Taher et al. International Journal of Emergency Medicine (2022) 15:62 https://doi.org/10.1186/s12245-022-00464-y
ORIGINAL RESEARCH
Improving safety and communication for healthcare providers caring for SARS-COV-2 patients Ahmed Taher1,2* , Peter Glazer1,3, Chris Culligan1,3, Stephanie Crump1, Steven Guirguis1, Jennifer Jones1, Alia Dharamsi1,2 and Lucas B. Chartier1,2
Abstract
Background: Decreasing healthcare provider (HCP) exposure to the severe acute respiratory syndrome coronavirus 2 (SARS-COV-2) virus in emergency departments (EDs) is crucial. Approaches include limiting the HCP presence and ensuring sealed isolation rooms, which can result in communication difficulties. This quality improvement (QI) initia- tive aimed to decrease by 50% duration of isolation room door opening and increasing HCP-perceived communica- tion clarity by one point on a five-point Likert scale.
Methods: This was a prospective, multi-stage project with three Plan-Do-Study-Act (PDSA) cycles between May and July 2020: (1) an educational intervention, (2) the introduction of a novel transceiver communication device, and (3) utilizing a clinical champion. Statistical Process Control XbarR charts were used to assess for special cause variation, and two-tailed Mann-Whitney U tests were used for statistical significance between Likert survey means. Qualitative responses underwent thematic analysis.
Results: Observation of 174 patient encounters was completed over 33 days, with 95 meeting the inclusion crite- ria. Door opening decreased from baseline (n=40; mean 72.97%) to PDSA 3 (n=21; mean 1.58%; p<0.0001). HCP- perceived communication clarity improved from baseline (n=36; mean 3.36) to PDSA-3 (n=49; mean 4.21; p<0.001). Survey themes included positive effects on communication and workflow, with some challenges on the integration of the new device into the clinical workflow. HCP-perceived errors, workarounds, and workflow pauses showed signifi- cant improvements.
Conclusion: This QI initiative with a novel transceiver showed significant decreases in isolation room door opening and increases in communication clarity. Future work will expand to operating rooms and intensive care units.
Keywords: Patient safety, Healthcare providers, Safety, SARS-CoV
© The Author(s) 2022. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http:// creat iveco mmons. org/ licen ses/ by/4. 0/. The Creative Commons Public Domain Dedication waiver (http:// creat iveco mmons. org/ publi cdoma in/ zero/1. 0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
Background The severe acute respiratory syndrome coronavirus 2 (SARS-COV-2) pandemic has caused over 173 mil- lion cases worldwide, with over 3.7 million deaths, by June 2021 [1]. Healthcare provider (HCP) infections are thought to comprise a notable proportion of these cases [2]. The emergency department (ED) is a high-risk set- ting of exposure to SARS-COV-2, especially given the performance of aerosol-generating medical procedures
Open Access
International Journal of Emergency Medicine
*Correspondence: ahmed.taher@mail.utoronto.ca
2 Department of Medicine, Division of Emergency Medicine, University of Toronto, C. David Naylor Building, 6 Queen’s Park Crescent West, Third Floor, Toronto, ON M5S 3H2, Canada Full list of author information is available at the end of the article
Page 2 of 8Taher et al. International Journal of Emergency Medicine (2022) 15:62
(AGMPs) such as endotracheal intubation [3]. There- fore, EDs have sought to follow international guidance on personal protective equipment (PPE) practices [4], increase isolation rooms [5], utilize dedicated intuba- tion teams [6] [7], increase the use of ED telemedicine [8], and implement updated resuscitation guidance [9].
Patients with SARS-COV-2 requiring intubation [10] represent a risk to HCPs [11] given the potential aero- solization of the virus [12]. The risk is increased, in part because of the duration of exposure and proximity to the patient increase [13]. Therefore, it is important to minimize the door opening of AGMP patients, which may transmit aerosolized material and to minimize the number of HCPs present inside isolation rooms. It is also necessary to minimize doffing and re-donning of PPE, which produces self-contamination risks and over-use of resources [14].
Given the importance of maintaining closed isolation room doors, it is important to continue clear commu- nication between HCPs across closed doors. This pre- sented multiple challenges for our ED during the first wave of the pandemic, which prompted multiple in situ simulation exercises [15] to prepare HCPs and increase safety practices with AGMPs. A variety of communi- cation methods across the closed doors were trialed, including a “call bell” system (paging the nurse desk), a whiteboard and marker through glass windows, and finally a commercial baby monitor system, which was in use during our baseline period. Communication remained the most common challenge cited in debriefs after ED resuscitations.
Given the communication challenges across isolation rooms along with the risks of the continued door open- ing and potential doffing and re-donning, we conducted a quality improvement (QI) initiative with the aim to decrease isolation room door opening and improve HCP communication clarity.
Methods Study setting Toronto General Hospital (TGH) is a quaternary care adult academic 371-bed medical center part of the Uni- versity Health Network in Toronto, Ontario, Canada. TGH is a referral center for SARS-COV-2 critically ill patients. The ED sees approximately 55,000 patients per year. The TGH ED has 24 beds in the acute zone, of which 17 are isolation rooms. Isolation rooms have glass doors with curtains outside them (for privacy) or anterooms leading to the patient room. Baseline com- munication with the outside team was through door opening or by using a commercial baby monitor system.
Participants Our HCP participants consisted of ED nurses, physi- cians, and trainees on shift during the data collection period. Patient encounters were included if the patient was greater than 16 years old, placed in an isolation room, and met any of the potential infectious symp- toms for SARS-COV-2 (Additional file 1: Appendix 1), regardless of presentation acuity.
Study design This QI initiative was completed in a prospective, multi-stage approach consisting of a baseline period and three discrete Plan-Do-Study-Act (PDSA) cycles. The SQUIRE 2.0 Guidelines were used for study design and reporting [16]. We received a formal exemption from our organization’s research ethics board. We were supported by a local grant: The Mount Sinai Hospital - University Health Network (MSH-UHN) Academic Medicine Organization COVID-19 Innovation Grant.
Interventions An initial period of stakeholder engagement was undertaken at daily nursing huddles and ED physician business meetings, which helped devise three sequen- tial PDSA cycles. PDSA-1 was an educational inter- vention, whereby HCPs were educated about the need and rationale for closing isolation room doors through email and daily departmental huddles.
Given the limitations of educational interventions, concurrent search was done for a better informa- tion communication technology (ICT). PDSA-2 was the introduction of an ICT developed by a co-author (CC) called the TQC 200 “the transceiver” [17]. The transceiver was initially developed for use in sports but was adapted to our local ED setting to replace the existing use of the commercial baby monitor system. The transceivers are wireless radiofrequency wearable paired headsets that allow for two-way uninterrupted communication across closed isolation room doors (Fig. 1). HCPs entering an isolation room would wear one transceiver headset as part of the PPE donning process, and the remaining team members would wear paired transceiver headsets. Team members would then be able to talk to each other freely without push- ing any buttons. Devices would be disinfected accord- ing to approved infection prevention and control protocols and plugged in to recharge in between use. PDSA-3 was mainly to embed and sustain the change that was noted. It included a local clinical champion (registered nurse) who was identified to demonstrate, remind, and support nurses and physicians in their use of the new technology available. During this phase,
Page 3 of 8Taher et al. International Journal of Emergency Medicine (2022) 15:62
further refinement of the transceiver was also made based on ongoing clinician feedback.
Data collection A dedicated research coordinator (SG; not involved in transceiver development) directly observed patient encounters that met the inclusion criteria throughout the baseline period and PDSA cycles. The total number of minutes HCPs spent during a clinical encounter (an episode of going into the room until exiting it) and the total number of minutes of door opening during that encounter were recorded. Data collection consisted of a convenience sample that occurred during research coor- dinator availability, between 09:00 and 15:00 on week- days, between May and July 2020.
Baseline surveys were distributed to nursing staff and included quantitative (Likert scale) and qualitative (two open-ended questions) portions. Following the con- clusion of PDSA 3, a follow-up survey was collected (Additional file 1: Appendix 2). Likert scale surveys included rating communication clarity, communication errors, the need for opening isolation doors to commu- nicate, using other means to communicate (e.g., hand gestures, whiteboard), or having to alter clinical work- flow to communicate adequately. The follow-up survey included questions comparing the baby monitor system to the transceiver. Finally, open-ended questions were asked about the team’s communication approach. Sur- veys were collected in the ED during HCP shifts. All surveys were developed locally and piloted with a sam- ple of HCP prior to administration. Transceiver-specific
qualitative data was collected during PDSAs 2 and 3 to allow for rapid improvements during the corresponding PDSA cycle.
Outcome and balancing measures The primary outcome measure was the duration of the isolation room door opening, which was the percentage of time the isolation room door was left open between the HCP(s) entering and exiting the room for that cor- responding patient encounter. Our aim was to decrease isolation room door opening by 50% over a period of 3 months. Our secondary outcome measure was HCP- perceived communication clarity across closed isolation room doors. This was measured using a Likert scale sur- vey. Our aim was to increase HCP-perceived communi- cation clarity using an ICT by one point on a five-point Likert scale over a period of 3 months.
Our balancing measures were HCP-perceived com- munication errors across the closed isolation room door during patient encounters, perceived use of worka- rounds across the closed door (e.g., hand gestures), and HCP pausing their workflow to communicate with team members outside rooms (all using a Likert scale). We also measured the percentage of patients seen in the ED who were placed in isolation rooms, i.e., failed triage screen presenting with one or more infectious symptoms to monitor the infectious landscape.
Data analysis We utilized Statistical Process Control ([SPC] or Shewhart) XbarR charts [18] to assess for special cause variation. SPC charts were completed with QI Macros© (Version 2018.04, KnowWare International Inc., Denver, CO, USA) for Microsoft© Excel© (Microsoft Corporation, Redmond, WA, USA, Version 14.5.9). Centerline calcula- tion utilized formulae [19] and control limits utilized rules recommended by the Institute for Healthcare Improve- ment [20]. Four discrete periods of data were collected (baseline and three PDSA cycles). For the baseline and follow-up surveys, we used a two-tailed Mann-Whitney U test to assess for statistical significance between means, with a significance level set at p<0.05. The qualitative (open-ended questions) underwent thematic analysis [21].
Exclusion criteria for points used to create the SPC charts included: subgroups with less than three data points according to accepted rules [22], patients who passed the triage infectious assessment (no infectious symptoms reported), patient interactions lasting less than 5 min (chosen as a minimum to include meaningful clinical interaction), and patient interactions when HCP were discharging patients from the ED (e.g., disconnect- ing from monitors and patient leaving).
Fig. 1 Transceiver headset
Page 4 of 8Taher et al. International Journal of Emergency Medicine (2022) 15:62
Results Direct observation of 174 patient encounters was completed over 33 days. Ninety-five encounters were included in the final analysis, while 79 met exclusion criteria and were removed from the analysis (38 did not fail the infectious screen; 25 encounters were less than 5 min; 16 subgroup days had less than three data points). Door opening (primary outcome) increased from base- line (n=40) with a mean of 72.97 to 96.93% in PDSA-1 (n=28; p=0.04). Then, there was a statistically signifi- cant decrease as compared to the baseline in PDSA-2 (n=6; mean 1.58%; p<0.001) and in PDSA-3 (n=21; mean 1.47%; p<0.0001). The primary outcome across the study
period is illustrated in the SPC chart shown in Fig. 2. The baseline period and PDSA-1 met the criteria for special cause variation, i.e., a possible external influence of the system that would need investigation. An improvement was noted by PDSA-2 and 3, with no further special cause variation.
The themes that emerged from the baseline nurs- ing staff qualitative survey (when HCPs were using baby monitors) were varied (Table 1). Positive find- ings included the facilitation of isolation door closure, increased HCP safety, and some positive effects on com- munication. Negative effects pertained to workflow and difficulty with the quality of communication, and they
Fig. 2 Isolation door opening XbarR statistical process control chart. CL center line; UCL upper control limit; LCL lower control limit; PDSA Plan-Do-Study-Act; medians are adjacent to the control limits
Table 1 Qualitative responses during baseline period
Facilitation of isolation door closure “Allowed room door to remain closed”
Increased healthcare worker safety “Allows the team to minimize exposure to potentially infected COVID patients”
Positive effects on communication “The clarity and quality of communication with the baby monitors is satisfactory most of the time”
Challenges with workflow Proximity to device: “Person has to be close to the baby monitor in order to hear orders” Lack of portability: “Having to walk to the baby monitor to speak into it” Multiple people taking: “Radio communication ➔ information not being heard if both sides are talking the same time or one side starts talking and other side has not realized or heard”
Poor quality of communication Ambient noise: “Hard to hear if lots of background noise or if department is busy” Poor clarity/static: “Hard to hear. Too much static” Hardware: “Requires occasionally troubleshooting, sound quality is often poor, especially through mask and shield. Issues with connecting between monitors”
Page 5 of 8Taher et al. International Journal of Emergency Medicine (2022) 15:62
exceeded the positive findings in both frequency and intensity.
The qualitative themes identified post-PDSA-3 nurs- ing staff survey included increased HCP safety, facilita- tion of door closure, positive communication effects, and increased patient confidentiality (Table 2). Moreover, positive effects were noted on workflow including being hands-free, improved efficiency, and portability. Themes also included some challenges with the new workflow, device design, and sound quality with the initial roll-out.
Our secondary outcome of HCP-perceived sound clarity had a statistically significant improvement on the Likert survey (5 is best) from baseline (n=36; mean 3.36) to post-PDSA-3 (n=49; mean 4.21; p<0.001). Our three balancing measures of HCP-perceived errors, using other means of communication and alterations in workflow did not show any worsening; in fact, they all showed statistically significant improvements (Table 3). The post-PDSA-3 survey also showed the majority of respondents agreeing or strongly agreeing with utiliz- ing the transceiver in making care for their patients less stressful, improved satisfaction with communication, and improved quality as compared to the previously used baby monitor (Fig. 3).
Throughout the PDSA cycles, specific feedback was also obtained to improve the transceiver system and bet- ter integrate it into the workflow. Feedback and result- ant actions are illustrated in Table 4. Finally, changes in
patients presenting to the ED who failed the infectious screen are shown in Additional file 1: Appendix 3 and appear to show a slow general decline in patients meeting the criteria as the study period progressed.
Discussion Our QI project with [1] HCP education, [2] the intro- duction of the transceiver, and [3] engagement through a local nurse champion substantially decreased door opening for infectious patients beyond the apriori 50% aim (primary outcome). Our secondary out- come also showed a significant improvement, but nar- rowly missing our a priori one Likert scale point. This improvement in communication was also supported
Table 2 Qualitative responses after PDSA-3
PDSA Plan-Do-Study-Act
Increased healthcare worker safety/stress “Being able to have less people in the room during codes which makes it less stressful”
Facilitation of isolation door closure “Good audio volume. Not needing to open main door. Having members on outside of code room being in communication loop”
Positive effects on workflow Hands free: “Not having to push any buttons e.g. baby monitor to talk - constant communication (even when in medication room, etc.)” Efficiency in patient care: “Not having to leave the room. Orders received right away and initiated - no lag time” Portability: “I was still able to hear the communication inside of the room when I was away to get blood work supplies”
Challenges with workflow Multiple/different groups of people speaking: “Sometimes people talk at the same time and it can get confusing at times” Difficulties with others not using transceiver: “Can’t hear people not on transceiver --> maybe have one ear shorter?” Length of use: “Needs longer battery life for long codes”
Positive effects on communication Clarity: “Clear communication between members wearing headsets. Range is very good (when in medication room the sound is clear)” Ambient Noise: “Can hear everyone clearly. No background noise.”
Increased patient confidentiality “Definitely better than baby monitors in terms of patient confidentiality because people outside the room need not to shout for clearer communica- tion”
Challenges with device design “Uncomfortable after sometime”
Challenges with communication quality “Sometimes the sound quality was not the best but improved with the new set”
Table 3 Baseline and post-PDSA-3 nurse Likert survey
PDSA Plan-Do-Study-Act; Survey Items correspond to Appendix 2
Measure Survey item Baseline mean (n=36)
Post-PDSA 3 mean (n=49)
P value
Outcome Communication clarity
3.36 4.21 <0.001
Balancing Errors 2.61 2.13 0.026
Balancing Door opening 3.00 2.06 <0.001
Balancing Workarounds 3.89 2.46 <0.001
Balancing Alerting workflow 4.08 2.21 <0.001
Page 6 of 8Taher et al. International Journal of Emergency Medicine (2022) 15:62
by qualitative feedback, which demonstrated a positive experience with the transceiver as a communication device. While there have been many reports of frontline HCPs using baby monitors in the ED [23], operating room [24], and intensive care units [25], our qualita- tive surveys revealed substantial problems with their use pertaining to communication clarity and workflow, especially in high-acuity situations when they are even more important for safe patient care.
The special cause variation in the baseline period (Fig. 2) coincided with the addition of new curtains placed in front of isolation room glass doors, which was organized by ED leadership and was unrelated to the conduct of our project. As a result of this increased privacy, HCPs often kept the glass doors open, thereby posing a potential risk to others. A significant decrease in door opening was noted after the introduction of the transceiver by PDSA-2, and it was maintained through PDSA-3. A search for potential confounders leading to
this notable decrease such as a new local policy, incen- tive, or critical incident did not reveal any contributors.
The educational intervention was not noted to have a difference on decreased door opening, but we posit it may have helped prime the participants towards further PDSA cycles along with the transceiver introduction. Moreover, we leveraged a local nurse champion in PDSA-3 to further support and coach HCPs in keeping with the technology acceptance model [26]. While this was anecdotally appre- ciated, a further change was not captured on the SPC chart given the already low values by PDSA-2.
We coupled the introduction of the transceiver with ongoing feedback. The introduction of (ICT) into a com- plex work ecology such as in the ED necessitates ongoing stakeholder engagement and feedback, also referred to as participatory design [27]. This feedback led to ongoing improvements in design and use as illustrated in Table 4. Moreover, key factors that enable the successful adoption and integration of ICTs in healthcare include portability,
Fig. 3 Post-PDSA-3 Likert survey questions. PDSA Plan-Do-Study-Act
Table 4 Feedback and actions for transceiver system improvement
Period Feedback Resultant actions
Baseline period Bulky headsets; casing not robust Changed casing–decreased size of the headset
Foam ear covers (infection control) Eliminated foam ear covers Staff educated on device cleaning
Large ambient noise Noise cancellation optimization
Add an extra speaker in the room Adding a speaker was not found to be effective
Long wire connecting the headset Removed wire and mounted device on the headset
PDSA 1 Battery life short for long resuscitations Improved battery life to 2 h Staff educated on device docking and recharging
Static noted Adjusted settings to decrease static and ambient noise
Improve storage and accessibility New storage units placed beside accessible areas in the ED
PDSA 2 Not enough headsets per room for workflow Increased the number of headsets per room to 6 headsets
Casing handling issue Casing upgraded for more robust materials
PDSA 3 Multiple people talking at once Working towards possible 1 ear headset Working towards simulation exercises integrating headsets and new communication patterns
Page 7 of 8Taher et al. International Journal of Emergency Medicine (2022) 15:62
user satisfaction, and communicability [28]. The main focus of the QI approach was to engage relevant stake- holders at every stage of the project, thereby optimiz- ing adoption and integration into the workflow. A large proportion of HCPs found that using the transceiver decreased their stress during patient care and increased their satisfaction in communication with team members.
Qualitative surveys also highlighted further improve- ment opportunities, such as optimizing communica- tion strategies with multiple team members speaking in parallel. While allowing continuous communication for all users (instead of only one at a time like talkie-walkie push-to-talk systems) facilitated efficient communication, there was some confusion when multiple people spoke at the same time, as would be inherent in real-time in-per- son interactions. To optimize continuous communication and shared mental models for resuscitation patients [29], in situ simulation [30] may improve team performance.
Limitations Our pragmatic QI approach used a convenience sample based on the research coordinator availability. We also did not collect patient demographics (to reduce risk to participants), which may limit generalizability. Moreover, (CC) who developed the transceiver was not involved in study design, data collection, and data analysis to miti- gate conflict of interest. Data collection was done as soon as patients entered the isolation rooms prior to the per- formance of infectious illness screening done by nurses, which led to a proportion of encounters being excluded. Moreover, we could not identify an adequate validated survey for our initiative, so a new instrument was devel- oped to capture important measures instead.
The SPC chart shows special cause variation in the baseline period. Methodologically, the baseline should extend until no further special cause variation is noted. However, a change in practice was noted (the addition of curtains, leading to greater door opening), which high- lighted the need to proceed for potential HCP safety. We elected to continue with PDSA-1 after seeing the follow- ing week of baseline data stabilize.
Conclusion Our iterative QI approach with HCP education, trans- ceiver introduction, and leveraging a local champion dem- onstrated significant decreases in the door opening and improvement in communication clarity without increasing communication errors, workarounds, or alterations in the workflow. Future work may include scaling this initiative to other EDs, operating rooms, and intensive care units.
Abbreviations HCP: Healthcare provider; SARS-COV-2: Severe acute respiratory syndrome - coronavirus 2; ED: Emergency department; QI: Quality improvement; PDSA: Plan-Do-Study-Act; AGMPs: Aerosol-generating medical procedures; PPE: Protective equipment; MSH-UHN: Mount Sinai Hospital - University Health Network; ICT: Information communication technology; SPC: Statistical Process Control.
Supplementary Information The online version contains supplementary material available at https:// doi. org/ 10. 1186/ s12245- 022- 00464-y.
Additional file 1.
Acknowledgements We would like to acknowledge Jamie Forest for his contributions to the trans- ceiver design and Konika Nirmanalathan for her assistance with the project.
Authors’ contributions AT: study concept and design acquisition of the data, data analysis, manuscript drafting, and critical revision of the manuscript. PG: study concept and design, acquisition of funding, and critical revision of the manuscript. CC: intellectual content, acquisition of funding, and critical revision of the manuscript. SC: acquisition of the data, data analysis, and critical revision of the manuscript. SG: acquisition of the data, data analysis, and critical revision of the manu- script. JJ: study concept and design, statistical expertise, and critical revision of the manuscript. AD: data analysis and critical revision of the manuscript. LBC: study concept and design, acquisition of funding, drafting of the manuscript, critical revision of the manuscript for important intellectual content, and statistical expertise. The authors read and approved the final manuscript.
Funding The study was supported by a Mount Sinai Hospital - University Health Network (MSH-UHN) Academic Medicine Organization COVID-19 Innovation Grant. The funding body was not involved in the study design, data collection, analysis, interpretation, or manuscript writing.
Availability of data and materials The datasets during and/or analyzed during the current study are available from the corresponding author on reasonable request.
Declarations
Ethics approval and consent to participate We received formal exemption from the University Health Network Research Ethics Board. Given that this was a local quality improvement project with no patient information was collected and no research assistant interaction with patients; therefore, the exemption was obtained, and no consent was collected.
Consent for publication Not applicable.
Competing interests CC developed the transceiver [17], and this was adapted for ED use with the support of the innovation grant. The other authors declare no competing interests.
Author details 1 University Health Network, 200 Elizabeth Street R. Fraser Elliott Building, Ground Floor, Room 480, Toronto, ON M5G 2C4, Canada. 2 Department of Medicine, Division of Emergency Medicine, University of Toronto, C. David Naylor Building, 6 Queen’s Park Crescent West, Third Floor, Toronto, ON M5S 3H2, Canada. 3 Department of Family and Community Medicine, Division of Emergency Medicine, University of Toronto, 500 University Avenue, Fifth Floor, Toronto, ON M5G 1V7, Canada.
Page 8 of 8Taher et al. International Journal of Emergency Medicine (2022) 15:62
• fast, convenient online submission
•
thorough peer review by experienced researchers in your field
• rapid publication on acceptance
• support for research data, including large and complex data types
•
gold Open Access which fosters wider collaboration and increased citations
maximum visibility for your research: over 100M website views per year •
At BMC, research is always in progress.
Learn more biomedcentral.com/submissions
Ready to submit your researchReady to submit your research ? Choose BMC and benefit from: ? Choose BMC and benefit from:
Received: 18 October 2021 Accepted: 23 October 2022
References 1. COVID Dashboard by the center for systems science and engineer-
ing at Johns Hopkins University. https:// coron avirus. jhu. edu/ map. html Accessed June 6, 2021.
2. ICN calls for data on healthcare worker infection rates and deaths. International Council of Nurses. May 6 2020. https:// www. icn. ch/ news/ icn- calls- data- healt hcare- worker- infec tion- rates- and- deaths Accessed on 23 Aug 2020.
3. Tran K, Cimon K, Severn M, Pessoa-Silva CL, Conly J. Aerosol generating procedures and risk of transmission of acute respiratory infections to healthcare workers: a systematic review. PLoS One. 2012;7(4):e35797.
4. World Health Organization. Rational use of personal protective equip- ment for coronavirus disease 2019 (COVID-19). Published February 27, 2020. https:// apps. who. int/ iris/ bitst ream/ handle/ 10665/ 331215/ WHO- 2019- nCov- IPCPPE_ use- 2020.1- eng. pdf Accessed 23 Aug 2020.
5. Wee LE, Fua TP, Chua YY, et al. Containing COVID-19 in the emergency department: the role of improved case detection and segregation of suspect cases. Acad Emerg Med. 2020;27(5):379–87.
6. Yao W, Wang T, Jiang B, et al. Emergency tracheal intubation in 202 patients with COVID-19 in Wuhan, China: lessons learnt and international expert recommendations. Br J Anaesth. 2020;125(1):e28–e3.
7. Sorbello M, El-Boghdadly K, Di Giacinto I, et al. The Italian coronavirus dis- ease 2019 outbreak: recommendations from clinical practice. Anaesthesia. 2020 Jun;75(6):724–32.
8. Turer RW, Jones I, Rosenbloom ST, et al. Electronic personal protective equipment: a strategy to protect emergency department providers in the age of COVID-19. JAMIA. 2020;27(6):967–71.
9. Brewster DJ, Chrimes NC, Do TB, et al. Consensus statement: Safe Airway Society principles of airway management and tracheal intubation spe- cific to the COVID-19 adult patient group. Med J Aust. 2020;16:16.
10. Grasselli G, Zangrillo A, Zanella A, et al. Baseline characteristics and outcomes of 1591 patients infected with SARS-CoV-2 admitted to ICUs of the Lombardy Region, Italy. JAMA. 2020;323(16):1574–81.
11. Weissman DN, De Perio MA, Radonovich LJ. COVID-19 and risks posed to personnel during endotracheal intubation. JAMA. 2020;323(20):2027–8.
12. Simpson JP, Wong DN, Verco L, et al. Measurement of airborne particle exposure during simulated tracheal intubation using various proposed aerosol containment devices during the COVID-19 pandemic. Anaesthe- sia. 2020;19.
13. Wilson NM, Norton A, Young FP, et al. Airborne transmission of severe acute respiratory syndrome coronavirus-2 to healthcare workers: a narra- tive review. Anaesthesia. 2020;20.
14. Kwon JH, Burnham CA, Reske KA, et al. Assessment of healthcare worker protocol deviations and self-contamination during personal protec- tive equipment donning and doffing. Infect Control Hosp Epidemiol. 2017;38(9):1077–83.
15. Dharamsi A, Hayman K, Yi S, et al. Enhancing departmental preparedness for COVID-19 using rapid cycle in situ simulation. J Hosp Infect. 2020:13.
16. Ogrinc G, Davies L, Goodman D, et al. SQUIRE 2.0 (Standards for QUality Improvement Reporting Excellence): revised publication guidelines from a detailed consensus process. J Contin Educ Nurs. 2015;46(11):501–7.
17. Culligan C., Shellhammer B., (2015). US10250514B2. United States. Accessed online August 23 2020. https:// paten ts. google. com/ patent/ US201 70171 093A1/ en.
18. Fretheim A, Tomic O. Statistical process control and interrupted time series: a golden opportunity for impact evaluation in quality improve- ment. BMJ Qual Saf. 2015;24(12):748–52.
19. Xbar Chart Formulas. QI Macros. https:// www. qimac ros. com/ contr ol- chart- formu las/x- bar-r- chart- formu la/Accessed 20 Oct 2020.
20. Control Chart rules are used to perform stability analysis. QI Macros. 2019. Accessed online December 11 2019. https:// www. qimac ros. com/ contr ol- chart/ stabi lity- analy sis- contr ol- chart- rules/ Accessed 20 Oct 2020
21. Charmaz K, Belgrave LL. Grounded theory. The Blackwell encyclopedia of sociology; 2007. p. 15.
22. Provost LP, Murray S. The health care data guide: learning from data for improvement: John Wiley & Sons; 2011 Oct 4.
23. St. Joe’s turns to baby monitors to communicate during COVID-19 pandemic. Hamilton. CBC News. April 29 2020. https:// www. cbc. ca/ news/ canada/ hamil ton/ baby- monit ors- covid- 19-1. 55488 23 Accessed 2 Nov 2020.
24. Hospital Staff use Baby Monitors to communicate. Misericordia operating room adds two-way technology to enhance safety. The vital beat Jun 8 2020. https:// www. thevi talbe at. ca/ news/ hospi tal- staff- use- baby- monit ors- commu nicate/ Accesses 2 Nov 2020.
25. Baby monitors in the ICU: nurses get creative to save lives, critical equip- ment. ABC news. Aprl 1 2020. https:// abcne ws. go. com/ Health/ nurses- find- monit or- patie nts- virus- baby- monit ors/ story? id= 69907 152 Accessed 2 Nov 2020.
26. Davis FD. Perceived usefulness, perceived ease of use, and user accept- ance of information technology. MIS quarterly. 1989;1:319–40.
27. Sjöberg C, Timpka T. Participatory design of information systems in health care. JAMIA, 1998;5(2):177–83.
28. Chatterjee S, Chakraborty S, Sarker S, et al. Examining the success factors for mobile work in healthcare: a deductive study. Decis Support Syst. 2009;46(3):620–33.
29. McComb S, Simpson V. The concept of shared mental models in health- care collaboration. J Adv Nurs. 2014;70(7):1479–88.
30. Petrosoniak A, Auerbach M, Wong AH, et al. In situ simulation in emer- gency medicine: moving beyond the simulation lab. Emerg Med Australas. 2017;29(1):83–8.
Publisher’s Note Springer Nature remains neutral with regard to jurisdictional claims in pub- lished maps and institutional affiliations.
International Journal of Emergency Medicine is a copyright of Springer, 2022. All Rights Reserved.
- Improving safety and communication for healthcare providers caring for SARS-COV-2 patients
- Abstract
- Background:
- Methods:
- Results:
- Conclusion:
- Background
- Methods
- Study setting
- Participants
- Study design
- Interventions
- Data collection
- Outcome and balancing measures
- Data analysis
- Results
- Discussion
- Limitations
- Conclusion
- Acknowledgements
- References