Inhaled Nitric Oxide Delivery Systems for Mechanically Ventilated and Nonintubated Patients: A Review

Stefano Gianni, Ryan W Carroll, Robert M Kacmarek,y and Lorenzo Berra

Introduction

Nitric Oxide Generators and Delivery Systems

Cylinder-Based Systems

Electricity-Generated NO

Chemical NO Generator

NO-Releasing Solutions

NO-Releasing Nanoparticles

Summary

Nitric oxide (NO) is a biologically active molecule approved for the treatment of pulmonary hypertension

in newborn patients. Commercially available NO delivery systems use pressurized cylinders as the source

of NO and a sensor to control the concentrations of NO and nitrogen dioxide (NO2) delivered. Cylinder-

based delivery systems are safe and widely used around the world, but they are bulky, expensive, and

reliant on a robust supply chain. In the past few years, novel NO generators and delivery systems have

been developed to overcome these limitations. Electric NO generators produce NO from ambient air

using high-voltage electrical discharge to ionize air, which leads to the formation of NO, NO2, and ozone

(O3). A scavenging system is incorporated to reduce the concentration of the toxic byproducts generated

in this type of system. NO can also be generated by the reduction of NO2 by ascorbic acid or released

from liquid solutions or solid nanoparticles. The development of easy-to-use, safe, and portable NO deliv-

ery systems may enable the delivery of NO in the out-patient setting or at home. Furthermore, non–cyl-

inder-based NO generators reduce the cost of NO production and storage and may therefore make NO

delivery feasible in low-resource settings. Here we review commercially available systems that can gener-

ate and administer inhalable NO. Key words: nitric oxide; nitrogen dioxide; respiratory therapy; inhala- tion. [Respir Care 2021;66(6):1021–1028. © 2021 Daedalus Enterprises]

Introduction

Nitric oxide (NO) gas is a molecule that plays a pivotal

role in many physiological processes. In the cardiovascu-

lar system, NO is released by the endothelial cell and acts

as a vascular smooth muscle relaxant to induce systemic

and pulmonary vasodilation.1 Because NO is rapidly inac-

tivated by the reaction with oxyhemoglobin, inhaled NO

acts as a pure pulmonary vasodilator with negligible sys-

temic hemodynamic effects.2 In 1999, the U.S. Food and

Drug Administration (FDA) approved gaseous inhaled

NO for the treatment of “term and near-term (> 34 weeks)

neonates with hypoxic respiratory failure associated with

clinical or echocardiographic evidence of pulmonary

hypertension where it improves oxygenation and reduces

the need for extracorporeal membrane oxygenation.”3,4 In

addition to its pulmonary vasodilator effect, NO displays

a mild bronchodilator effect5 as well as antithrombotic6-7

and anti-inflammatory effects.8

In adult patients with ARDS, NO reduces pulmonary

artery pressure and improves ventilation-perfusion match-

ing.9-11 Moreover, NO reduces pulmonary vascular resist-

ance in patients with COPD12,13 and chronic pulmonary

hypertension.14,15

RESPIRATORY CARE � JUNE 2021 VOL 66 NO 6 1021

NO is normally produced by healthy endothelium and, in

the presence of intravascular hemolysis, is rapidly inacti-

vated by its reaction with free hemoglobin,16 inducing

systemic and pulmonary vasoconstriction. The potent pul-

monary vasodilator effect of inhaled NO has been demon-

strated to treat the systemic and pulmonary vasocon-

striction induced by free hemoglobin.17 Cardiopulmonary

bypass induces intravascular hemolysis, releasing hemoglo-

bin and subsequently scavenging endogenous NO, leading

to intrarenal vasoconstriction and causing post-cardiac sur-

gery kidney injury. A recent randomized controlled trial

demonstrated that the perioperative administration of NO

reduces the incidence of acute kidney injury in cardiac sur-

gery subjects.18 Furthermore, NO displays a broad antimi-

crobial effect against bacteria,19 fungi,20 and viruses,21

especially if administered at high concentrations (> 100 parts per million [ppm]).22-25 The antimicrobial properties of

high-concentration NO are now being tested in several clini-

cal trials (ClinicalTrials.gov registration NCT02498535,

NCT04685720, NCT04606407, NCT04305457).

The increased spectrum of pathologies treated with NO

gas has urged innovators and industries to find solutions to

produce versatile, inexpensive, and readily available NO

delivery systems for in-hospital, out-patient, and home use.

While offering novel solutions, companies are also focusing

on the safety of delivery systems. The main safety issue of

high-dose inhaled NO delivery is the generation of nitrogen

dioxide (NO2). When NO and oxygen (O2) mix in a gaseous

environment, NO2 is formed spontaneously by the following

reaction: NO þ NO þ O2 ! NO2 þ NO2. The speed of NO2 production has a first-order dependence on O2 concen-

tration and a second-order dependence on NO concentration

with a time constant of 1:1960:11 � 10�11=PPM � s.26 Inhaled NO reacts with water and generates nitric acid

(pH 1.0), causing chemical lesions to the bronchial mu-

cosa. Therefore, to avoid pulmonary injury, the Amer-

ican Conference of Governmental Industrial Hygienists

established a limit of 3 ppm NO2 in an 8-h time weighted av-

erage and 5 ppm as a short-term exposure limit.27 More con-

servatively, the National Institute for Occupational Safety

and Health recommends a short-term exposure limit of 1

ppm.27 It is therefore imperative that any NO delivery system

reduces the delivered NO2 levels below the safety thresholds.

In this review we will describe the systems currently devel-

oped to administer inhaled NO to patients (Table 1).

Nitric Oxide Generators and Delivery Systems

Cylinder-Based Systems

Cylinder-based systems represent the vast majority of com-

mercially available NO delivery systems. (Table 2). Pressurized

cylinders contain various concentrations of NO buffered with

an inert gas (ie, not containing oxygen) such as nitrogen (N2) to

avoid the generation of NO2 within the cylinder.

The N2/NO pressurized cylinder is connected to a flow-

regulated injector that delivers a set NO flow and therefore

the target NO concentration into the inspiratory arm of a re-

spiratory circuit. A central processing unit continuously

measures the gas flow delivered to the patient (ie, from a ven-

tilator) and regulates the NO flow from the pressurized cylin-

der in real time. The targeted NO concentration is obtained

by maintaining the ratio between the NO flow from the pres-

surized cylinder and the total gas flow delivered to the

patient. The NO administration can be continuous throughout

the respiratory cycle or synchronized with ventilation. All

commercially available NO delivery systems are able to con-

tinuously measure NO, NO2, and O2 concentrations through

an electrochemical sensor cell.28 The schema of a cylinder-

based NO delivery system is summarized in Figure 1.

Specific cylinder-based delivery systems that should be

highlighted are the INODD (Novoteris, Garden Grove,

California) and the INOpulse (Bellerophon Therapeutics,

Warren, New Jersey). The INODD is, up to now, the only

commercially available cylinder-based system designed to

deliver high-dose NO (160 ppm) and is currently employed

in a trial testing the safety and efficacy of 160 ppm NO on

multi-resistant bacterial lung infection and in subjects with

coronavirus disease-2019 (COVID-19) (NCT03331445).

The INOpulse is a portable system designed to deliver

NO in an ambulatory/home environment using a 0.16-L

mini-cylinder as the NO source. The INOpulse delivers a

set pulsed volume of NO at the beginning of each breath

via a special nasal cannula connected to the device.

Consequently, the INOpulse delivers a constant dose of NO

over time (expressed in mg/kg/h)12 that is independent of minute ventilation and inspiratory flow. In spontaneously

breathing patients, the administration of a pulsed NO dose

has 2 potential advantages. First, a pulsed NO dose can be

Drs Gianni and Berra are affiliated with the Department of Anaesthesia,

Critical Care and Pain Medicine, Massachusetts General Hospital,

Boston, Massachusetts. Dr Carroll is affiliated with the Department of

Pediatrics, Massachusetts General Hospital, Boston, Massachusetts.

Dr Kacmarek was affiliated with Respiratory Care Department,

Massachusetts General Hospital, Boston, Massachusetts. All authors are

affiliated with Harvard Medical School, Boston, Massachusetts.

Dr Gianni has disclosed no conflicts of interest. Dr Carroll has disclosed

a relationship with Unitaid. Dr Kacmarek had disclosed relationships

with Medtronic, Orange Med, Medtronic, and Venner Medical. Dr Berra

has disclosed relationships with the NIH, iNO Therapeutics, Praxair,

Masimo, and the Mercatus Center of George Mason University.

†Deceased.

Correspondence: Lorenzo Berra MD, Massachusetts General Hospital,

Harvard Medical School, 55 Fruit St, Boston, MA 02141. E-mail:

[email protected].

DOI: 10.4187/respcare.08856

INHALED NITRIC OXIDE DELIVERY SYSTEMS

1022 RESPIRATORY CARE � JUNE 2021 VOL 66 NO 6

precisely delivered using a nasal cannula without the need

for a snug, tight-fitting mask, improving patient comfort in

an ambulatory or home setting. Second, the brief pulse of

NO minimizes the amount of drug dispensed and reduces

the environmental exhaust.29

Clinical trials are ongoing to test the efficacy of pulsed

inhaled NO for the treatment of pulmonary hypertension

associated with interstitial lung disease (NCT02734953),

COPD,12 and sarcoidosis (NCT03727451). A phase 3 clini-

cal trial (NCT02725372) testing the effect of pulsed inhaled

NO in subjects with symptomatic pulmonary arterial hyper-

tension was stopped for futility. Furthermore, there are 2

phase 3 clinical trials under way to test whether the admin-

istration of NO through INOpulse can reduce mortality and

acute respiratory failure in spontaneously breathing sub-

jects with COVID-19 (NCT04421508 and NCT04388683).

Table 2. Technical Characteristics of Cylinder-Based NO Delivery Systems

System What Is Set NO Delivery Concentration Range,

ppm NO Injector Flow Sensor

INOMax (Mallinkrodt) NO concentration Synchronized, continuous 0–80 Reusable Reusable

NOxBOXi (Praxair) NO concentration Synchronized, continuous 0–80 Single use Single use

SoKINOX (Air Liquide) NO concentration Synchronized, continuous 0–80 Single use Single use

AeroNOx 2.0

(International

Biomedical)

NO flow Continuous Depends on NO/N2 cyl-

inder concentration

and set flow

Single use Single use

NO-A (EKU) NO concentration Synchronized, continuous 0–100 Single use Single use

NOXTEC 1000 (ITC) NO concentration Continuous 0–100 Single use Single use

NOx plus (Zysense) NO flow Continuous Depends on NO/N2 cyl-

inder concentration

and set flow

Reusable Reusable

The operator can set the desired NO concentration or the NO flow necessary to achieve it. The NO delivery can be continuous (ie, the system maintains a stable NO concentration throughout the respira-

tory cycle) or synchronized with ventilation (ie, the system delivers an NO flow proportional to the inspiratory flow delivered by the ventilator). All the delivery systems use an electrochemical cell sensor

to measure NO and NO2 concentration.

NO ¼ nitric oxide NO2 ¼ nitrogen dioxide ppm ¼ parts per million

Table 1. Summary of the Available Nitric Oxide Delivery Systems

Manufacturer Product Market Release

Cylinder-based systems

Mallinckrodt INOMax DSIR North America, Europe, Australia

Praxair NOxBOXi North America, Europe

EKU NO-A Europe, Asia

Air Liquide Healthcare SoKINOX North America, Europe, Australia

Ingenierı́a y Técnicas Clı́nicas NOXTEC Europe

Novoteris INODD Research use only (NCT02498535)

International Biomedical Aeronox North America, Europe

Zysense NOx plus North America, Europe, South America

Bellerophon INOpulse Research use only (NCT03267108, NCT04421508, NCT03727451, NCT02652429)

Electric NO systems

Third Pole Therapeutics eNOX-200 Under development

Beyond Air LungFit Research use only (NCT04685720, NCT04397692, NCT04606407)

Chemical-based NO systems

Vero Biotech Genosyl Ds United States of America

Nu-Med plus Clinical Unit Under FDA approval

NO-releasing solutions

SaNOtize NORS Research use only (NCT04163978, NCT04443868, NCT04337918)

Nanoparticle NO technology

NMB Therapeutics NanoNOx Under development

Vast Therapeutics TBD Under development

INHALED NITRIC OXIDE DELIVERY SYSTEMS

RESPIRATORY CARE � JUNE 2021 VOL 66 NO 6 1023

Systems that utilize cylinder-based delivery are the more

widely available NO delivery systems, with > 450,000 patients treated worldwide.30 They are safe and reliable and

can deliver a wide range of NO concentrations by simply

changing the flow from the NO cylinder. There are 2 main

disadvantages of cylinder-based delivery systems: (1) pres-

surized gas cylinders require an extensive supply chain, in-

ventory at the hospital, and trained personnel to connect

cylinders to ventilators or noninvasive respiratory systems;

(2) cylinder-based NO therapy is expensive for institutions

themselves,31 with the average charge for providing NO ther-

apy for 5 d to a newborn patient with persistent pulmonary

hypertension being � $14,000.30 These factors greatly reduce the feasibility of NO gas in several settings, such as

referral hospitals, out-patient clinics, in the home, and in

low-resource settings.

Electricity-Generated NO

Electricity-generated NO systems use high-voltage elec-

trical discharges to generate NO from air or another gas

mixture (Fig. 2). The application of an electric potential

between 2 electrodes ionizes air (or other gas mixture),

causing an electron flow from the cathode to the anode.

Gas Inlet

Gas Pump Flow sensor

Ca(OH)� scavenger 0.22-�m filter

Sparking chamber

Gas Outlet

Gas Pump Flow sensor

Ca(OH)� scavenger 0.22-�m filter

Fig. 2. Schema of a nitric oxide electric generator. Ca(OH)2 ¼ calcium hydroxide.

Set NO concentration

CPU Flow sensor

Sensors

NO NO� O�

Ventilator Patient

N O

ta nk

Flow regulator

Fig. 1. Schema of a cylinder-based NO delivery system. NO ¼ nitric oxide; NO2 ¼ nitrogen dioxide; CPU ¼ central processing unit.

INHALED NITRIC OXIDE DELIVERY SYSTEMS

1024 RESPIRATORY CARE � JUNE 2021 VOL 66 NO 6

The frequent electron collisions due to the high current

raise the temperature between the 2 electrodes up to 10,000

Kelvin, causing the dissociation of N2 and O2 into a plasma

state and the subsequent generation of NO, NO2, and ozone

(O3). Air flows used are in the range of 0.9–4.5 L/min. The

NO and NO2 production increase with frequency of

the sparking pulse, increase with greater atmospheric pres-

sure, and decrease with increasing air flow.32,33

Because NO electric generators produce both NO and

NO2, it is imperative to minimize the NO2/NO ratio to

achieve more efficient and safer NO production. One of the

main determinants of the NO2/NO ratio is the temperature of

the plasma: the higher the plasma temperature, the lower the

NO2/NO ratio. The plasma temperature is dependent on the

capacitance of the pulse-forming capacitor (increasing with

the increase in capacitance, up to 20nF), and on the current

running through the sparker (plasma temperature increases

from 900 to 1130 K as the current increases from 600 to

1,400 ampere).34 The NO2/NO ratio is also dependent on the

inter-electrode gap: increasing the inter-electrode gap from

1.0 to 7.5 mm decreases the NO2/NO ratio from 0.12 to 0.09.

Namihira et al35 measured how NO production is affected by

variations of O2 + N2 gas mixture using an NO electric gen-

erator with the following settings: 1.5 L/min of total gas flow

and pulse frequency of 3 Hz. The highest NO production

was observed with an FIO2 of 0.26. The NO2/NO ratio was

also found to be dependent on the electrode material, ranging

from 0.13 for tungsten to 0.05 for iridium, across an array of

metals (ie, tungsten > carbon > nickel > iridium). To administer pure NO, it is critical to remove the

byproducts such as NO2, ozone (O3), and brass particles

emitted by the electrode during arc discharges. Because

NO2 is continuously produced whenever oxygen and NO

are mixed, a system that removes NO2 from the outlet of

the NO generator is always necessary, and 2 mechanisms

can be used: the catalytic conversion of NO2 into NO, and

the selective adsorption of NO2.

There are 4 primary methods to convert NO2 into NO.

One method is to use a molybdenum wire33,36 or a mol-

ybdenum powder filter that can convert NO2 into NO

when heated to � 870 Kelvin through the reaction 3NO2 þ Mo ! 3NO þ MoO3, Namihira et al33 reported a decrease of NO2 concentration from 138 ppm to 48

ppm accompanied by an increase of NO concentration

from 455 ppm to 540 ppm when employing this reaction.

Another method utilizes ascorbic acid loaded on silica

gel pellets (see Chemical NO Generators). A third

method is to have barium oxide (BaO) react with NO2,

producing NO according to the following reaction:

3NO2 þ BaO ! BaðNO3Þ2 þ NO. The fourth method involves calcium hydroxide (Ca(OH)2), which is capable

of selectively reacting with NO2 though the following

reaction37: CaðOHÞ2 þ NO2 þ NO ! CaðNO2Þ2 þ H2O. Yu et al38 reported a reduction of NO2 levels between

70% and 90% using a scavenger containing Ca(OH)2,

when using both in-line and offline electric plasma

generators.

An electrical discharge can also produce ozone (O3) as a

potential toxic by-product. O3 can be detected in the output

arm of the generator and ranges from 10 ppm32 to 18 parts

per billion (ppb).38 The U.S. Environmental Protection

Agency dictates that O3 levels must be kept below 0.07

ppm (70 ppb). O3 is removed (< 0.1 ppb) by bubbling the gas through water32 or by using a Ca(OH)2 filter.

38

Brass and platinum nanoparticles can be generated from

the etching of the electrode material during electric dis-

charges and may be present in the unfiltered gas. A high-ef-

ficiency particulate (HEPA) air filter can remove the metal

particles from the NO generated by electrical discharge.39

The generation of NO from air using pulsed electrical

discharges is a growing field of research, and a handful of

companies are building NO generators and delivery sys-

tems using this technology. The FDA approval process for

these devices is ongoing. The opportunity to generate NO

from air without the need for expensive and bulky gas cyl-

inders is promising and could make NO widely available in

underrepresented regions, both domestically and abroad.

Electric NO generators, when compared to cylinder-

based delivery systems, have 2 potential limitations: NO

production and safety. Because the NO production

decreases when the air flowing through the sparker

increases, the NO generated may not be sufficient to main-

tain the desired NO concentration (eg, in the setting of a

patient receiving NO through a mechanical ventilator deliv-

ering high minute ventilation). Another limitation of the

electric NO generators is the higher risk of safety issues

compared to cylinder-based delivery systems. In case of a

system malfunction (ie, damage to the HEPA filter or to the

Ca(OH)2 scavenger), the device may deliver unmeasured

toxic products such as metal brass particles or ozone (O3).

Chemical NO Generator

In 2011 Lovich et al40 described the generation of NO

from the reduction of NO2 using ascorbic acid. The authors

developed a cartridge composed of a thermoplastic struc-

ture wherein powdered ascorbic acid and silica gel are

adhered to the inside of a solid tube. The pores in the silica

gel provide a matrix in which ascorbic acid and NO2 react

at high efficiency. NO2 flows through the inlet of this car-

tridge and is directed to the tube margins, where it per-

meates radially through the wall of a tube of ascorbic

acid/silica gel particles imbedded in the thermoplastic

structure, ultimately yielding NO. The authors reported the

complete conversion of NO2 (80 ppm) into NO with an out-

put NO2 concentration < 1 ppm (0.6–0.8 ppm using 1 car- tridge with FIO2 of 1.0).

INHALED NITRIC OXIDE DELIVERY SYSTEMS

RESPIRATORY CARE � JUNE 2021 VOL 66 NO 6 1025

In a subsequent manuscript, Lovich et al41 described an

NO generator and delivery system based on this concept.

NO generation follows a 2-step process: the first step is the

generation of NO2 from the vaporization of liquid dinitro-

gen tetroxide (N2O4), and the second step is the conversion

of NO2 to NO using the previously described ascorbic acid

cartridge (Fig. 3).

Liquid N2O4 is contained in a 2-mL evaporation chamber;

the walls of this chamber are stainless steel and transmits

heat to the contained liquid N2O4 to promote the formation

of gaseous NO2. Then the NO2 flows through a capillary

tube of defined cross-sectional area and length and is mixed

with air or nitrogen before entering the reactor cartridge and

being converted to purified NO. Consequently, NO produc-

tion is regulated only by the evaporation chamber tempera-

ture and by cross-sectional area and length of the capillary

tube linking the N2O4 reservoir to the manifold. For a fixed

resistance and for a fixed carrier stream flow, the pressure

and the generated NO2 concentration are directly propor-

tional. The authors reported that the natural logarithm of the

NO2 concentration in the carrier stream is linearly related to

the inverse of the temperature.

At the time of this writing, the only FDA-approved NO

delivery system that uses this technology is the Genosyl DS

(Vero Biotech, Atlanta, Georgia).42 This system can deliver

NO at a concentration of 20 ppm with NO2 concentration

< 2 ppm. The main advantage of this device is the reduced device size and weight compared to cylinder-based delivery

systems. Despite being more portable, the Genosyl DS

shares the same drawback of the cylinder-based delivery

systems: the need for an extensive production and supply

chain to allow widespread use.

NO-Releasing Solutions

NO-releasing solutions are liquid solutions that, when

exposed to specific conditions, can produce and release

NO. Stenzler et al43 patented the use of a solution composed

of an NO-releasing compound (such as sodium nitrite) and

a water- or saline-based solution. The production and

release of NO is dependent on the pH of the solution: if the

pH is > 4, the production of NO is negligible; with pH val- ues < 4, the production of NO increases in a pH-dependent manner. Consequently, the solution is kept in a “dormant”

state at a pH > 4, which allows it to be easily prepared, stored, and transported without losing any appreciable

amount of NO gas. Then, for administration, the solution is

activated by adding citric acid monohydrate, driving the

pH < 4.44 Interestingly, the NO produced (generating NO2 that, reacting with the aqueous solution produced nitric

acid) keeps the pH of the solution < 4, auto-maintaining the NO release.

The NO production from the described NO-releasing so-

lution is dependent on the concentration of sodium nitrite

(NaNO2) and on the pH of the solution. The authors

described an NO production of up to 300 ppm (in the pres-

ence of a continuous air flow of 3 L/min) from 64 mL of a

60 mM NaNO2 solution.

The target clinical use of this technique is the use of

high-dose NO to treat topical infections (eg, bacteria,

fungi, and viruses). In addition, phase 1 and phase 2 trials

are ongoing to test the efficacy of intranasal administra-

tion of a NO-releasing solution for the treatment of

COVID-19 (NCT04337918) and chronic bacterial sinusi-

tis (NCT04163978).

Set a NO concentration

Heated water bath

N�O�

NO� NO

Sensors

NO NO� O�

Patient

Ascorbic acid cartridge

CPU

Fig. 3. Schema of the Genosyl DS NO generator and delivery system. NO ¼ nitric oxide; NO2 ¼ nitrogen dioxide; N2O4 ¼ dinitrogen tetroxide; CPU ¼ central processing unit.

INHALED NITRIC OXIDE DELIVERY SYSTEMS

1026 RESPIRATORY CARE � JUNE 2021 VOL 66 NO 6

NO-Releasing Nanoparticles

A NO-releasing nanoparticle consists of a small particle

(measured in nanometers) that contains either NO or an

inactive NO precursor in a stable form that, when applied

to the target tissue, releases NO in a controlled manner.

Friedman et al45 reported that NO can be efficiently gener-

ated by the conversion of nitrites enclosed in a solid matrix

(derived from trehalose glass or hydrogel/glass composite).

The nitrites are stable in dry form, and NO is released from

the solid matrix when exposed to moisture. The pattern of

NO released from the nanoparticle is dependent on the

composition of the matrix: a glassy matrix releases NO

more rapidly than does a hydrogel/glass matrix.

Like the NO-releasing solutions described above, NO-

releasing nanoparticles have been used to topically admin-

ister antimicrobial doses of NO (eg, treatment of cutaneous

bacterial or mycotic infection). Furthermore, the nanopar-

ticles, due to their small diameter, could potentially be

delivered through an aerosol directly into the lung.46

Further preclinical and clinical trials are needed to address

the safety and the efficacy of this new NO delivery system.

Summary

NO gas is used worldwide for the treatment of neonates

with hypoxic respiratory failure and pulmonary hyperten-

sion. The growing interest in administering high-dose NO

for its antimicrobial effect19-21 poses a major challenge to

the available NO generation and delivery systems both in

terms of cost and safety. In recent years, an effort has been

made, both by researchers and biomedical and pharmaceu-

tical companies, to create NO delivery systems that are

more portable and easier to use. These technological

advancements will allow the use of gaseous NO outside the

hospital (ie, for patients with COPD or primary pulmonary

hypertension). Furthermore, the newly designed NO gener-

ators, by reducing NO production and storage costs, will

make NO administration feasible in low-resource settings.

REFERENCES

1. Ignarro LJ, Buga GM, Wood KS, Byrns RE, Chaudhuri G.

Endothelium-derived relaxing factor produced and released from ar-

tery and vein is nitric oxide. Proc Natl Acad Sci U S A 1987;84

(24):9265-9269.

2. Frostell CG, Blomqvist H, Hedenstierna G, Lundberg J, Zapol WM.

Inhaled nitric oxide selectively reverses human hypoxic pulmonary va-

soconstriction without causing systemic vasodilation. Anesthesiology

1993;78(3):427-435.

3. Barrington KJ, Finer N, Pennaforte T, Altit G. Nitric oxide for respira-

tory failure in infants born at or near term. Cochrane Database Syst

Rev 2017;1:CD000399.

4. Clark RH, Kueser TJ, Walker MW, Southgate WM, Huckaby JL,

Perez JA, et al. Low-dose nitric oxide therapy for persistent pulmonary

hypertension of the newborn. N Engl J Med 2000;342(7):469-474.

5. Bender KA, Alexander JA, Enos JM, Skimming JW. Effects of

inhaled nitric oxide in patients with hypoxemia and pulmonary hyper-

tension after cardiac surgery. Am J Crit Care 1997;6(2):127-131.

6. Gries A, Herr A, Kirsch S, Günther C, Weber S, Szabo G, et al.

Inhaled nitric oxide inhibits platelet-leukocyte interactions in patients

with acute respiratory distress syndrome. Crit Care Med 2003;31

(6):1697-1704.

7. Beghetti M, Sparling C, Cox PN, Stephens D, Adatia I. Inhaled NO

inhibits platelet aggregation and elevates plasma but not intraplatelet

cGMP in healthy human volunteers. Am J Physiol Heart Circ Physiol

2003;285(2):H637-642.

8. Elahi MM, Worner M, Khan JS, Matata BM. Inspired nitric oxide and

modulation of oxidative stress during cardiac surgery. Curr Drug Saf

2009;4(3):188-198.

9. Gerlach H, Keh D, Semmerow A, Busch T, Lewandowski K, Pappert

DM, et al. Dose–response characteristics during long-term inhalation

of nitric oxide in patients with severe acute respiratory distress syn-

drome: a prospective, randomized, controlled study. Am J Respir Crit

Care Med 2003;167(7):1008-1015.

10. Troncy E, Collet JP, Shapiro S, Guimond JG, Blair L, Ducruet T, et al.

Inhaled nitric oxide in acute respiratory distress syndrome: a pilot

randomized controlled study. Am J Respir Crit Care Med 1998;157(5

Pt 1):1483-1488.

11. Taylor RW, Zimmerman JL, Dellinger RP, Straube RC, Criner GJ,

Kenneth DJ, et al. Low-dose inhaled nitric oxide in patients with acute

lung injury: a randomized controlled trial. JAMA 2004;291(13):1603-

1609.

12. Hajian B, de Backer J, Vos W, van Holsbeke C, Ferreira F, Quinn D,

et al. Pulmonary vascular effects of pulsed inhaled nitric oxide in

COPD patients with pulmonary hypertension. Int J Chron Obstruct

Pulmon Dis 2016;11:1533-1541.

13. Vonbank K. Controlled prospective randomised trial on the effects on

pulmonary haemodynamics of the ambulatory long term use of nitric

oxide and oxygen in patients with severe COPD. Thorax 2003;58

(4):289-293.

14. Nathan SD, Flaherty KR, Glassberg MK, Raghu G, Swigris J, Alvarez

R, et al. A randomized, double-blind, placebo-controlled study of

pulsed, inhaled nitric oxide in subjects at risk of pulmonary hyperten-

sion associated with pulmonary fibrosis. Chest 2020;158(2):637-645.

15. Berra L, Rodriguez-Lopez J, Rezoagli E, Yu B, Fisher DF, Semigran

MJ, et al. Electric plasma–generated nitric oxide: hemodynamic

effects in patients with pulmonary hypertension. Am J Respir Crit

Care Med 2016;194(9):1168-1170.

16. Minneci PC, Deans KJ, Zhi H, Yuen PST, Star RA, Banks SM, et al.

Hemolysis-associated endothelial dysfunction mediated by accelerated

NO inactivation by decompartmentalized oxyhemoglobin. J Clin

Invest 2005;115(12):3409-3417.

17. Marrazzo F, Larson G, Sherpa Lama TT, Teggia Droghi M, Joyce M,

Ichinose F, et al. Inhaled nitric oxide prevents systemic and pulmonary

vasoconstriction due to hemoglobin-based oxygen carrier infusion: a

case report. J Crit Care 2019;51:213-216.

18. Lei C, Berra L, Rezoagli E, Yu B, Dong H, Yu S, et al. Nitric oxide

decreases acute kidney injury and stage 3 chronic kidney disease after

cardiac surgery. Am J Respir Crit Care Med 2018;198(10):1279-1287.

19. Miller CC, Hergott CA, Rohan M, Arsenault-Mehta K, D€oring G,

Mehta S. Inhaled nitric oxide decreases the bacterial load in a rat

model of Pseudomonas aeruginosa pneumonia. J Cyst Fibros 2013;12

(6):817-820.

20. Deppisch C, Herrmann G, Graepler-Mainka U, Wirtz H, Heyder S,

Engel C, et al. Gaseous nitric oxide to treat antibiotic resistant bacterial

and fungal lung infections in patients with cystic fibrosis: a phase I

clinical study. Infection 2016;44(4):513-520.

21. Keyaerts E, Vijgen L, Chen L, Maes P, Hedenstierna G, Van Ranst M.

Inhibition of SARS-coronavirus infection in vitro by S-nitroso-N-

INHALED NITRIC OXIDE DELIVERY SYSTEMS

RESPIRATORY CARE � JUNE 2021 VOL 66 NO 6 1027

acetylpenicillamine, a nitric oxide donor compound. Int J Infect Dis

2004;8(4):223-226.

22. Miller C, McMullin B, Ghaffari A, Stenzler A, Pick N, Roscoe D, et al.

Gaseous nitric oxide bactericidal activity retained during intermittent

high-dose short duration exposure. Nitric Oxide 2009;20(1):16-23.

23. Safaee Fakhr B, Wiegand SB, Pinciroli R, Gianni S, Morais CCA,

Ikeda T, et al. High concentrations of nitric oxide inhalation therapy in

pregnant patients with severe coronavirus disease 2019 (COVID-19).

Obstet Gynecol 2020;136(6):1109-1113.

24. Tal A, Greenberg D, Av-Gay Y, Golan-Tripto I, Feinstein Y, Ben-

Shimol S, et al. Nitric oxide inhalations in bronchiolitis: a pilot,

randomized, double-blinded, controlled trial. Pediatr Pulmonol

2018;53(1):95-102.

25. Wiegand SB, Safaee Fakhr B, Carroll RW, Zapol WM, Kacmarek

RM, Berra L. Rescue treatment with high-dose gaseous nitric oxide in

spontaneously breathing patients with severe coronavirus disease

2019. Crit Care Explor 2020;2(11):e0277.

26. Sokol GM, Van Meurs KP, Wright LL, Rivera O, Thorn WJ, Chu PM,

et al. Nitrogen dioxide formation during inhaled nitric oxide therapy.

Clin Chem 1999;45(3):382-387.

27. The National Institute for Occupational Safety and Health. Nitrogen

dioxide. Available at: https://www.cdc.gov/niosh/pel88/10102-44.

html. Accessed April 12, 2021.

28. Privett BJ, Shin JH, Schoenfisch MH. Electrochemical nitric oxide

sensors for physiological measurements. Chem Soc Rev 2010;39

(6):1925-1935.

29. Heinonen E, H€ogman M, Meriläinen P. Theoretical and experimental comparison of constant inspired concentration and pulsed delivery in

NO therapy. Intensive Care Med 2000;26(8):1116-1123.

30. Todd Tzanetos DR, Housley JJ, Barr FE, May WL, Landers CD.

Implementation of an inhaled nitric oxide protocol decreases direct

cost associated with its use. Respir Care 2015;60(5):644-650.

31. Pierce CM, Peters MJ, Cohen G, Goldman AP, Petros AJ. Cost of ni-

tric oxide is exorbitant. BMJ 2002;325(7359):336.

32. Namihira T, Tsukamoto S, Wang D, Katsuki S, Hackam R, Okamoto

K, et al. Production of nitric monoxide using pulsed discharges for a

medical application. IEEE Trans Plasma Sci 2000;28(1):109-114.

33. Namihira T, Katsuki S, Hackam R, Akiyama H, Okamoto K.

Production of nitric oxide using a pulsed arc discharge. IEEE Trans

Plasma Sci 2002;30(5):1993-1998.

34. Hu H, Chen W, Zhang J, Lu X, He J. Influence of plasma temperature

on the concentration of NO produced by pulsed arc discharge. Plasma

Sci Technol 2012;14(3):257-262.

35. Namihira T, Wang D, Tsukamoto S, Katsuki S, Akiyama H. Effect of

gas composition for NOx removal using pulsed power. IEEJ Transfm

1999;119(10):1190-1195.

36. Hu H, Liang H, Li J, Zhao Q, He J. Study on production of inhaled ni-

tric oxide for medical applications by pulsed discharge. IEEE Trans

Plasma Sci 2007;35(3):619-625.

37. Weimann J, Hagenah JU, Motsch J. Reduction in nitrogen dioxide

concentration by soda lime preparations during simulated nitric oxide

inhalation. Br J Anaesth 1997;79(5):641-644.

38. Yu B, Muenster S, Blaesi AH, Bloch DB, Zapol WM. Producing nitric

oxide by pulsed electrical discharge in air for portable inhalation ther-

apy. Sci Transl Med 2015;7(294):294ra107.

39. Yu B, Ferrari M, Schleifer G, Blaesi AH, Wepler M, Zapol WM, et al.

Development of a portable mini-generator to safely produce nitric ox-

ide for the treatment of infants with pulmonary hypertension. Nitric

Oxide 2018;75:70-76.

40. Lovich MA, Bruno NK, Plant CP, Wei AE, Vasquez GB, Johnson BJ,

et al. Use of ultra pure nitric oxide generated by the reduction of nitro-

gen dioxide to reverse pulmonary hypertension in hypoxemic swine.

Nitric Oxide 2011;24(4):204-212.

41. Lovich MA, Fine DH, Denton RJ, Wakim MG, Wei AE, Maslov MY,

et al. Generation of purified nitric oxide from liquid N2O4 for the

treatment of pulmonary hypertension in hypoxemic swine. Nitric

Oxide 2014;37:66-72.

42. Center for Drug Evaluation and Research. Application number:

202860Orig1s000. Available at: https://www.accessdata.fda.gov/

drugsatfda_docs/nda/2019/202860Orig1s000Approv.pdf. Accessed

April 12, 2021.

43. Stenzler A, inventor; Sanotize Research And Development Corp, as-

signee. Compositions and methods for treating diseases or disorders

using extended release nitric oxide releasing solutions. U.S. Patent

Application 20150328256A1. 2015.

44. Stenzler A, Han S, Miller CC, Regev-Shoshani G, inventors.

Preparation and delivery of sustained nitric oxide releasing solutions.

WIPO Application WO2015058127A1. 2015.

45. Friedman AJ, Han G, Navati MS, Chacko M, Gunther L, Alfieri A,

et al. Sustained release nitric oxide releasing nanoparticles: characteri-

zation of a novel delivery platform based on nitrite containing hydro-

gel/glass composites. Nitric Oxide 2008;19(1):12-20.

46. Ding Y, Riediker M. A system to create stable nanoparticle aerosols

from nanopowders. J Vis Exp 2016;2016(113):54414.

INHALED NITRIC OXIDE DELIVERY SYSTEMS

1028 RESPIRATORY CARE � JUNE 2021 VOL 66 NO 6