) testing: a College of American Pathologists survey of 757 laboratories. Arch Pathol Lab Med 134:728034, 2010

**REVISED** 08/17/2016 COM.01400 PT Attestation Statement Phase II

The proficiency testing attestation statement is signed by the laboratory director or qualified designee and all individuals involved in the testing process.

NOTE: Physical signatures must appear on a paper version of the attestation form. A listing of typed names on the attestation statement does not meet the intent of the requirement. The signature of the laboratory director or designee need not be obtained prior to reporting results to the proficiency testing provider.

Designees must be qualified through education and experience to meet the defined regulatory requirements associated with the complexity of the testing as defined in the Personnel section of the Laboratory General Checklist.

● For high complexity testing, it may be delegated to an individual meeting the qualifications of a technical supervisor or section director (GEN.53400). For the specialties of Histocompatibility, Cytogenetics, and Transfusion Medicine, refer to specific requirements for the qualifications of section directors/technical supervisors in the associated checklists (HSC.40000, CYG.50000, and TRM.50050).

● For moderate complexity testing, it may be delegated to an individual meeting the qualifications of a technical consultant (GEN.53625).

Evidence of Compliance: ✓ Appropriately signed attestation statement from submitted PT result forms

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 1992(Feb 28):7146 [42CFR493.801(b)(1)]

COM.01500 Alternative Performance Assessment Phase II

For tests for which CAP does not require PT, the laboratory at least semi-annually exercises an alternative performance assessment system for determining the reliability of analytic testing.

NOTE 1: Appropriate alternative performance assessment procedures include participation in an external PT program not required by CAP; participation in an ungraded/educational PT program; split sample analysis with referral or other laboratories, split sample analysis with an established in-house method, use of assayed materials, clinical validation by chart review, or other suitable and documented means. It is the responsibility of the laboratory director to define such alternative assessment procedures and the criteria for successful performance in accordance with good clinical and scientific laboratory practice.

NOTE 2: For FISH testing and other complex molecular and sequencing-based tests (including but not limited to microarray-based tests, multiplex PCR-based tests, and next generation sequencing-based tests), alternative assessment may be performed by method or specimen type rather than for each analyte or tested abnormality. For tests such as allergen testing, alternative assessment may be performed in batches of analogous tests.

14 of 52

All Common Checklist 08.17.2016

NOTE 3: Semiannual alternative performance assessment must be performed on tests for which external PT is not available.

NOTE 4: This checklist requirement applies to both waived and nonwaived tests.

The list of analytes for which CAP requires proficiency testing is available on the CAP website [http://www.cap.org/] or by phoning 800-323-4040 (or 847-832-7000), option 1.

Evidence of Compliance: ✓ List of tests defined by the laboratory as requiring alternative assessments AND ✓ Records of those assessments

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24):7184 [42CFR493.1236(c)(1)] 2) Shahangian S, et al. A system to monitor a portion of the total testing process in medical clinics and laboratories. Feasibility of a split-

specimen design. Arch Pathol Lab Med. 1998;122:503-511 3) Shahangian S, Cohn RD. Variability of laboratory test results. Am J Clin Pathol. 2000;113:521-527 4) CLSI. Assessment of Laboratory Tests When Proficiency Testing is not Available; Approved Guideline-Second Edition. CLSI

Document GP29-A2. [ISBN 1-56238-673-5]. CLSI, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898 USA, 2008 5) Marks V. False-positive immunoassay results: a multicenter survey of erroneous immunoassay results from assays of 74 analytes in

10 donors from 66 laboratories in seven countries. Clin Chem. 2002;48:2008-2016 6) Schrijver I, Aziz N, Jennings L, Richards CS, Voelkerding KV, Weck KE. Methods-Based Proficiency Testing in Molecular Genetic

Pathology. J Mol Diagn. May 2014;16(3):283-287.

**REVISED** 08/17/2016 COM.01600 PT Integration Routine Workload Phase II

The laboratory integrates all proficiency testing samples within the routine laboratory workload, and those samples are analyzed by personnel who routinely test patient/client samples, using the same primary method systems as for patient/client/donor samples.

NOTE: Repetitive analysis of any proficiency sample by one or more individuals is acceptable only if patient/client specimens are routinely analyzed in the same manner. With respect to morphologic examinations (identification of cell types and microorganisms; review of electrophoretic patterns, etc.), group review and consensus identifications are permitted only for unknown samples that would ordinarily be reviewed by more than one person on an actual patient sample.

Laboratories that are subject to regulation by the Centers for Medicare and Medicaid Services (CMS) are not permitted to test the same analyte from the same PT product on more than one instrument or method unless that is how the laboratory tests patient specimens and laboratory procedures are written to reflect that process.

If the laboratory (under one CLIA license) uses multiple methods for an analyte, proficiency samples must be analyzed by the primary method at the time of the PT event, or rotated among primary methods each PT shipment. Laboratories subject to CMS regulation are not allowed to order multiple PT kits for the purpose of testing the same sample/analyte on multiple instruments or methods prior to the due date for submitting results to the provider.

The educational purposes of proficiency testing are best served by a rotation that allows all testing personnel to be involved in the proficiency testing program. Proficiency testing records must be retained and can be an important part of the competency and continuing education records in the personnel files of the individuals. When external proficiency testing materials are not available, the semiannual alternative performance assessment process should also be integrated within the routine workload, if practical.

The US Department of Defense (DOD) and the Department of Veterans Affairs (VA) laboratories are subject to different regulations. For both the DOD and the VA, multiple proficiency testing kits may be ordered, with results reported, from the same proficiency testing provider on the same analyte; however, laboratories may not compare results from multiple kits until after the deadline for submission of results to the provider.

15 of 52

All Common Checklist 08.17.2016

Evidence of Compliance: ✓ Written policy describing proper handling of PT specimens AND ✓ Instrument printout and/or work records AND ✓ Completed attestation pages from submitted PT result forms

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 1992(Feb 28):7146 [42CFR493.801(b)] 2) Shahangian S, et al. Toward optimal PT use. Med Lab Observ. 2000;32(4):32-43 3) Parsons PJ. Evaluation of blood lead proficiency testing: comparison of open and blind paradigms. Clin Chem. 2001;47:322-330

**REVISED** 08/17/2016 COM.01700 PT Evaluation Phase II

There is ongoing evaluation of PT and alternative assessment results, with appropriate corrective action taken for each unacceptable result.

NOTE: Primary records related to PT and alternative assessment testing are retained for at least two years (five years for transfusion medicine). These include all instrument tapes, work cards, computer printouts, evaluation reports, evidence of review, and records of follow-up or corrective action.

For laboratories outside the US, PT failures relating to problems with shipping and specimen stability should include working with local customs and health regulators to ensure appropriate transit of proficiency testing specimens.

Evidence of Compliance: ✓ Records of ongoing review of all PT reports and alternative assessment results by the

laboratory director or designee AND ✓ Records of investigation of each "unacceptable" PT and alternative assessment result

including records of corrective action appropriate to the nature and magnitude of the problem

REFERENCES 1) Ehrmeyer SS, et al. Use of alternative rules (other than the 1-2s) for evaluating interlaboratory performance data. Clin Chem.

1988:34:250-256 2) Klee GG, Forsman RW. A user's classification of problems identified by proficiency testing surveys. Arch Pathol Lab Med.

1988;112:371-373 3) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 1992(Feb 28):7173 [42CFR493.1407(e)(4)(iv)] 4) Steindel SJ, et al. Reasons for proficiency testing failures in clinical chemistry and blood gas analysis. A College of American

Pathologists Q-Probes study in 655 laboratories. Arch Pathol Lab Med. 1996;120:1094-1101 5) Clinical and Laboratory Standards Institute (CLSI). Using Proficiency Testing to Improve the Clinical Laboratory; Approved Guideline

—Second Edition. CLSI document GP27-A2 (ISBN 1-56238-632-8). Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2007

6) Shahangian S, et al. Toward optimal PT use. Med Lab Observ. 2000;32(4):32-43 7) Zaki Z, et al. Self-improvement by participant interpretation of proficiency testing data from events with 2 to 5 samples. Clin Chem.

2000;46:A70

**REVISED** 08/17/2016 COM.01800 PT Interlaboratory Communication Phase II

There is no interlaboratory communication about proficiency testing samples until after the deadline for submission of data to the proficiency testing provider.

NOTE: Results must be reported by personnel within the laboratory. The written proficiency testing policies must strictly prohibit interlaboratory communications about proficiency testing samples or results until after the deadline for submission of data to the proficiency testing provider. The laboratory director is responsible for enforcing this prohibition. Records of training on the handling of PT samples and prevention of interlaboratory communication are strongly recommended. The laboratory must maintain the records of the proficiency testing event, including a copy of the proficiency testing program's report forms, instrument printouts, and work records.

16 of 52

All Common Checklist 08.17.2016

Proficiency testing records must not be shared with and should be inaccessible to personnel of other laboratories, including an affiliated laboratory until after the deadline for submission of results. Laboratories that share a common computer system must take appropriate steps to ensure that records are not readily accessible by other laboratories.

Evidence of Compliance: ✓ Written policy prohibiting interlaboratory communication about PT specimens AND ✓ Proficiency testing records

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 1992(Feb 28):7146 [42CFR493.801(b)(3)] 2) Bierig JR. Comparing PT results can put a lab's CLIA license on the line. Northfield, IL: College of American Pathologists CAP Today.

2002;16(2):84-87

**REVISED** 08/17/2016 COM.01900 PT Referral Phase II

Proficiency testing specimens are not referred to other laboratories and are not accepted from other laboratories for analysis.

NOTE: The written proficiency testing policies must strictly prohibit referral or acceptance of proficiency testing specimens for analysis from other laboratories. This prohibition takes precedence over the requirement that proficiency testing specimens be handled in the same manner as patient specimens. For example, a laboratory's routine procedure for review of abnormal blood smears might be referral of the smear to a pathologist located at another site. For proficiency testing specimens, the referring laboratory must NOT follow its routine procedure in this situation. Rather, the laboratory must submit a PT result indicating that the test is not performed since the review does not occur within the referring laboratory.

For laboratories subject to US regulations, this applies even if the second laboratory is in the same health care system. It is the responsibility of the laboratory director to ensure that this prohibition is enforced.

Records of training on referral and acceptance of PT samples is strongly recommended.

Refer to 'Tips for Avoiding Proficiency Testing Referral' on the CAP website (http://www.cap.org) through e-LAB Solutions Suite.

Evidence of Compliance: ✓ Written policy prohibiting PT specimen referral or acceptance from other laboratories AND ✓ Proficiency testing records

REFERENCES 1) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Clinical laboratory improvement amendments

of 1988; final rule. Fed Register. 1992(Feb 28): [42CFR493.801(b)(4)] 2) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Brochure #8. Proficiency Testing, Dos

and Don'ts. September 2008. https://www.cms.gov/Regulations-and-Guidance/Legislation/CLIA/Downloads/CLIAbrochure8.pdf. Accessed December 23, 2015.

**NEW** 08/17/2016 COM.01950 Cease Patient Testing for Repeat PT Failures Phase II

If the laboratory was instructed by the CAP to cease patient testing for an analyte or subspecialty due to repeat unsuccessful proficiency testing, laboratory records demonstrate that no patient results were released until after the laboratory received approval from the CAP to resume patient testing.

NOTE: In order to resume patient testing, the laboratory must meet the conditions as outlined in the cease patient testing notification.

17 of 52

All Common Checklist 08.17.2016

Evidence of Compliance: ✓ Records of communication notifying staff/physicians that testing is suspended for the

required period of time OR ✓ LIS report verifying that no patient results were reported for the affected analyte or

subspecialty during the cease testing time frame OR ✓ Patient reports indicating name and address of laboratory where testing was performed

during the affected period OR ✓ Send-out log to referral laboratory

REFERENCES 1) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Clinical laboratory improvement amendments

of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.807]. 2) Olson JD, Karon BS. PT failures: steps to preventing a cease testing. CAP Today. 2015;29(9):5-8.

QUALITY MANAGEMENT

GENERAL ISSUES

Inspector Instructions:

● Sampling of QM policies and procedures ● QM/QC program, including pre-analytic, analytic and post-analytic monitor records

and corrective action when indicators do not meet threshold ● Incident/error log and corrective action ● Records of high school graduate high complexity test review by supervisor ● Records of monthly review of instrument/equipment maintenance and function checks ● Semiannual instrument/method comparison records

● How do you evaluate data on the incident/error log? How do you determine appropriate corrective action?

● As a staff member, what is your role in quality management? ● How do you detect and correct laboratory errors?

● Follow an incident identified on the incident/error log and follow actions including notification and resolution

● Select several problems identified by the QM plan and follow tracking and corrective action. Determine if the methods used led to discovery and effective correction of the problem.

● Review two or three instruments or items of equipment critical for patient testing. Determine if function check and maintenance records are adequate and if the laboratory performed the appropriate follow-up when irregularities were found.

**REVISED** 08/17/2016 COM.04000 Written QM Program Phase II

The laboratory has a written quality management (QM) program.

NOTE: The program must ensure quality throughout the pre-analytic, analytic, and post-analytic (reporting) phases of testing, including patient identification and preparation; specimen collection, identification, preservation, transportation, and processing; and accurate, timely result reporting. The program must be capable of detecting problems in the laboratory's systems, and identifying

18 of 52

All Common Checklist 08.17.2016

opportunities for system improvement. The laboratory must be able to develop plans of corrective action based on data from its QM system.

All QM requirements in the Laboratory General Checklist pertain to the laboratory.

Evidence of Compliance: ✓ Records reflecting conformance with the program as designed AND ✓ Results of quality surveillance

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 1992(Feb 28):7176 [42CFR493.1445(e)(5)] 2) Clinical and Laboratory Standards Institute. Quality Management System: A Model for Laboratory Services; Approved Guideline. 4th

ed. CLSI Document QMS01-A4. Clinical and Laboratory Standards Institute, Wayne, PA; 2011. 3) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1289]

**REVISED** 08/17/2016 COM.04050 Error Detection and Correction Phase II

There is a written procedure for the detection and correction of significant clerical and analytical errors, and unusual laboratory results, in a timely manner.

NOTE: One common method is review of results by a qualified person (technologist, supervisor, pathologist) before release from the laboratory, but there is no requirement for supervisory review of all reported data for tests that do not include interpretation. In computerized laboratories, there should be automatic "traps" for improbable results. The system for detecting clerical errors, significant analytical errors, and unusual laboratory results should provide for timely correction of errors, i.e. before results become available for clinical decision making. For confirmed errors detected after reporting, corrections must be promptly made and reported to the ordering physician or referring laboratory, as applicable.

If laboratories use delta checks as a mechanism to detect errors prior to the reporting of patient results, the laboratory must have written procedures describing the actions to be taken when acceptability criteria are exceeded and a process for approval of new or changed delta checks by the laboratory director or designee.

Each procedure must include a listing of common situations that may cause analytically inaccurate results, together with a procedure to address such analytic errors or interferences. This may require alternate testing methods; in some situations, it may not be possible to report results for some or all of the tests requested.

The intent of this requirement is NOT to require verification of all results outside the reference (normal) range.

Evidence of Compliance: ✓ Records of review of results OR records of consistent implementation of the error detection

system(s) defined in the procedure AND ✓ Records of timely corrective action of identified errors

REFERENCES 1) Dufour D, et al. The clinical significance of delta checks. Am J Clin Pathol. 1998;110:531 2) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1281(b)] 3) Clinical and Laboratory Standards Institute. Use of Delta Checks in the Medical Laboratory; 1st ed. CLSI Document EP33-ED1.

Clinical and Laboratory Standards Institute. Wayne, PA; 2016.

COM.04100 Supervisory Result Review Phase II

In the absence of on-site supervisors, high complexity testing performed by trained high school graduates qualifying as high complexity testing personnel is reviewed by the laboratory director or supervisor/general supervisor within 24 hours.

19 of 52

All Common Checklist 08.17.2016

NOTE: The CAP does NOT require supervisory review of all test results before or after reporting to patient records. Rather, this requirement is intended to address only that situation for "high complexity testing" performed by trained high school graduates qualifying under the CLIA regulation 42CFR493.1489(b)(5)(i) when a qualified supervisor/general supervisor is not present.

The qualifications to perform high complexity testing can be accessed using the following link: CAP Personnel Requirements by Testing Complexity.

Evidence of Compliance: ✓ Written policy defining the personnel and test results requiring review AND ✓ Records of result review for specified personnel

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 1992(Feb 28):7182 [42CFR493.1463(a)(3) and 42CFR493.1463(c)]: 7183 [42CFR493.1489(b)(1)] and [42CFR493.1489(b)(5)]

**REVISED** 07/28/2015 COM.04200 Instrument/Equipment Record Review Phase II

Instrument and equipment maintenance and function check records are reviewed and assessed at least monthly by the laboratory director or designee.

NOTE: The review of the records related to tests that have an approved IQCP must include an assessment of whether further evaluation of the risk assessment and quality control plan is needed based on problems identified (e.g. trending for repeat failures, etc.).

COM.04250 Comparability of Instruments and Methods - Nonwaived Testing Phase II

If the laboratory uses more than one nonwaived instrument/method to test for a given analyte, the instruments and methods are checked against each other at least twice a year for comparability of results.

NOTE: This requirement applies to tests performed on the same or different instrument makes/ models or by different methods. The purpose of the requirement is to evaluate the relationship between test results using different methodologies, instruments, or testing sites. This comparison is required only for nonwaived instruments/methods accredited under a single CAP number. The laboratory must establish a written procedure for this check that includes acceptance criteria. This requirement is not applicable to calculated parameters.

Quality control data may be used for this comparison for tests performed on the same instrument platform, with both control materials and reagents of the same manufacturer and lot number.

Otherwise, the use of human samples, rather than stabilized commercial controls, is preferred to avoid potential matrix effects. The use of pooled patient samples is acceptable since there is no change in matrix. In cases when availability or pre-analytical stability of patient/client specimens is a limiting factor, alternative protocols based on QC or reference materials may be necessary but the materials used should be validated (when applicable) to have the same response as fresh human samples for the instruments and methods involved.

This requirement only applies when the instruments/reagents are producing the same reportable result. For example, some laboratories may use multiple aPTT reagents with variable sensitivity to the lupus anticoagulant. If these are defined as separate tests, then this requirement does not apply unless each type of aPTT test is performed on more than one analyzer.

For Microbiology testing, this requirement applies when two instruments (same or different manufacturers) are used to detect the same analyte. Two or more detectors or incubation cells connected to a single data collection, analysis and reporting computer need not be considered separate systems (e.g. multiple incubation and monitoring cells in a continuous monitoring blood culture instrument, two identical blood culture instruments connected to a single computer system, or multiple thermocycler cells in a real time polymerase chain reaction instrument). This

20 of 52

All Common Checklist 08.17.2016

checklist requirement does not apply to multiple analytical methods (e.g. antigen typing versus culture or detection of DNA versus a biochemical characteristic) designed to detect the same analyte.

Evidence of Compliance: ✓ Written procedure for performing instrument and method comparison AND ✓ Records of comparability studies reflecting performance at least twice per year with

appropriate specimen types

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Medicare, Medicaid and CLIA programs;

CLIA fee collection; correction and final rule. Fed Register. 2003(Jan 24):5236 [42CFR493.1281(a)] 2) Podczasy JJ, et al. Clinical evaluation of the Accu-Chek Advantage blood glucose monitoring system. Lab Med. 1997;28:462-466 3) Ross JW, et al. The accuracy of laboratory measurements in clinical chemistry: a study of eleven analytes in the College of

American Pathologists Chemistry Survey with fresh frozen serum, definitive methods and reference methods. Arch Pathol Lab Med. 1998;122:587-608

4) Miller WG, Myers GL, Ashwood ER, et al. State of the Art in Trueness and Inter-Laboratory Harmonization for 10 Analytes in General Clinical Chemistry. Arch Pathol Lab Med 2008;132:838-846

5) Clinical and Laboratory Standards Institute. Verification of Comparability of Patient Results within One Healthcare System: Approved Guideline (Interim Revision). CLSI Document EP31-A-IR. Clinical and Laboratory Standards Institute, Wayne, PA; 2012.

6) Miller WG, Erek A, Cunningham TD, et al. Commutability limitations influence quality control results with different reagent lots. Clin Chem. 2011;57:76-83

COM.04300 Comparability Criteria - Nonwaived Testing Phase II

Acceptability criteria are defined for comparability of nonwaived instruments and methods used to test the same analyte, with records of action when the criteria are not met.

NOTE: Statistically defined acceptability limits should be used for quantitative assays.

Evidence of Compliance: ✓ Records of comparability studies with evidence of review and action taken, as appropriate

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1282(A)] 2) Clinical and Laboratory Standards Institute (CLSI). Protocol for the Evaluation, Validation, and Implementation of Coagulometers:

Approved Guideline. CLSI document H57-A (ISBN 1-56238-656-5).Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2008.

21 of 52

All Common Checklist 08.17.2016

SPECIMEN COLLECTION AND HANDLING

Inspector Instructions:

● Sampling of specimen collection and handling policies and procedures ● Sampling of specimen rejection records/log

● Sampling of patient specimens and derivatives of the primary specimen used during testing (specimen labeling, presentation, integrity)

● What is your course of action when you receive unacceptable or sub-optimal specimens?

● How does your laboratory ensure specimen integrity throughout processing and testing?

● What identifiers are in place on specimens (e.g. slides, aliquots, etc.) derived from the primary specimen?

COM.06000 Specimen Collection Manual Phase II

There are written procedures describing methods for patient identification, patient preparation, specimen collection and labeling, specimen preservation, and conditions for transportation, and storage before testing, consistent with good laboratory practice.

NOTE: The proximity of the patient to the test site does not preclude the need for proper identification systems to prevent reporting of one patient's result to another's record. Refer to the Specimen Collection section of the Laboratory General Checklist for additional information on patient identification. The procedure may be in paper or electronic form.

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Oct 1):1034 [42CFR493.1242(a)]

**NEW/REVISED** 08/17/2016 COM.06100 Primary Specimen Container Labeling Phase II

All primary specimen containers are labeled with at least two patient-specific identifiers.

NOTE: A primary specimen container is the innermost container that holds the original specimen prior to processing and testing. This may be in the form of a specimen collection tube, cup, syringe, swab, slide or other form of specimen storage. Data files received from other laboratories for analysis are considered a specimen and must contain acceptable patient identifiers. Criteria for acceptable specimen labeling and the handling of sub-optimal specimens must be defined in laboratory policy.

Examples of acceptable identifiers include, but are not limited to: patient name, date of birth, hospital number, social security number, requisition number, accession number, unique random

22 of 52

All Common Checklist 08.17.2016

number. A location (e.g. hospital room number) is not an acceptable identifier. Identifiers may be in a machine readable format, such as a barcode.

For prepared slides submitted to the laboratory, if the slides only contain one identifier, they must be securely submitted in a container labeled with two identifiers.

In limited situations, a single identifier may be used if it can uniquely identify the specimen. For example, in a trauma situation where a patient's identification is not known, a specimen may be submitted for testing labeled with a unique code that is traceable to the trauma patient. Other examples may include forensic specimens, coded or de-identified research specimens, or donor specimens labeled with a unique code decryptable only by the submitting location.

For specimens where site of origin is critical to the analysis (e.g. site specific cultures, surgical and cytology specimens), the primary specimen container and/or the requisition must clearly identify the site of origin, and as appropriate, the laterality of the specimen (right versus left). If more than one specimen container is submitted with one requisition, each container must be labeled in a manner to ensure linkage of the specimen to the site of origin and laterality.

This requirement does not apply to the labeling of specimens collected for immediate bedside patient testing performed in the presence of the patient. If the specimens are (or may be) utilized for testing away from the patient, the labeling criteria defined in this requirement apply.

Evidence of Compliance: ✓ Written policy with criteria for acceptable labeling of primary specimen containers AND ✓ Specimen collection procedures with defined labeling specifications OR ✓ Records of compliance audits for specimen labeling

REFERENCES 1) Clinical and Laboratory Standards Institute. Specimen Labels: Content and Location, Fonts, and Label Orientation; Approved

Standard. CLSI Document AUTO12-A. Clinical and Laboratory Standards Institute. Wayne, PA; 2011. 2) So You're Going to Collect a Blood Specimen. An Introduction to Phlebotomy, 12th ed. Northfield, IL: College of American

Pathologists, 2007. 3) Clinical and Laboratory Standards Institute. Laboratory Automation: Bar Codes for Specimen Container Identification; Approved

Standard. 2 nd

ed. CLSI document AUTO02-A2. Clinical and Laboratory Standards Institute. Wayne, PA; 2006.

**NEW** 07/28/2015 COM.06200 Secondary Specimen Container Labeling Phase II

Adequate specimen identification is provided on specimen containers throughout all phases of testing, including, but not limited to aliquots, dilution, tubes, slides, blocks, culture plates, reaction units, nucleic acids and other extracts, data extract files, images, and other secondary specimens created during the processing or testing of a specimen.

NOTE: A single, unique identifier may be used to label materials derived from the primary specimen for use in subsequent phases of testing. The specimen identification system used must provide reliable identification of the secondary specimen and be linked to the full particulars of patient identification, collection date, specimen type, etc. The specimen identifier(s) must be indelible, legible, and able to withstand all stages of processing and conditions of storage. Identification may be text-based, numeric, bar-coded, etc. The form of this system is entirely at the discretion of each laboratory and must be defined in laboratory procedure.

Slides prepared from specimens in the laboratory are considered secondary specimen containers. Slides prepared in the patient setting and brought to the laboratory (e.g. fine needle aspiration, bone marrow preparations) are considered primary specimen containers and must follow the labeling requirements for primary specimen containers.

For histology specimens, each block of tissue must be identified by a unique identifier traceable to the primary specimen (e.g. accession number) assigned to the case and by any descriptive letter(s)/number(s) added by the prosector during the dissection. If additional blocks are prepared later, all lists and logs must reflect these additions. Identification number and letter(s)/number(s) must be affixed to all blocks in a manner that remains legible. Each slide must be identified by the unique identifier traceable to the primary specimen and descriptive letters unique to the block

23 of 52

All Common Checklist 08.17.2016

from which it is cut. Other appropriate identifiers should be included as applicable (e.g. levels of sectioning). Automated prelabeling systems are acceptable.

REFERENCES 1) Clinical and Laboratory Standards Institute. Specimen Labels: Content and Location, Fonts, and Label Orientation; Approved

Standard. CLSI Document AUTO12-A. Clinical and Laboratory Standards Institute. Wayne, PA; 2011. 2) Clinical and Laboratory Standards Institute. Laboratory Automation: Bar Codes for Specimen Container Identification; Approved

Standard. 2nd ed. CLSI document AUTO02-A2. Clinical and Laboratory Standards Institute. Wayne, PA; 2006. 3) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Oct 1):1034 [42CFR493.1242(a)]

**NEW** 07/28/2015 COM.06300 Specimen Rejection Criteria Phase II

There are written criteria for the rejection of unacceptable specimens, instructions for the special handling of sub-optimal specimens, and records of disposition of all unacceptable specimens in the patient/client report and/or quality management records.

NOTE: The test report must indicate information regarding the condition and disposition of specimens that do not meet the laboratory's criteria for acceptability.

This requirement applies to specimens received for all types of testing and does not imply that all "unsuitable" specimens are discarded or not analyzed. If there is a problem with a specimen (e.g. improperly collected or stored, insufficient quality/quantity of specimen, inadequate labeling or requisition information, broken slides, hemolysis, lipemia, gross contamination, etc.), there must be a mechanism to notify clinical personnel responsible for patient care. If the treating physician desires the result, then the laboratory must note the condition of the specimen on the report. Some or all tests may be incorrect on such a specimen. The laboratory may wish to record that a dialogue was held with the physician, when such occurs.

For newborn screening specimens, rejection criteria must be consistent with the criteria defined in the current edition of the CLSI NBS01 Standard, Blood Collection on Filter Paper for Newborn Screening Programs.

Evidence of Compliance: ✓ Records of rejected specimens AND ✓ Instructions for special handling of sub-optimal specimens AND ✓ Records of disposition of unacceptable specimens

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24):7183 [42CFR493.1249(a) and (b)] 2) Jones BA, et al. Chemistry specimen acceptability. A College of American Pathologists Q-Probes study of 453 laboratories. Arch

Pathol Lab Med. 1997;121:19-26 3) Hill DM, et al. Effects of temperature on stability of blood homocysteine in collection tubes containing 3-deazaadenosine. Clin Chem.

2002;48:2017-2022 4) Kilinç AS, et al. Falsely increased free triiodothyronine in sera stored in serum separator tubes. Clin Chem. 2002;48:2296-2297 5) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1283(a)(3)] 6) Clinical and Laboratory Standards Institute. Accuracy in Patient and Sample Identification; Approved Guideline. CLSI Document

GP-33A. Clinical and Laboratory Standards Institute, Wayne, PA; 2010. 7) Clinical and Laboratory Standards Institute. Blood Collection on Filter Paper for Newborn Screening Programs; Approved Standard.

6 th

ed. CLSI Document NBS01-A6. Clinical and Laboratory Standards Institute. Wayne, PA; 2013. 8) Boos MS, et al. Temperature and storage dependent changes in hematologic variable and peripheral blood morphology. Am J Clin

Pathol. 1998; 110:537. 9) United Network for Organ Sharing. Bylaws Appendix B. Standard for histocompatibility testing C2.140. Richmond, VA: UNOS, 2001. 10) Clinical and Laboratory Standards Institute. Enumeration of Immunologically Defined Cell Populations by Flow Cytometry; Approved

Guideline. 2 nd

ed. CLSI Document H42-A2. Clinical and Laboratory Standards Institute, Wayne, PA; 2007.

PROCEDURE MANUAL

The procedure manual should be used by personnel at the workbench and must include the following elements, when applicable to the test procedure:

24 of 52

All Common Checklist 08.17.2016

1. Principle and clinical significance 2. Requirements for patient preparation; specimen collection, labeling, storage, preservation,

transportation, processing, and referral; and criteria for specimen acceptability and rejection 3. Microscopic examination, including the detection of inadequately prepared slides 4. Step-by-step performance of the procedure, including test calculations and interpretation of results 5. Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in

testing 6. Calibration and calibration verification procedures 7. The analytic measurement range for test results for the test system, if applicable* 8. Control procedures 9. Corrective action to take when calibration or control results fail to meet the laboratory's criteria for

acceptability 10. Limitations in the test methodology, including interfering substances 11. Reference intervals (normal values) 12. Imminently life-threatening (critical) test results 13. Pertinent literature references 14. The laboratory's system for entering results in the patient record and reporting patient results

including, when appropriate, the procedure for reporting imminently life-threatening (critical) results 15. Description of the course of action to take if a test system becomes inoperable

(*The analytic measurement range may not apply to qualitative or semi-quantitative tests.)

The manual should address relevant pre-analytic and post-analytic considerations, as well as the analytic activities of the laboratory. The specific style and format of procedure manuals are at the discretion of the laboratory director.

Inspector Instructions:

● Representative sample of procedures for completeness, laboratory director approval, and review. Current practice must match contents of policies and procedures.

● How do you access procedures? ● What procedure has most recently been implemented or modified? ● How do you ensure all copies of procedures are up to date? ● How are changes in procedures documented and communicated to staff? ● How are discontinued policies and procedures removed from general access? ● Show me how you access procedures when your network is down

● Identify a newly-implemented procedure in the prior two years and follow the steps through authoring, laboratory director approval, and staff training

**REVISED** 08/17/2016 COM.10000 Procedure Manual Phase II

A complete procedure manual is available in a paper-based, electronic, or web-based format at the workbench or in the work area.

NOTE 1: The use of inserts provided by manufacturers is not acceptable in place of a procedure manual. However, such inserts may be used as part of a procedure description, if the insert

25 of 52

All Common Checklist 08.17.2016

accurately and precisely describes the procedure as performed in the laboratory. Any variation from this printed or electronic procedure must be detailed in the procedure manual. In all cases, procedures must match the laboratory's practice, the laboratory's practice must follow written procedure, and appropriate reviews must occur.

NOTE 2: A manufacturer's procedure manual for an instrument/reagent system may be acceptable as a component of the overall departmental procedures. Any modification to or deviation from the manufacturer's manual must be clearly recorded and approved.

NOTE 3: Card files or similar systems that summarize key information are acceptable for use as quick reference at the workbench provided that:

■ A complete manual is available for reference ■ The card file or similar system corresponds to the complete manual and is

subject to document control

NOTE 4: Electronic manuals accessed by computer are fully acceptable. There is no requirement for paper copies to be available for the routine operation of the laboratory, as long as the electronic versions are readily available to all personnel and personnel have been trained on how to access them. However, procedure manuals must be available to laboratory personnel when the electronic versions are inaccessible (e.g. during laboratory information system or network downtime); thus, the laboratory must maintain paper copies, electronic copies on CD or other digital media, or have an approved alternative mechanism to access web-based files during network downtimes. All procedures, in either electronic or paper form, must be readily available for review by the inspector at the time of the CAP inspection.

Electronic procedure manuals and electronic copies of procedures are subject to proper document control (see GEN.20375), and there must be records of biennial review. Review of electronic procedures may be recorded by including statements such as “reviewed by [name of reviewer] on [date of review]” in the electronic record. Records of review by a secure electronic signature are NOT required. Alternatively, paper review sheets may be used to record review of electronic procedures.

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24):7164 [42CFR493.1251(a) (b) (1-14)(c)(d)(e)] 2) Borkowski A, et al. Intranet-based quality improvement documentation at the Veterans Affairs Maryland health care system. Mod.

Pathol. 2001;14:1-5 3) Clinical and Laboratory Standards Institute (CLSI). Quality Management System: Development and Management of Laboratory

Documents; Approved Guideline - Sixth Edition. CLSI document QMS02-A6 (ISBN 1-56238-869-X). Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087 USA, 2013.

COM.10100 Procedure Manual Review Phase II

There are records of review of all technical policies and procedures by the current laboratory director or designee at least every two years.

NOTE: The laboratory director must ensure that the collection of testing policies and technical procedures is complete, current, and has been thoroughly reviewed by a knowledgeable person. Technical approaches must be scientifically valid and clinically relevant. To minimize the burden on the laboratory and reviewer(s), it is suggested that a schedule be developed whereby roughly 1/24 of all technical policies and procedures are reviewed monthly. Paper/electronic signature review must be at the level of each procedure, or as multiple signatures on a listing of named procedures. A single signature on a Title Page or Index is not a sufficient record that each policy or procedure has been carefully reviewed. Signature or initials on each page of a policy or procedure is not required.

Only technical policies and procedures are addressed in this requirement. Biennial review is not required for other controlled documents.

REFERENCES

26 of 52

All Common Checklist 08.17.2016

1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 1992(Feb 28):7173 [42CFR493.1407(e)(13)]

2) Borkowski A, et al. Intranet-based quality improvement documentation at the Veterans Affairs Maryland health care system. Mod. Pathol. 2001;14:1-5

COM.10200 New Procedure Review Phase II

The laboratory director reviews and approves all new technical policies and procedures, as well as substantial changes to existing documents, before implementation.

NOTE: This review may not be delegated to designees in laboratories subject to the CLIA regulations.

Paper or electronic signature review of records is required. A secure electronic signature is desirable, but not required.

Evidence of Compliance: ✓ Policy on procedure review AND ✓ Records of new policy or procedure approval

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24):7164 [42CFR493.1251(d)]

COM.10250 New Procedure Review (Not Subject to US Regulations) Phase II

For laboratories not subject to US regulations, the laboratory director or designee who meets CAP director qualifications reviews and approves all new technical policies and procedures, as well as substantial changes to existing documents before implementation.

NOTE: Paper or electronic signature review of records is required. A secure electronic signature is desirable, but not required.

Evidence of Compliance: ✓ Policy on procedure review AND ✓ Records of new policy or procedure approval

COM.10300 Knowledge of Policies and Procedures Phase II

The laboratory has a defined process and records indicating that all personnel are knowledgeable about the contents of the policies and procedures (including changes) relevant to the scope of their testing activities.

NOTE: The form of this system is at the discretion of the laboratory director. Annual procedure sign-off by testing personnel is not specifically required.

Evidence of Compliance: ✓ Records indicating that the testing personnel have read the policies and procedures, new

and revised, OR records of another written method approved by the laboratory director

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1251(a)]

**REVISED** 08/17/2016 COM.10500 Discontinued Policies and Procedures Phase II

When a policy or procedure is discontinued, a paper or electronic copy is maintained for at least two years (five years for transfusion medicine), recording initial date of use, and retirement date.

NOTE: For testing on minors (under the age of 21), stricter state regulations may apply.

27 of 52

All Common Checklist 08.17.2016

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24):7164 [42CFR493.1105(a)(2)], [42CFR493.1251(e)]

RESULTS REPORTING

Inspector Instructions:

● Sampling of critical patient results/log

● How do you record the reporting of critical results? Who do you contact?

● Follow a critical result from testing, reporting and recording of notification

COM.29950 Reference Intervals Phase II

All patient/client results are reported with reference (normal) intervals or interpretations as appropriate.

NOTE: The laboratory must report reference intervals or interpretations with patient/client results, where such exist. This is important to allow proper interpretation of patient/client data. Age- and/or sex-specific reference intervals or interpretive ranges must be reported with patient test results, as applicable. In addition, the use of high and low flags (generally available with a computerized laboratory information system) is recommended. It is not necessary to include reference intervals when test results are reported as part of a treatment protocol that includes clinical actions, which are based on the test result.

Under some circumstances it may be appropriate to distribute lists or tables of reference intervals to all users and sites where reports are received. This system is usually fraught with difficulties, but if in place and rigidly controlled, it is acceptable.

REFERENCES 1) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Clinical laboratory improvement amendments

of 1988; final rule. Fed Register. 2003(Jan 24):7162 [42CFR493.1291(d)] 2) Clinical and Laboratory Standards Institute. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory;

Approved Guideline. 3 rd

ed. CLSI Document EP28-A3c. Clinical and Laboratory Standards Institute. Wayne, PA; 2008.

**REVISED** 08/17/2016 COM.30000 Critical Result Notification Phase II

The laboratory has written procedures for immediate notification of a physician (or other clinical personnel responsible for the patient's care) when results of designated tests exceed established "critical" values that are important for prompt patient management decisions. Records of notification are maintained.

28 of 52

All Common Checklist 08.17.2016

NOTE: Alert or critical results are those results that may require rapid clinical attention to avert significant patient morbidity or mortality. Each laboratory may define the critical values and critical results that pertain to its patient population. The laboratory may establish different critical results for specific patient subpopulations (for example, dialysis clinic patients). Critical results should be defined by the laboratory director, in consultation with the clinicians served. For changes to anatomic pathology and cytopathology reports, refer to ANP.12175 and CYP.06450 instead.

Allowing clinicians to "opt out" of receiving critical results is strongly discouraged.

Records must be maintained showing prompt notification of the appropriate clinical individual after obtaining results in the critical range. These records must include: date, time, responsible laboratory individual, person notified (the person's first name alone is not adequate documentation), and test results. Any problem encountered in accomplishing this task should be investigated to prevent recurrence.

Referral laboratories may report critical results directly to clinical personnel, or to the referring laboratory. The referral laboratory should have a written agreement with the referring laboratory that indicates to whom the referral laboratory reports critical results.

In the point-of-care setting, the identity of the testing individual and person notified need not be recorded when the individual performing the test is the same person who treats the patient. In this circumstance, however, there must be a record of the critical result, date, and time in the test report or elsewhere in the medical record.

REFERENCES 1) Kost GJ. Critical limits for urgent clinician notification at US medical centers. JAMA. 1990;263:704-707 2) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1291(g)] 3) Steindel SJ, Heard NV. Critical values: data analysis and critique. Q-Probes 92-04. Northfield, IL: College of American Pathologists,

1992 4) Emancipator K. Critical values. ASCP practice parameter. Am J Clin Pathol. 1997:108:247-253

COM.30100 Critical Result Read-Back Phase I

When critical results are communicated by phone, “read-back” of the results is requested and recorded.

NOTE: Transmission of critical results by electronic means (FAX or computer) is acceptable. If critical results are transmitted electronically, the laboratory must confirm receipt of the result by the intended recipient (e.g. by a phone call); however, no read-back is necessary.

Evidence of Compliance: ✓ Records of critical result notification, including read-back as necessary

REAGENTS

Inspector Instructions:

● Sampling of test procedures for reagent handling ● Sampling of new reagent/shipment confirmation of acceptability records ● Sampling of ambient temperature logs (if reagents stored at ambient temperature)

● Sampling of reagents (expiration date, labeling, storage)

29 of 52

All Common Checklist 08.17.2016

● How do you store the reagents and controls used in test procedures? ● How do you confirm the acceptability of new reagent lots? ● If you identify a problem with a reagent in use (e.g. expired vial, unacceptable storage

conditions, etc.), what is your process for evaluating the potential impact on patients? ● What are your laboratory's criteria for mixing components from one lot number of

reagent kit with components from another lot number of kit? ● How does your laboratory manage and control reagent inventory?

**REVISED** 08/17/2016 COM.30250 Reagent Storage and Handling - Waived Tests Phase II

For waived tests, the laboratory follows manufacturer instructions for handling and storing reagents, cartridges, test cards, etc.

NOTE: There is no requirement to routinely label individual containers with "date opened"; however, a new expiration date must be recorded if opening the container changes the expiration date, storage requirement, etc.

If the manufacturer defines a required storage temperature range, the temperature of storage areas must be monitored daily. Refer to the Temperature-Dependent Instruments, Equipment, and Environment section of the checklist for requirements for monitoring and recording temperature.

If the laboratory identifies a problem with a reagent that was used for patient testing (e.g. expired vial or reagent subjected to unacceptable storage conditions, etc.), the laboratory must evaluate the potential impact on patient test results and maintain records of the evaluation and actions taken.

Evidence of Compliance: ✓ Records of reagent storage and handling consistent with manufacturer's instructions,

including refrigerator, freezer and room temperature monitoring

The remaining checklist requirements in the REAGENTS section do not apply to waived tests.

**REVISED** 07/28/2015 COM.30300 Reagent Labeling Phase II

Reagents, calibrators, controls, stains, chemicals, and solutions are properly labeled, as applicable and appropriate, with the following elements.

1. Content and quantity, concentration or titer 2. Storage requirements 3. Date prepared, filtered or reconstituted by laboratory 4. Expiration date

NOTE: The above elements may be recorded in a log (paper or electronic), rather than on the containers themselves, providing that all containers are identified so they are traceable to the appropriate data in the log. While useful for inventory management, labeling with "date received" is not routinely required. There is no requirement to routinely label individual containers with "date opened"; however, a new expiration date must be recorded if opening the container changes the expiration date, storage requirement, etc.

This requirement also applies to the labeling of chemicals used in the laboratory to prepare reagents or during the preanalytic and analytic phases of the testing process. Requirements relating to precautionary labeling for hazardous chemicals are included in the Chemical Safety section of the Laboratory General Checklist.

30 of 52

All Common Checklist 08.17.2016

Evidence of Compliance: ✓ Written procedure defining elements and requirements for reagent labeling

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24):7164 [42CFR493.1252(c)]

**REVISED** 08/17/2016 COM.30350 Reagent Storage and Handling Phase II

All reagents and media are stored and handled as defined by the laboratory and following the manufacturer's instructions.

NOTE: Reagents and media must be stored and handled in a manner that will prevent environmentally-induced alterations that could affect reagent stability and test performance. Prepared reagents must be properly stored, mixed, when appropriate, and discarded when stability parameters are exceeded.

If the manufacturer defines a required storage temperature range, the temperature of storage areas must be monitored daily. Refer to the Temperature-Dependent Instruments, Equipment, and Environment section of the checklist for requirements for monitoring and recording temperature.

If the laboratory identifies a problem with a reagent that was used for patient testing (e.g. expired vial or reagent subjected to unacceptable storage conditions, etc.), the laboratory must evaluate the potential impact on patient test results and maintain records of the evaluation and actions taken.

Evidence of Compliance: ✓ Records of reagent and media storage and handling consistent with manufacturer's

instructions, including refrigerator, freezer and room temperature monitoring

REFERENCES 1) Gonzales Y, Kampa IS. The effect of various storage environments on reagent strips. Lab Med. 1997;28:135-137 2) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1252(b)] 3) Clinical and Laboratory Standards Institute (CLSI). One-Stage Prothrombin Time (PT) Test and Activated Partial Thromboplastin

Time (aPTT) Test; Approved Guideline—Second Edition. CLSI document H47-A2 (ISBN 1-56238-672-7). Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087 USA, 2008.

**REVISED** 07/28/2015 COM.30400 Reagent Expiration Date Phase II

All reagents, chemicals, and media are used within their indicated expiration date.

NOTE: The laboratory must assign an expiration date to any reagents and media that do not have a manufacturer-provided expiration date. The assigned expiration date should be based on known stability, frequency of use, storage conditions, and risk of deterioration.

This requirement also applies to the labeling of chemicals used in the laboratory. Requirements relating to precautionary labeling for hazardous chemicals are included in the Chemical Safety section of the Laboratory General Checklist.

Separate requirements for rare blood banking reagents are included in the Transfusion Medicine Checklist.

For laboratories not subject to US regulations and military laboratories in overseas locations, expired reagents may be used only under the following circumstances: 1) The reagents are unique, rare or difficult to obtain; or 2) Delivery of new shipments of reagents is delayed through causes not under control of the laboratory. The laboratory must maintain records of verification of the performance of expired reagents in accordance with written laboratory procedure. The laboratory must also maintain records of its efforts to obtain reagents in a timely manner and the rationale for continuing to perform the test instead of referring it to another laboratory.

31 of 52

All Common Checklist 08.17.2016

Laboratories subject to US regulations must not use expired reagents.

Evidence of Compliance: ✓ Written procedure for evaluating reagents and media lacking manufacturer's expiration date

AND ✓ Records confirming acceptability of any reagent used beyond its expiration date (in

jurisdictions where allowed)

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24):7164 [42CFR493.1252(d)]

COM.30450 New Reagent Lot Confirmation of Acceptability Phase II

New reagent lots and shipments are checked against old reagent lots or with suitable reference material before or concurrently with being placed in service.

NOTE: The purpose of this check is to confirm that the use of new reagent lots and shipments do not affect patient results. Matrix interferences between different lots of reagents may impact the calibration status of instruments and consistency of patient results. Improper storage conditions during shipping of reagents may have a negative impact on their ability to perform or exhibit the same levels of reactivity as intended.

Qualitative: For qualitative nonwaived tests, minimum cross-checking includes retesting at least one positive and negative sample with known reactivity against the new reagent lot. A weakly positive sample should also be used in systems where patient results are reported in that fashion.

Examples of suitable reference materials for qualitative tests include:

1. Positive and negative patient samples tested on a previous lot; 2. Previously tested proficiency testing materials; 3. External QC materials tested on the previous lot; 4. Control strains of organisms or previously identified organisms for microbiology

reagents used to detect or evaluate cultured microorganisms.

For flow cytometry reagents, please refer to the Reagents section of the Flow Cytometry Checklist.

Quantitative: For quantitative nonwaived tests, patient specimens should be used to compare a new lot against the old lot, when possible. Manufactured materials, such as proficiency testing (PT) or QC materials may be affected by matrix interference between different reagent lots, even if results show no change following a reagent lot change. The use of patient samples confirms the absence of matrix interference. Other than patient samples, the following materials may also be used:

1. Reference materials or QC products provided by the method manufacturer with method specific and reagent lot specific target values;

2. Proficiency testing materials with peer group established means; 3. QC materials with peer group established means based on interlaboratory

comparison that is method specific and includes data from at least 10 laboratories; 4. Third party general purpose reference materials if the material is affirmed in the

package insert or by the method manufacturer to be commutable with patient specimens for the method.

5. QC material used to test the current lot is adequate alone to check a new shipment of the same reagent lot, as there should be no change in potential matrix interactions between the QC material and different shipments of the same lot number of reagents.

For hematology analyzers, reservoirs containing testing reagents and cleaning/decontaminating solutions must be checked according to manufacturer's instructions.

32 of 52

All Common Checklist 08.17.2016

Evidence of Compliance: ✓ Written procedure for the confirmation of acceptability of new lots and shipments, with

defined acceptability criteria AND ✓ Records for the introduction of new lots and shipments, including lot number(s) tested and

comparison of results to the acceptability criteria

REFERENCES 1) Clinical and Laboratory Standards Institute. Evaluation of Matrix Effects; Approved Guideline. 3rd ed. CLSI Document EP14-A3.

Clinical and Laboratory Standards Institute, Wayne, PA; 2014 2) Clinical and Laboratory Standards Institute. Verification of Comparability of Patient Results within One Healthcare System: Approved

Guideline (Interim Revision). CLSI Document EP31-A-IR. Clinical and Laboratory Standards Institute, Wayne, PA; 2012. 3) Miller WG, Myers GL, Rej R. Why commutability matters. Clin Chem. 2006;52:553-554 4) Clinical and Laboratory Standards Institute. User Evaluation of Between-Reagent Lot Variation; Approved Guideline. CLSI Document

EP26-A. Clinical and Laboratory Standards Institute. Wayne, PA; 2013.

COM.30500 Reagent Kit Components Phase II

If there are multiple components of a reagent kit, the laboratory uses components of reagent kits only within the kit lot unless otherwise specified by the manufacturer.

Evidence of Compliance: ✓ Written policy defining allowable exceptions for mixing kit components from different lots

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1252(e)]

INSTRUMENTS AND EQUIPMENT

A variety of instruments and equipment are used to support the performance of analytical procedures. Examples of equipment include, but are not limited to centrifuges, microscopes, incubators, heat blocks, refrigerators, freezers, biological safety cabinets, fume hoods, glassware, pipettes, etc. This section contains general requirements that apply to most laboratory sections and types of testing. The laboratory is also responsible for any additional instrument and equipment requirements found in the discipline-specific checklists, as applicable.

INSTRUMENT AND EQUIPMENT MAINTENANCE/FUNCTION CHECKS

Inspector Instructions:

● Sampling of instrument/equipment policies and procedures ● Sampling of function check and performance verification records for instruments/

equipment ● Sampling of instrument/equipment maintenance logs and repair records

● Instrument/equipment records (promptly retrievable)

COM.30525 Maintenance and Function Checks - Waived Tests Phase II

For waived tests, the laboratory follows manufacturer's instructions for instrument and equipment maintenance and function checks.

Evidence of Compliance: ✓ Written procedure consistent with manufacturer's instructions for each waived test AND

33 of 52

All Common Checklist 08.17.2016

✓ Records for instrument/equipment maintenance and function checks as required by the manufacturer

The remaining checklist requirements in the INSTRUMENT AND EQUIPMENT MAINTENANCE/FUNCTION CHECK section do not apply to waived tests.

**REVISED** 07/28/2015 COM.30550 Instrument/Equipment Performance Verification Phase II

The performance of all instruments and equipment is verified upon installation and after major maintenance or service to ensure that they run according to expectations.

NOTE: Performance verification is necessary after repairs or replacement of critical components of an instrument or item of equipment.

Evidence of Compliance: ✓ Written procedure for performance verification AND ✓ Records of performance verification

REFERENCES 1) Clinical and Laboratory Standards Institute. Laboratory Instrument Implementation, Verification, and Maintenance; Approved

Guideline. CLSI Document GP31-A. Clinical and Laboratory Standards Institute, Wayne, PA; 2009.

**REVISED** 07/28/2015 COM.30575 Instrument Operation Phase II

There are written procedures for start-up, operation and shutdown of instruments and equipment, as applicable.

NOTE: These procedures must readily be available to the operator in the immediate vicinity of the instrument, and ideally should include a procedure for emergency shutdown and for handling workload during instrument downtime. These may be separate approved procedures or included in the testing procedure for a specific analyte.

COM.30600 Maintenance/Function Checks Phase II

Appropriate maintenance and function checks are performed and records maintained for all instruments (e.g. analyzers) and equipment (e.g. centrifuges) following a defined schedule, at least as frequent as specified by the manufacturer.

NOTE: There must be a schedule and procedure at the instrument for appropriate function checks and maintenance. These may include (but are not limited to) cleaning, electronic, mechanical and operational checks. The procedure and schedule must be as thorough and as frequent as specified by the manufacturer.

Function checks should be designed to detect drift, instability, or malfunction, before the problem is allowed to affect test results.

For equipment that has no standard frequency or requirement for maintenance and function checks, each laboratory should establish a schedule and procedure that reasonably reflects the workload and specifications of its equipment.

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1254] 2) Clinical and Laboratory Standards Institute. Laboratory Instrument Implementation, Verification, and Maintenance; Approved

Guideline. CLSI Document GP31-A. Clinical and Laboratory Standards Institute, Wayne, PA; 2009.

**REVISED** 07/28/2015 COM.30625 Function Check Tolerance Limits Phase II

34 of 52

All Common Checklist 08.17.2016

Tolerance limits for acceptable function are defined for specific instruments and equipment wherever appropriate, with records of action when the limits are exceeded.

NOTE: The defined tolerance limits must follow the manufacturer's specified limits. Function checks must be within the defined tolerance limits prior to use for testing patient samples.

The action related to tests that have an approved Individualized Quality Control Plan (IQCP) must include an assessment of whether further evaluation of the risk assessment and quality control plan is needed based on problems identified (e.g. trending for repeat failures, etc.).

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1254]

COM.30650 Instrument Troubleshooting Phase II

Instructions are provided for minor troubleshooting and repairs of instruments (such as manufacturer's service manual).

COM.30675 Instrument and Equipment Records Phase II

Instrument and equipment maintenance, function check, performance verification, and service and repair records (or copies) are promptly available to, and usable by, the technical staff operating the equipment.

NOTE: Effective utilization of instruments and equipment by the technical staff depends upon the prompt availability of the records (copies are acceptable) to detect trends or malfunctions. Off- site storage, such as with centralized medical maintenance or computer files, is acceptable if the inspector is satisfied that the records can be promptly retrieved.

REFERENCES 1) Clinical and Laboratory Standards Institute. Laboratory Instrument Implementation, Verification, and Maintenance; Approved

Guideline. CLSI Document GP31-A. Clinical and Laboratory Standards Institute, Wayne, PA; 2009.

THERMOMETERS

Inspector Instructions:

● Records of traceability to NIST Standards ● Sampling of verification records for non-certified thermometers ● Sampling of policies and procedures for thermometer verification

COM.30700 Thermometric Standard Device Phase II

An appropriate thermometric standard device of known accuracy (certified to meet NIST Standards or traceable to NIST Standards) is available.

NOTE: Thermometric standard devices must be recalibrated, recertified, or replaced prior to the date of expiration of the guarantee of calibration or they are subject to requirements for non- certified thermometers.

Thermometers should be periodically evaluated for damage (e.g. separation of columns). Thermometers with obvious damage must be rechecked for continued use.

Evidence of Compliance: ✓ Thermometer certificate of accuracy AND ✓ Policy for the use of thermometers after the date of expiration of the guarantee of calibration

and records of recertification

35 of 52

All Common Checklist 08.17.2016

REFERENCES 1) Clinical and Laboratory Standards Institute. Laboratory Instrument Implementation, Verification, and Maintenance; Approved

Guideline. CLSI Document GP31-A. Clinical and Laboratory Standards Institute, Wayne, PA; 2009.

COM.30725 Non-certified Thermometers Phase II

All non-certified thermometers in use are checked against an appropriate thermometric standard device before initial use and as defined by laboratory policy.

NOTE: Non-certified thermometers used in transfusion medicine, including blood-warmer thermometers, must be checked at least annually.

If digital or other displays of temperatures on equipment are used for daily monitoring, the laboratory must verify that the readout is accurate. The display must be checked initially and following manufacturer's instructions.

Evidence of Compliance: ✓ Written procedure defining verification of non-certified thermometers AND ✓ Written policy for rechecking of non-certified thermometers AND ✓ Records of verification

REFERENCES 1) Clinical and Laboratory Standards Institute. Laboratory Instrument Implementation, Verification, and Maintenance; Approved

Guideline. CLSI Document GP31-A. Clinical and Laboratory Standards Institute, Wayne, PA; 2009.

TEMPERATURE-DEPENDENT INSTRUMENTS, EQUIPMENT, AND ENVIRONMENTS

Inspector Instructions:

● Sampling of temperature logs (refrigerator, freezer, water bath, heat block, incubator ambient, etc.)

**REVISED** 08/17/2016 COM.30750 Temperature Checks Phase II

Temperatures are checked and recorded each day of use for all temperature-dependent equipment and environments using a calibrated thermometer.

NOTE: Temperature-dependent equipment (e.g. refrigerators, freezers, incubators) containing reagents and/or patient/client specimens must be monitored daily, as equipment failures could affect accuracy of patient/client test results. Items such as water baths and heat blocks used for procedures need only be checked on days of patient/client testing. For heat blocks or dry baths, thermocouple probes may be used as an alternative method for checking the temperature.

If specific instruments, equipment, kits, or supplies have specified ambient temperature ranges for proper operation, storage, or use, there must be records that the specified ambient temperature is maintained and corrective action taken when tolerance limits are exceeded.

Temperatures may be recorded either manually, or using a recording device or system by: 1) recording the numerical temperature, or 2) placing a mark on a graph that corresponds to a numerical temperature. If temperatures are recorded manually, the identity of the individual recording the temperature(s) must be recorded (initials of the individual are adequate).

If an automated (including remote) temperature monitoring system is used instead of manual temperature monitoring, laboratory personnel must have ongoing immediate access to the

36 of 52

All Common Checklist 08.17.2016

temperature data so that appropriate corrective action can be taken if a temperature is outside of the acceptable range. System records must demonstrate daily functionality of the system.

If a minimum/maximum thermometer is used to perform continuous monitoring of temperatures between daily temperature readings or following a laboratory downtime (e.g. laboratory closure for weekend or holiday), both the low and high temperatures must be recorded. To ensure correct temperature readings, the minimum/maximum thermometer device must be reset prior to the monitoring period.

A frost-free freezer may be used to store reagents and controls provided that the function of these materials is not compromised. Storage conditions must remain within the specifications of the manufacturer of the reagent or control. Temperatures may be recorded using a continuous monitoring system or a maximum/minimum thermometer. Thermal containers within the freezer may be used.

Patient samples may be stored in a frost-free freezer only if protected from thawing. The laboratory must maintain records showing that the temperatures stay within the defined range.

REFERENCES 1) Clinical and Laboratory Standards Institute. Laboratory Instrument Implementation, Verification, and Maintenance; Approved

Guideline. CLSI Document GP31-A. Clinical and Laboratory Standards Institute, Wayne, PA; 2009.

COM.30775 Temperature Range Phase II

Acceptable ranges have been defined for all temperature-dependent equipment and environments (including test-dependent ambient temperature) in accordance with the manufacturer's instructions.

Evidence of Compliance: ✓ Temperature log or record with defined acceptable range

COM.30800 Temperature Corrective Action Phase II

There is evidence of corrective action taken if acceptable temperature ranges for temperature-dependent equipment and environmental temperatures are exceeded, including evaluation for adverse effects.

NOTE: If acceptable temperature ranges are exceeded, stored reagents, controls, calibrators, etc. must be checked to confirm the accuracy or quality of the material before use, with records maintained. The check should follow a defined procedure.

TEST METHOD VALIDATION AND VERIFICATION

Inspector Instructions:

● Policies and procedures for the introduction of new tests, methods, or instruments ● Sampling of assay validation and verification studies with emphasis on tests

introduced in the past two years ● Sampling of patient reports for laboratory-developed assays

● Which laboratory tests or instruments have been implemented in the past two years, particularly those that are not FDA-cleared/approved?

● Do you follow the manufacturer's instructions exactly for all FDA-cleared/approved diagnostic kits or devices?

● How does your laboratory validate or verify assay performance prior to test implementation?

37 of 52

All Common Checklist 08.17.2016

● How has your laboratory verified or established the reportable range (AMR) for the test?

● How does your laboratory establish or verify reference intervals? ● How does your laboratory validate clinical claims made by the laboratory about

LDTs?

● Select a representative sampling of new assays introduced during the past two years for evaluation.

● In addition, select assays in place greater than two years if recurrent problems have been identified in proficiency testing results, quality control, competency assessment, or physician complaints for evaluation, as applicable.

● Review evaluation statements and supporting validation or verification data to confirm that studies were performed using an adequate number of cases, data were reviewed, and evaluations were approved prior to implementation. If the data showed discordances or unacceptable variations, investigate how they were resolved.

**NEW** 08/17/2016 COM.30980 Waived Test Implementation and Approval Phase II

For each waived test, the laboratory follows manufacturer's instructions for the introduction of the instrument or device and there are records that the test(s) is approved for use by the laboratory director, or designee meeting CAP director qualifications, prior to use in patient testing.

NOTE: Waived testing must be performed following the manufacturer's instructions as written. If the laboratory modifies a waived test, the checklist requirements for high complexity testing apply, including the requirements for validation of the method performance specifications.

The laboratory director's signature on the written test procedure may be used to show approval of the test for use in patient testing.

Evidence of Compliance: ✓ Records of test approval

**REVISED** 08/17/2016 COM.40000 Method Validation and Verification Approval - Nonwaived Tests Phase II

For each nonwaived test, there is an evaluation of the test method validation or verification study (accuracy, precision, etc.) signed by the laboratory director, or designee meeting CAP director qualifications, prior to use in patient testing to confirm the acceptability of the data and approve each nonwaived test for clinical use.

NOTE: This checklist requirement is applicable only to nonwaived tests implemented after June 15, 2009; however, all nonwaived tests must have records of completed analytical validation or verification, regardless of their implementation date.

If multiple identical instruments or devices are in use, there must be records showing that the method performance specifications have been separately verified for each test and instrument or device. The evaluation and data must clearly support the decision to approve the test for clinical use.

The evaluation must include: 1) a written assessment of each component of the validation or verification study, including the acceptability of the data; 2) a signed approval statement, such as, "I have reviewed the verification (or validation) data for accuracy, precision, reportable range, and reference range studies (insert other components, as required) for the (insert instrument/test name), and the performance of the method is considered acceptable for patient testing." If data

38 of 52

All Common Checklist 08.17.2016

include discordant results, there must be a record of the discordance and investigation of any impact on the approval of the test for clinical use.

For an FDA-cleared/approved test, an evaluation of the verification data must address analytical performance specifications, including analytical accuracy, precision, interferences, and reportable range, as applicable.

In addition, for modified FDA-cleared/approved tests or LDTs, the evaluation must address analytical sensitivity, analytical specificity, and any other parameter that is considered important, to assure the analytical performance of a test (e.g. specimen stability, reagent stability, linearity, carryover, and cross-contamination, etc. as appropriate and applicable).

If the laboratory makes clinical claims about its tests, the evaluation must address the validation of these claims.

See the Method Performance Specifications section for details concerning validation and verification.

Evidence of Compliance: ✓ Evaluation of validation and verification studies with review and approval

REFERENCES 1) Lawrence Jennings, Vivianna M. Van Deerlin, Margaret L. Gulley (2009) Recommended Principles and Practices for Validating

Clinical Molecular Pathology Tests. Archives of Pathology & Laboratory Medicine: Vol. 133, No. 5, pp. 743-755 2) Lacbawan FL, Weck KE, Kant JA, Feldman GL, Schrijver; Biological and Molecular Genetic Resource Committee of the College of

American Pathologists. Verification of performance specifications of a molecular test: cystic fibrosis carrier testing using the Luminex liquid bead array. Arch Pathol Lab Med. 2012. Jan; 136(1):14-9.

METHOD PERFORMANCE SPECIFICATIONS - NONWAIVED TESTS

NOTE: This subsection on METHOD PERFORMANCE SPECIFICATIONS does not apply to waived tests.

ANALYTICAL VALIDATION/VERIFICATION

Laboratories are required to perform analytical validation or verification of each nonwaived test, method, or instrument system before use in patient testing, regardless of when it was first introduced by the laboratory, including instruments of the same make and model and temporary replacement (loaner) instruments. There is no exception for analytical validation or verification of tests introduced prior to a specific date. The laboratory must have data for the validation or verification of the applicable method performance specifications and retain the records as long as the method is in use and for at least two years after discontinuation.

The method performance specifications must be validated or verified in the location in which patient testing will be performed. If an instrument is moved, the laboratory must verify the method performance specifications (i.e. accuracy, precision, reportable range) after the move to ensure that the test system was not affected by the relocation process or any changes due to the new environment (e.g. temperature, humidity, reagent storage conditions, etc.). The laboratory must follow manufacturer's instructions for instrument set up, maintenance, and system verification.

Not all method performance specifications apply to qualitative tests. For qualitative tests, the laboratory must establish/verify the method performance specifications that are applicable and clinically relevant.

LABORATORIES SUBJECT TO US REGULATIONS:

● For unmodified FDA-cleared or approved tests, the laboratory may use information from manufacturers, or published literature, but the laboratory must verify such outside information on accuracy, precision and reportable range.

● For tests that are not FDA-cleared or approved (including tests developed in-house), or for FDA-cleared/approved tests modified by the laboratory, the laboratory must establish accuracy, precision, analytical sensitivity, interferences, analytical specificity, and reportable range, as

39 of 52

All Common Checklist 08.17.2016

applicable; data on interferences may be obtained from manufacturers or published literature, as applicable.

LABORATORIES NOT SUBJECT TO US REGULATIONS:

● The laboratory must verify or establish analytical accuracy, precision, analytical sensitivity, analytical specificity (interfering substances) and reportable range for each test. Laboratories may use information from manufacturers, published literature, or studies performed in other laboratories, but should verify such outside information, whenever practical.

LABORATORY-DEVELOPED TESTS: For the purposes of interpreting the checklist requirements, a laboratory-developed test (LDT) is defined as follows: A test used in patient management that has both of the following features:

1. The test is performed by the clinical laboratory in which the test was developed wholly or in part; AND

2. The test is neither FDA-cleared nor FDA-approved.

COM.40100 Intermittent Testing Phase II

When a test is put back into production, the following requirements must be met:

1. PT or alternative assessment performed within 30 days prior to restarting patient testing

2. Method performance specifications verified, as applicable, within 30 days prior to restarting patient testing

3. Competency assessed for analysts within 12 months prior to restarting patient testing

NOTE: This requirement applies to tests that are taken out of production for a time (for example, seasonal testing for influenza). A test is considered to be taken out of production when (1) patient testing is not offered AND (2) PT or alternative assessment, as applicable, is suspended. It does not apply to situations where a proficiency testing challenge is not performed due to a temporary, short-term situation, such as a reagent back order or an instrument breakdown. In those situations, the laboratory must perform alternative assessment for that testing event.

The laboratory should have written procedures for putting intermittent tests into production.

For tests for which PT is required by CAP, if a PT challenge is not offered during the 30-day period prior to restarting patient testing, the laboratory may perform an alternative assessment of the test. The laboratory must participate in the next scheduled PT event, if the Laboratory Accreditation Program requires external PT for that analyte.

**REVISED** 07/28/2015 COM.40200 LDT and Modified Test List Phase I

The laboratory maintains a list of laboratory-developed tests (LDTs) and modified FDA- cleared/approved tests implemented by the laboratory.

NOTE: The list must include tests developed in-house, and for laboratories subject to US regulations, tests using analyte-specific reagents (ASRs), and FDA-cleared/approved tests that have been modified by the laboratory.

A form is available on the CAP website that may be used for maintaining this list and can be downloaded from the CAP website (http://www.cap.org) through e-LAB Solutions Suite.

COM.40250 Manufacturer's Instructions Phase II

40 of 52

All Common Checklist 08.17.2016

The laboratory follows manufacturer's instructions or provides validation records if the test has been modified.

NOTE: Following manufacturer's instructions includes performing quality control, calibration, calibration verification, and related functions as applicable to the scope of testing. Reagents, fluids, and disposable materials supplied by the laboratory must meet the specifications in the instructions.

If the laboratory modifies manufacturer's instructions, the test is no longer an FDA-cleared/ approved test, and the modification(s) must be validated by the laboratory. Changes in the specimen type or collection device are examples of common modifications (see "modification of manufacturer's instructions" in the Definition of Terms). Additional requirements for validation/ verification may be found in the discipline-specific checklists.

For waived and moderately complex tests, if manufacturer instructions are modified, requirements for high complexity testing apply.

Evidence of Compliance: ✓ Validation records of established performance specifications (accuracy, precision, analytical

sensitivity, analytical specificity, interferences, reference range, and reportable range) of any test that has been modified.

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1252]

**REVISED** 08/17/2016 COM.40300 Accuracy and Precision Phase II

The laboratory verifies or establishes analytical accuracy and precision for each test.

NOTE: Accuracy is established or verified by comparing results to a definitive or reference method, or an established comparative method. Use of matrix-appropriate reference materials, patient specimens (altered or unaltered), or other commutable materials with known concentrations or activities may be used to establish or verify accuracy. The use of routine quality control materials or calibrators that were used to calibrate the method is not appropriate.

Precision is established by repeat measurement of samples at varying concentrations or activities within-run and between-run over a period of time.

Evidence of Compliance: ✓ Written procedure for determining method performance characteristics, including accuracy

and precision AND ✓ Records of verification or establishment of accuracy and precision for each test

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24) [42CFR493.1253] 2) Clinical and Laboratory Standards Institute. Measurement Procedure Comparison and Bias Estimation Using Patient Samples;

Approved Guideline. 3 rd

ed. CLSI Document EP09-A3. Clinical and Laboratory Standards Institute, Wayne, PA; 2013. 3) Voss EM, et al. Determining acceptability of blood glucose meters. Statistical methods for determining error. Lab Med.

1996;27:601-606 4) Clinical and Laboratory Standards Institute (CLSI). Statistical Quality Control for Quantitative Measurement Procedures: Principles

and Definitions; Approved Guideline—Third Edition. CLSI document C24-A3 (ISBN 1-56238-613-1). Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2006

5) Clinical and Laboratory Standards Institute. Preliminary Evaluation of Quantitative Clinical Laboratory Methods; Approved Guideline.

3 rd

ed. CLSI Document EP10-A3-AMD. Clinical and Laboratory Standards Institute, Wayne, PA; 2014. 6) Clinical and Laboratory Standards Institute. A Framework for Using CLSI Documents to Evaluate Clinical Laboratory Measurement

Procedures. 2 nd

ed. CLSI report EP19-Ed2. Clinical and Laboratory Standards Institute, Wayne, PA; 2015. 7) Clinical and Laboratory Standards Institute. Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline.

3 rd

ed. CLSI document EP05-A3. Clinical and Laboratory Standards Institute, Wayne, PA; 2014. 8) Clinical and Laboratory Standards Institute. Evaluation of the Commutability of Processed Samples; Approved Guideline. 3rd ed.

CLSI document EP14-A3. Clinical and Laboratory Standards Institute, Wayne, PA; 2014.

41 of 52

All Common Checklist 08.17.2016

**NEW** 07/28/2015 COM.40350 Accuracy - Modified FDA-cleared/approved and LDTs Phase I

For modified FDA-cleared/approved and laboratory-developed tests (LDTs), validation of analytical accuracy includes testing with an appropriate number of samples.

NOTE: An appropriate number of samples is defined as the following: ● For quantitative tests, a minimum of 20 samples with analyte concentrations distributed

across the analytical measurement range should be used. Proportionate mixtures of samples may be used to supplement the study population.

● For qualitative tests, a minimum of 20 samples, including positive, negative, and low- positive samples with concentrations near the lower level of detection should be used; equivocal samples should not be used.

● For certain methods that test multiple analytes (e.g. next-generation sequencing, FISH, HPLC, GC-MS, MALDI-TOF, etc.), analytic accuracy may be established for each method (not necessarily each analyte), as appropriate.

If the laboratory uses fewer samples, the laboratory director must record the criteria used to determine the appropriateness of the sample size. In many cases, a validation study with more samples is desirable.

For modified FDA-cleared/approved tests and LDTs in use prior to July 31, 2016, for which limited validation studies are recorded, ongoing data supporting acceptable test performance may be used to meet the above minimum sample requirement, unless the laboratory director has recorded the criteria used to determine the acceptability of a smaller sample size. Examples of such ongoing data include records of proficiency testing, alternative performance assessment, and quality control.

This checklist requirement does not apply to LDTs that employ the following methods: ● Manual microscopy (e.g. histopathologic and cytologic interpretation, microscopic

examination of blood or body fluids, Gram stains) ● Conventional microbiologic cultures and disc/broth/tube susceptibility studies

Evidence of Compliance: ✓ Written procedure for determining method performance characteristics, including accuracy

AND ✓ Records of establishment of analytical accuracy for each test

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24) [42CFR493.1253] 2) Clinical and Laboratory Standards Institute. Measurement Procedure Comparison and Bias Estimation Using Patient Samples;

Approved Guideline. 3 rd

ed. CLSI Document EP09-A3. Clinical and Laboratory Standards Institute, Wayne, PA; 2013. 3) Clinical and Laboratory Standards Institute. Statistical Quality Control for Quantitative Measurement Procedures: Principles and

Definitions; Approved Guideline. 3 rd

ed. CLSI Document C24-A3. Clinical and Laboratory Standards Institute, Wayne, PA; 2006. 4) Clinical and Laboratory Standards Institute. Preliminary Evaluation of Quantitative Clinical Laboratory Methods; Approved Guideline.

3 rd

ed. CLSI Document EP10-A3-AMD. Clinical and Laboratory Standards Institute, Wayne, PA; 2014. 5) Clinical and Laboratory Standards Institute. A Framework for Using CLSI Documents to Evaluate Clinical Laboratory Measurement

Procedures. 2 nd

ed. CLSI report EP19-Ed2. Clinical and Laboratory Standards Institute, Wayne, PA; 2015.

COM.40400 Analytical Sensitivity Phase II

For modified FDA-cleared/approved tests or laboratory-developed tests (LDTs), the laboratory establishes the analytical sensitivity (lower detection limit) of each assay, as applicable.

Evidence of Compliance: ✓ Written procedure for determining method performance characteristics, including analytical

sensitivity AND ✓ Records of establishment of analytical sensitivity for each assay

REFERENCES

42 of 52

All Common Checklist 08.17.2016

1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):[42CFR493.1253(b)(2)]

2) Clinical and Laboratory Standards Institute (CLSI). Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline—Second Edition. CLSI document EP17-A2 (ISBN 1-56238-795-2). Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 2500, Wayne, Pennsylvania 19087-1898 USA, 2012.

3) Clinical and Laboratory Standards Institute. A Framework for Using CLSI Documents to Evaluate Clinical Laboratory Measurement

Procedures. 2 nd

ed. CLSI report EP19-Ed2. Clinical and Laboratory Standards Institute, Wayne, PA; 2015.

COM.40450 Analytical Specificity Phase II

For modified FDA-cleared/approved tests or laboratory-developed tests (LDTs), the results of each validation study include a sufficient number of samples to establish the test's analytical specificity.

NOTE: The analytical specificity refers to the ability of a test or procedure to correctly identify or quantify an entity in the presence of interfering or cross-reactive substances that might be expected to be present. Laboratories are encouraged to review the published literature for guidance and provided confidence intervals to estimated performance characteristics.

Evidence of Compliance: ✓ Records of validation studies and published references used to establish analytical specificity

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1253(b)(2)] 2) Clinical and Laboratory Standards Institute (CLSI). Interference Testing in Clinical Chemistry; Approved Guideline—Second Edition.

CLSI document EP7-A2 (ISBN 1-56238-584-4). Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2005.

3) Clinical and Laboratory Standards Institute. A Framework for Using CLSI Documents to Evaluate Clinical Laboratory Measurement

Procedures. 2 nd

ed. CLSI report EP19-Ed2. Clinical and Laboratory Standards Institute, Wayne, PA; 2015.

COM.40500 Analytical Interferences Phase II

The laboratory understands the analytical interferences for each test, and has an appropriate plan of action when they are present.

NOTE: Interfering substances may pose a significant problem to the clinical laboratory and healthcare providers who may be misled by laboratory results that do not reflect patient clinical status. The laboratory must be aware of common interferences by performing studies or referencing studies performed elsewhere (such as by the instrument-reagent manufacturer).

Evidence of Compliance: ✓ Written procedure for determining method performance characteristics, including analytical

interferences AND ✓ Document listing known interferences for each test and plan of action when they are present

REFERENCES 1) Clinical and Laboratory Standards Institute. Interference Testing in Clinical Chemistry; Approved Guideline. 2nd ed. CLSI document

EP07-A2. Clinical and Laboratory Standards Institute, Wayne, PA, 2005. 2) Ho C-H. The hemostatic effect of packed red cell transfusion in patients with anemia. Transfusion. 1998;38:1011-1014 3) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Clinical laboratory improvement amendments

of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1253]

COM.40600 Reportable Range Phase II

The reportable range is verified or established for each analytical procedure before implementation.

NOTE: The analytical measurement range (AMR) is the range of analyte values that a method can directly measure on the specimen without any dilution or concentration.

43 of 52

All Common Checklist 08.17.2016

Expanded definitions and details of the AMR are provided in some of the section-specific checklists (e.g. Chemistry). Verification of the AMR may not apply to certain assays (for example, in immunology and coagulation).

The limits of the AMR are based on meeting accuracy and precision requirements such as the minimal limit of quantification or sensitivity, when applicable. In some cases, clinically relevant limits may be narrower than the potential analytical range, and the clinically relevant limit would be used as the limit of the reportable range.

Evidence of Compliance: ✓ Written policy for determining method performance characteristics, including reportable range

AND ✓ Records of verification or establishment of reportable ranges for each test

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 1992(Feb 28): [42CFR493.1253] 2) Clinical and Laboratory Standards Institute. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory;

Approved Guideline. 3 rd

ed. CLSI document EP28-A3c. Clinical and Laboratory Standards Institute, Wayne, PA; 2008.

**NEW** 07/28/2015 COM.40610 Calibration and Quality Control Procedures Phase II

For laboratory-developed tests and modified FDA-cleared/approved tests, the laboratory defines written procedures for calibration and quality control based on the studies performed to evaluate the method performance specifications.

NOTE: The procedures must define the frequency, number, and concentration of calibrators and controls to be used.

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24) [42CFR493.1253(b)(3)]

**REVISED** 08/17/2016 COM.40620 Body Fluid Analysis Phase II

Methods for body fluid analysis have been validated or verified and metrics for interpretation have been established.

NOTE: This requirement applies directly to body fluid testing that the laboratory offers as a routine, orderable test. If the test is routinely performed on the fluid, there must be a written procedure. The requirement COM.40000 for a method validation or verification approval applies. Method performance specifications for blood specimens may be used for body fluids if the laboratory can reasonably exclude the existence of matrix interferences affecting the latter either by reference in the procedure manual to published literature or by evaluation for interferences due to matrix effects by performing an appropriate study (e.g. a dilution study using admixtures of samples, spiking samples, further dilution). Alternative performance assessment is required (COM.01500) and may be performed using clinical assessment by chart review.

The reference ranges must be defined and reported with the results, unless the concentration of the analyte is reported in comparison to its concentration in a contemporaneously collected blood specimen. If the result is to be interpreted by comparison to the patient's blood, serum, or plasma, such results must be accompanied by an appropriate comment such as, "The reference range and other method performance specifications are unavailable for this body fluid. Comparison of this result with the concentration in the blood, serum, or plasma is recommended." Reference range citations from the manufacturer's insert or published literature citations may be used to determine the range (COM.50000). However, reference ranges have not been published for many body fluid analytes, and obtaining normal fluids to establish a reference range may not be feasible.

44 of 52

All Common Checklist 08.17.2016

A request for a test on a body fluid specimen that is not listed on the laboratory's test menu that requires clearance by the section director or designee is considered a clinically unique specimen, rather than a routine, orderable test. Typically, these specimens are submitted due to an unusual clinical concern in a specific patient or situation (e.g. pathologic states where the analyte is not normally found in the fluid type) and it may not be possible to establish a comparative metric. In such cases, the result must be accompanied by a comment such as, "The reference range and other method performance specifications have not been established for this body fluid. The test result must be integrated into the clinical context for interpretation."

Evidence of Compliance: ✓ Records of validation or verification studies with review and approval AND ✓ Records of reference range study OR records of verification of manufacturer's stated range

or published literature OR other methods approved by the laboratory/section director

REFERENCES 1) Clinical and Laboratory Standards Institute (CLSI). Body Fluids in Clinical Chemistry; Approved Guideline. CLSI document C49-

A (ISBN 1-56238-638-7). Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2007

2) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1253]

3) Block DR, Algeciras-Schimnich A. Body fluid analysis; clinical utility and applicability of published studies to guide interpretation of today's laboratory testing in serous fluids. Crit Rev Clin Lab Sci. 2013 July-Oct;50(4-5):107-24. doi: 10.3109/10408363.2013.844679.

**REVISED** 08/17/2016 COM.40630 LDT Reporting Phase I

Reports for laboratory-developed tests (LDTs) contain a statement that the assay was developed by the laboratory.

NOTE: This requirement does not apply to traditional methods, such as manual microscopy, conventional microbiologic cultures, conventional cytogenetics, and manual hematology and immunology tests.

Requirements for reports are given in the Results Reporting sections of the checklists. Laboratories subject to US regulations often include an LDT disclaimer as follows: "This test was developed and its performance characteristics determined by <insert laboratory/company name>. It has not been cleared or approved by the FDA. The laboratory is regulated under CLIA as qualified to perform high-complexity testing. This test is used for clinical purposes. It should not be regarded as investigational or for research."

The report should also contain a brief description of the method and any performance characteristics needed for clinical use, unless the information is readily available to the clinician in an equivalent form (e.g. test catalog).

REFERENCES 1) CLSI. Establishing Molecular Testing in Clinical Laboratory Environments: CLSI document MM19-A (ISBN 1-56238-773-1). Clinical

and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2011

**REVISED** 07/28/2015 COM.40640 LDT Clinical Claims Validation Phase II

All clinical claims made by the laboratory are validated for the following types of tests:

● Laboratory-developed tests (LDTs) ● FDA-cleared/approved tests for which the laboratory makes a clinical claim(s) not

included in manufacturer instructions

NOTE: Clinical claims include statements about a test's diagnostic sensitivity and specificity, ability to predict the risk of a disease or condition, clinical usefulness, or cost-effectiveness. Clinical claims may be found on the test report or in other information distributed by the laboratory (websites, test catalogues, newsletters, memoranda, advertisements, etc.).

45 of 52

All Common Checklist 08.17.2016

Laboratories are not required to make clinical claims about a test, but any claims made by the laboratory must be validated.

In order to adequately support a claim about diagnostic sensitivity and specificity and/or ability to predict risk of a disease or condition, the laboratory must perform a clinical validation study, unless the clinical validity of the test is documented in peer-reviewed literature or textbooks. The clinical validation study must include at least 20 samples and must include both positive and negative samples. If the laboratory uses fewer samples, the laboratory director must record the criteria used to determine the appropriateness of the sample size.

Evidence of Compliance: ✓ Records of clinical studies performed by the laboratory OR peer-reviewed literature that

reasonably substantiates all claims made by the laboratory about a test

REFERENCES 1) Clinical and Laboratory Standards Institute. A Framework for Using CLSI Documents to Evaluate Clinical Laboratory Measurement

Procedures. 2 nd

ed. CLSI report EP19-Ed2. Clinical and Laboratory Standards Institute, Wayne, PA; 2015.

**REVISED** 07/28/2015 COM.40700 Method Performance Specifications Availability Phase II

For current test methods, the laboratory makes the following available to clients and the inspection team upon request:

● Summary of the analytical performance specifications for each method, validated or verified by the laboratory to include analytical accuracy, precision, analytical sensitivity, analytical specificity (test method interferences), reference range, and reportable range, as applicable; and

● Supporting data for clinical performance claims, if applicable, validated or verified by the laboratory or obtained from peer-reviewed literature.

NOTE: For the purposes of providing this information to clients, it may be presented in a summary format referring to the supporting data, statistics, and published studies, as appropriate. Clients include healthcare entities, other laboratories, and licensed independent practitioners. This requirement does not apply to patients or their authorized representatives.

The laboratory may require clients to treat the data as confidential and not to use such proprietary information for its own test development or share such data with any other party except as required by law. The CAP inspection team is instructed to treat all such data as confidential and to review them solely for accreditation purposes.

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24):7163 [42CFR493.1291(e)]

COM.40800 Analytical Methodology Changes Phase II

If the laboratory changes its analytical methodology so that test results or their interpretations may be SIGNIFICANTLY different, the change is explained to clients.

NOTE: This requirement can be accomplished in any of several different ways, depending on local circumstances. Some methods include directed mailings, laboratory newsletters or part of the test report itself.

Evidence of Compliance: ✓ Records such as directed mailings, laboratory newsletters or comment on the patient report

advising of the change

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 2003(Jan 24):7163 [42CFR493.1291(e)]

46 of 52

All Common Checklist 08.17.2016

REFERENCE INTERVALS

COM.50000 Reference Intervals Established/Verified Phase II

The laboratory establishes or verifies its reference intervals (normal values).

NOTE: Reference intervals are important to allow a clinician to assess patient results against an appropriate population. The reference range must be established or verified for each analyte and specimen source (e.g. blood, urine, cerebrospinal fluid), when appropriate. For example, a reference interval can be verified by testing samples from 20 healthy representative individuals; if no more than two results fall outside the proposed reference interval, that interval can be considered verified for the population studied (refer to CLSI guideline EP28-A3c, reference below).

If a formal reference interval study is not possible or practical, then the laboratory should carefully evaluate the use of published data for its own reference ranges, and retain records of this evaluation. For many analytes (e.g. therapeutic drugs and CSF total protein), literature references or a manufacturer's package insert information may be appropriate.

Evidence of Compliance: ✓ Record of reference range study OR records of verification of manufacturer's stated range

when reference range study is not practical (e.g. unavailable normal population) OR other methods approved by the laboratory/section director

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 1992(Feb 28):7164 [42CFR493.1253] 2) Van der Meulen EA, et al. Use of small-sample-based reference limits on a group basis. Clin Chem. 1994;40:1698-1702 3) Clinical and Laboratory Standards Institute (CLSI). Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory

- Approved Guideline- Third Edition - CLSI Document EP28-A3c. (ISBN 1-56238-682-4) Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, PA, 19087-1898, USA, 2010.

COM.50100 Reference Interval Evaluation Phase II

The laboratory evaluates the appropriateness of its reference intervals and takes corrective action if necessary.

NOTE: Criteria for evaluation of reference intervals include: 1. Introduction of a new analyte to the test repertoire 2. Change of analytic methodology 3. Change in patient population

If it is determined that the range is no longer appropriate for the patient population, corrective action must be taken.

Evidence of Compliance: ✓ Records of evaluation and corrective action, if indicated

REFERENCES 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement

amendments of 1988; final rule. Fed Register. 1992(Feb 28):7164 [42CFR493.1253] 2) Van der Meulen EA, et al. Use of small-sample-based reference limits on a group basis. Clin Chem. 1994;40:1698-1702 3) Clinical and Laboratory Standards Institute (CLSI). Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory

- Approved Guideline- Third Edition - CLSI Document EP28-A3c. (ISBN 1-56238-682-4) Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, PA, 19087-1898, USA, 2010.

4) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24): [42CFR493.1282(b)(1)(iii)]

47 of 52

All Common Checklist 08.17.2016

INDIVIDUALIZED QUALITY CONTROL PLAN (IQCP)

This section applies to laboratories using an IQCP approved by the laboratory director for nonwaived testing to reduce external control analysis to a frequency less than the limits defined in the CLIA regulations and CAP checklists. Note that development of an IQCP only impacts quality control requirements. All other checklist requirements remain unchanged and applicable.

This section does not apply to tests where an IQCP was implemented, but the type and frequency of quality control defined in the plan already meets or exceeds minimum quality control requirements defined in the CLIA regulations and CAP checklist requirements. Quality control requirements in other sections of the All Common Checklist and discipline-specific checklists will be used for inspection in those situations.

If a laboratory is located in a state that does not accept IQCP as an option for reducing the frequency of external quality control, the laboratory must follow the state regulations and perform external daily quality control following the frequency defined in the state regulations and CAP checklists.

Eligibility for use of an IQCP is limited to testing meeting all of the following criteria:

● Nonwaived tests that employ an internal (electronic/procedural/built-in) quality control system ❍ Exception: Microbiology media and reagents used for microbial identification and susceptibility

testing may implement an IQCP as defined in the checklist ● Tests performed in specialties other than Anatomic Pathology and Cytopathology

❍ Exception: If an Anatomic Pathology or Cytopathology test can be assigned to a different CMS subspecialty, it may qualify.

Laboratories may develop their own model for designing an IQCP or use the Clinical and Laboratory Standards Institute (CLSI) Guideline EP23-A, the Centers for Medicare and Medicaid Services resources, a manufacturer protocol, or use other commercially available products.

NOTE: A laboratory may not implement an IQCP that allows for quality control to be performed less frequently than indicated in the manufacturer's instructions. The components of the quality control plan must meet regulatory and CAP accreditation requirements and be in compliance with the manufacturer instructions, at minimum.

The following table contains information on quality control related requirements that are eligible for IQCP under CAP's accreditation program and the related CLIA regulations (Title 42CFR):

48 of 52

All Common Checklist 08.17.2016

QC Requirement Related CLIA Regulation (Title 42 in the CFR)

Quantitative testing includes two levels of quality control at different concentrations at least daily

493.1256(d)(3)(i)

Qualitative testing includes positive and negative controls at least daily 493.1256(d)(3)(ii) Semi-quantitative testing with graded or titered results include a control material of graded or titered reactivity at least daily

493.1256(d)(iii)

Tests with an extraction phase include two levels of quality control, one of which goes through the extraction phase at least daily

493.1256(d)(iv)

Tests with molecular amplification procedures include two control materials and a control material capable of detecting inhibition, as applicable

493.1256(d)(3)(v)

Each new lot and shipment of reagents, disks, stains, antisera, and identification systems (systems using two or more substrates or two or more reagents, or a combination) are checked for positive and negative reactivity, as well as graded reactivity, if applicable

493.1256(e)(1)

Each batch of commercially prepared media is checked for sterility, if sterility is required for testing, before or concurrent with initial use

493.1256(e)(4)(i)

Each batch of commercially prepared media is checked for its ability to support growth and, as appropriate, select or inhibit specific organisms or produce a biochemical response, before or concurrent with initial use

493.1256(e)(4)(ii)

Antimicrobial susceptibility tests include appropriate control organism(s) to check the procedure each day tests are performed

493.1261(b)(1)

Antifungal susceptibility tests include appropriate control organism(s) to check the procedure each day tests are performed

493.1263(b)(2)

Blood gas testing includes one control (combination of low and high values used) every eight hours of patient testing and one control sample each time a specimen is tested unless the method is auto-calibrated every 30 minutes

493.1267(b)(c)

Automated coagulation testing includes two levels of controls every 8 hours of patient testing and when a reagent is changed

493.1269(b)

Inspector Instructions:

● Policies and procedures for the implementation of an IQCP ● Sampling of IQCP records with emphasis on tests with IQCPs implemented in the

past two years for the following:

❍ Risk assessment, including laboratory data and summary of findings ❍ Manufacturer's product inserts and published data ❍ Signed quality control plan defining all aspects monitored ❍ Ongoing quality assessment monitoring records for QC, instrument/

equipment maintenance and function checks, complaints, errors, and corrective actions

❍ Reassessment of quality control plan at least annually

If an IQCP is in use, the laboratory is required to complete the following forms provided by the CAP and provide a copy to the inspector:

● List of Individualized Quality Control Plans by Instrument/Device/Test - identifies all tests, instruments and devices using an IQCP

● Individualized Quality Control Plan Summary - provides key information on implementation and monitoring of the IQCP

Use the completed forms to identify an appropriate sampling of records to review.

Sampling of IQCP records to include: 1) a mix of manual and automated tests using an IQCP in the last two years; 2) a mix of tests using an IQCP where there are variations in the testing environment, personnel, multiple testing devices, etc.; and 3) a mix of tests using an IQCP

49 of 52

All Common Checklist 08.17.2016

that have had recurring problems with proficiency testing, quality control, instrument failure, errors, or physician complaints.

● What sources of information are used to perform a risk assessment prior to IQCP implementation?

● What steps were taken to ensure that tests already in place with internal quality control processes for daily QC (e.g. equivalent quality control) are in compliance with the IQCP requirements?

● How is the ongoing assessment of the IQCP quality control plan performed? ● How are physician complaints about the validity of test results for tests using an IQCP

handled? ● What is the process to review errors for tests using an IQCP? ● Have there been any adverse patient events related to a test using an IQCP?

● Review an IQCP and confirm that all elements of the quality control plan are being monitored

● Review an IQCP that is shared by more than one testing location to verify that the risk assessment included an evaluation of each site or location and that each location is monitored as defined in the IQCP

● Review IQCP risk assessment summary, supporting data and approved quality control plan to confirm that the plan was approved by the laboratory director prior to implementation

● Review ongoing quality assessment data and error/incident logs to confirm that effective corrective actions have been taken

**NEW** 07/28/2015 COM.50200 IQCP Test List/Summary Phase II

The laboratory has identified all tests using an IQCP and completed the CAP's forms for laboratories using an individualized quality control plan.

NOTE: The CAP requires the completion of the following forms if an IQCP is in use by the laboratory: List of Individualized Quality Control Plans and the Individualized Quality Control Plan Summary. The forms may be downloaded from the IQCP Resources page (IQCP Resources) on the CAP website.

The use of the forms is required, even if standardized forms and templates are used by the laboratory. The laboratory is responsible for maintaining the accuracy of the data on the form and for providing a current copy to the inspector during an on-site CAP inspection. The form is intended to be used as an inspector tool and does not meet the checklist requirements for documenting the IQCP risk assessment or quality control plan.

REFERENCES 1) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Brochure #12. CLIA Individualized Quality

Control Plan, Considerations When Deciding to Develop an IQCP. November 2014. http://www.cms.gov/Regulations-and-Guidance/ Legislation/CLIA/Downloads/CLIAbrochure12.pdf

**NEW** 07/28/2015 COM.50300 Risk Assessment Phase II

The IQCP for a test/device/instrument includes a risk assessment to evaluate potential sources of error to include all of the following:

● Preanalytic, analytic, and postanalytic phases of the testing process

50 of 52

All Common Checklist 08.17.2016

● Intended medical uses of the test and impact if inaccurate results are reported (clinical risk)

● Components of the tests including reagents, environment, specimen, testing personnel, and test system

● Variations in the components based on use of the tests (e.g. use in different environments, by different personnel, or multiple identical devices)

● Data from the laboratory's own environment, instrument/equipment performance, and testing personnel

● Manufacturer's instructions and recommendations

NOTE: The risk assessment must include a process to identify the sources of potential failures and errors for a testing process, and evaluate frequency and impact of those failures and sources of error.

The laboratory director must consider the laboratory's clinical and legal responsibilities for providing accurate and reliable patient test results. Published data and information may be used to supplement the risk assessment, but is not a substitute for the laboratory's own studies and evaluation. The laboratory must involve a representative sample of testing personnel in the process of conducting the risk assessment. It is not necessary for all personnel to be involved.

The risk assessment for laboratories with multiple identical devices must show that an evaluation was performed if there are differences in testing personnel or environments where testing is performed, with customization of the quality control plan, as needed.

The QC study performed to assess the performance and stability of the tests must support the QC frequency and elements defined in the laboratory's quality control plan. The study must include data representing, at a minimum the maximum interval between runs of external quality control. The laboratory may use historical data during the risk assessment for tests already in place.

For affiliated laboratories (e.g. systems) with integrated procedures, each accredited laboratory must have its own IQCP approved by the laboratory director. There must be records demonstrating that risks specific to the site were evaluated involving a representative sample of local testing personnel to conduct the risk assessment and that laboratory-specific QC data were used in the study to support the defined frequency of quality control. Laboratories may use data from other sites to supplement risk assessments and to support their findings.

REFERENCES 1) Clinical and Laboratory Standards Institute. Laboratory Quality Control Based on Risk Management; Approved Guideline CLSI

Document EP23-A. Clinical and Laboratory Standards Institute. Wayne, PA; 2011. 2) Centers for Medicare and Medicaid Services (CMS). Individual Quality Control Plan (IQCP) for Clinical Laboratory Improvement

Amendments (CLIA) laboratory nonwaived testing. http://www.cms.gov/Regulations-and-Guidance/Legislation/CLIA/Downloads/ IQCP-announcement-letter-for-CLIA-CoC-and-PPM-labs.pdf Accessed January 12, 2016.

3) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Brochure #13. CLIA Individualized Quality Control Plan, What is an IQCP? November 2014. http://www.cms.gov/Regulations-and-Guidance/Legislation/CLIA/Downloads/ CLIAbrochure13.pdf

4) Nichols JH. Laboratory quality control based on risk management. Ann Saudi Med 2011; 31:223-228. 5) Yundt-Pacheco J, Parvin CA. Validating the performance of QC procedures. Clin Lab Med 2013; 33:75-88. 6) US Department of Health and Human Services. Developing an IQCP - A Step-by-Step Guide. https://wwwn.cdc.gov/clia/Documents/

IQCP%20Layout.pdf Accessed January 12, 2016.

**NEW** 07/28/2015 COM.50400 Quality Control Plan Approval Phase II

The IQCP includes a written quality control plan approved by the laboratory director prior to implementation.

NOTE: The quality control plan may be part of a test procedure or be a separate written plan. As an efficiency, a single plan may address multiple tests performed on one device. A separate, quality control plan approved by the laboratory director must be in place for each laboratory with a separate CAP and CLIA number.

REFERENCES

51 of 52

All Common Checklist 08.17.2016

1) Clinical and Laboratory Standards Institute. Laboratory Quality Control Based on Risk Management; Approved Guideline CLSI Document EP23-A. Clinical and Laboratory Standards Institute. Wayne, PA; 2011.

2) Centers for Medicare and Medicaid Services (CMS). Individual Quality Control Plan (IQCP) for Clinical Laboratory Improvement Amendments (CLIA) laboratory nonwaived testing. http://www.cms.gov/Regulations-and-Guidance/Legislation/CLIA/Downloads/ IQCP-announcement-letter-for-CLIA-CoC-and-PPM-labs.pdf Accessed January 12, 2016.

3) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Brochure #13. CLIA Individualized Quality Control Plan, What is an IQCP? November 2014. http://www.cms.gov/Regulations-and-Guidance/Legislation/CLIA/Downloads/ CLIAbrochure13.pdf Accessed January 12, 2016.

**NEW/REVISED** 08/17/2016 COM.50500 Quality Control Plan Elements Phase II

The individualized quality control plan must define all aspects monitored based on the potential errors identified during the risk assessment, including the following parameters as applicable:

● The number, type (external and internal quality control systems), and frequency of quality control

● Criteria for acceptable performance ● Monitoring of the testing environment and reagents ● Specimen quality ● Instrument calibration, maintenance, and function checks ● Training and competency of testing personnel ● Provisions for multiple identical devices and variation for uses covered under one

IQCP

NOTE: The components of the quality control plan must meet regulatory and CAP accreditation requirements and be in compliance with the manufacturer instructions, at minimum. The quality control plan must control the quality of the test process and ensure accurate and reliable test results.

External control material samples must be analyzed at least every 31 days and with new lots and shipments of reagents or more frequently if indicated in the manufacturer's instructions. The provision to perform external controls at least every 31 days does not apply to microbiology media and identification systems.

REFERENCES 1) Clinical and Laboratory Standards Institute. Laboratory Quality Control Based on Risk Management; Approved Guideline CLSI

Document EP23-A. Clinical and Laboratory Standards Institute. Wayne, PA; 2011. 2) Centers for Medicare and Medicaid Services (CMS). Individual Quality Control Plan (IQCP) for Clinical Laboratory Improvement

Amendments (CLIA) laboratory nonwaived testing. http://www.cms.gov/Regulations-and-Guidance/Legislation/CLIA/Downloads/ IQCP-announcement-letter-for-CLIA-CoC-and-PPM-labs.pdf Accessed January 12, 2016.

3) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Brochure #13. CLIA Individualized Quality Control Plan, What is an IQCP? November 2014. http://www.cms.gov/Regulations-and-Guidance/Legislation/CLIA/Downloads/ CLIAbrochure13.pdf Accessed January 12, 2016.

4) US Department of Health and Human Services. Developing an IQCP - A Step-by-Step Guide. https://wwwn.cdc.gov/clia/Documents/ IQCP%20Layout.pdf Accessed January 12, 2016.

**NEW** 07/28/2015 COM.50600 Quality Assessment Monitoring Phase II

Ongoing quality assessment monitoring is performed by the laboratory to ensure that the quality control plan is effective in mitigating the identified risks for the IQCP and includes the following:

● Review of quality control and instrument/equipment maintenance and function check data at least monthly

● Evaluation of errors relating to preanalytic, analytic and post analytic phases of the testing process

● Review of complaints from clinicians and other healthcare providers regarding the quality of testing to confirm the clinical efficacy of testing, and

● Evaluation of corrective actions taken if problems are identified

52 of 52

All Common Checklist 08.17.2016

● Reapproval of the quality control plan by the laboratory director or designee at least annually

NOTE: If ongoing assessments identify failures in one or more components of the quality control plan, the laboratory must investigate the cause and consider if modifications are needed to the quality control plan to mitigate potential risk.

REFERENCES 1) Clinical and Laboratory Standards Institute. Laboratory Quality Control Based on Risk Management; Approved Guideline CLSI

Document EP23-A. Clinical and Laboratory Standards Institute. Wayne, PA; 2011. 2) Centers for Medicare and Medicaid Services (CMS). Individual Quality Control Plan (IQCP) for Clinical Laboratory Improvement

Amendments (CLIA) laboratory nonwaived testing. http://www.cms.gov/Regulations-and-Guidance/Legislation/CLIA/Downloads/ IQCP-announcement-letter-for-CLIA-CoC-and-PPM-labs.pdf Accessed January 12, 2016.

3) Department of Health and Human Services, Centers for Medicare & Medicaid Services. Brochure #13. CLIA Individualized Quality Control Plan, What is an IQCP? November 2014. http://www.cms.gov/Regulations-and-Guidance/Legislation/CLIA/Downloads/ CLIAbrochure13.pdf Accessed January 12, 2016.

4) US Department of Health and Human Services. Developing an IQCP - A Step-by-Step Guide. https://wwwn.cdc.gov/clia/Documents/ IQCP%20Layout.pdf Accessed January 12, 2016.

Al Matar The Alere BinaxNOW ® Influenza A B Test Training Quiz.docx

The Alere BinaxNOW ® Influenza A & B

Competency Test

St. Luke’s Magic Valley Point of Care

Employee Name: __________________________________ Date: _____________________

1. Which temperature should BinaxNOW ® Influenza A & B is storage?

a. 4 °C.

b. -24 °C.

c. 37 °C

d. 24 °C.

2. Which temperature can be nasal washes held?

a. – 4 °C.

b. 37 °C.

c. 4 °C.

d. 24 °C.

3. In nasopharyngeal swabs, how many milliliter of sample should be romove?

a. 1.0 mL.

b. 0.5 mL.

c. 1.5 mL.

d. 2.0 mL.

4. Which is true about the result interpretation?

a. For a negative sample, the blue control line turns a pink to purple color. No other line appears.

b. For a Flu A positive sample, the blue control line turns a pink to purple color. A second line appears above it in the top third of the window.

c. For a Flu B positive control, the Blue control Line turns a pink to purple color. A second line appears above it in the middle third of the window.

d. For a Flu A and Flu B positive control, No control line appears.

5. Eluted swab samples can be held at 36-46 °F (2-8°C) for up to __________ before testing in the BinaxNOW ® Test.

a. 8 hours.

b. 18 hours.

c. 1 hour.

d. 24 hours.

Answer Key:

1. D.

2. C.

3. B

4. A.

5. D.